ICU Medical CL3520Recalls & reported problems
Icu Medical, Inc. · 40" (102 cm) Appx 5.2 ml, Admin Set w/2 ChemoLock™, 20 Drop In-Line Drip Chamber, Bag Hanger · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887709057036), received between Aug 2, 2017 and Dec 21, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Connection Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
59 other recalls cover devices cleared under the same FDA submission (K131549); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- May 12, 2026Class IIOpen, Classified
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.
Z-3161-2026 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Connection Problem1
- Fluid/Blood Leak1
- Leak/Splash1
Patient problems
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 21, 2018Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER REPORTED THAT AN ADMINISTRATION SET HAD A HOLE IN THE TOP OF THE WHITE PLASTIC OF THE DRIP CHAMBER AND CHEMOTHERAPY CONTINUE TO DRIP DESPITE CLAMPING. THERE WAS UNKNOWN PATIENT INVOLVEMENT AND UNKNOWN ADVERSE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8190601 ↗MalfunctionOct 27, 2017Leak/Splash; Leak/Splash
COMPLAINT RECEIVED REGARDING ONE CL3520, REPORT STATES: CHEMOTHERAPY WAS HUNG USING THE CHEMO LOCK SECONDARY SET (CL3520). AS SOON AS THE SECONDARY SET WAS ATTACHED TO THE CHEMO BAG IT STATED LEAKING WHERE THE TUBING ATTACHES TO THE BLUE CONNECTOR ABOVE THE DRIP CHAMBER. IT WAS A CONTINUOUS DRIP UNTIL THE NURSE DETACHED THE CL3520 FROM THE CHEMO BAG AND REPLACED IT WITH ANOTHER CL3520 AND LEAKAGE DISCONTINUED. PROPER PPE WAS UTILIZED. NO PATIENT INVOLVEMENT OR ADVERSE CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6984725 ↗MalfunctionAug 2, 2017Connection Problem; Connection Problem
LEAK NOTED BELOW THE UPPER CHEMOLOCK OF CL3520. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6762002 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K131549. One submission often covers many models, so check the product and lot details in each record.
- Z-3159-2026 · May 12, 2026 · Open, ClassifiedOncology sets with drip chambers
- Z-2210-2026 · Mar 24, 2026 · Open, ClassifiedOncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL
- Z-0356-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical Syringe Transfer Set w/MicroClave, ChemoLock Port REF: CL-34
- Z-0358-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical 8" (20 cm) Ext Set w/MicroClave Clear, ChemoLock Port, Y-Connector, Rotating Luer, REF: CL-3950
- Z-0352-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical Graduated Connector w/ChemoLock Port REF: CL3900
- Z-0343-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical ChemoLock Universal Vented Vial Spike REF: CL-70
- Z-0357-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical ChemoLock Port w/Spiros, Red Cap REF: CL-35
- Z-0344-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical ChemoLock Vented Vial Spike, 20mm REF: CL-70S
- Z-0362-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical Oncology Kit w/2 ChemoLock", 13mm Closed Vial Spike w/ChemoLock" Port, ChemoLock" Bag Spike, Ext Set w/ChemoLock" Port, Clamp, Graduated Adapter REF: CL4188
- Z-0364-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical ChemosafeLock Connecter REF: KL-FNU3
- Z-0349-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical ChemoLock 13mm Closed Vial Spike REF: CL-82
- Z-0348-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical ChemoLock Closed Vial Spike w/Skirt, 5 Units REF: CL-80S-5
- Z-0353-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical 6.5" Bag Spike w/ChemoLock Port Additive Port, ChemoLock Port REF: CL3955
- Z-0340-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical 73" (185 cm) Appx 2.5 ml, Ext Set, Smallbore w/ChemoLock Port, Clamp, ChemoLock w/Red Cap REF: CL4176
- Z-0331-2026 · Sep 8, 2025 · Open, ClassifiedICU Medical Clave Bag Spike w/ChemoLock Port REF: CL2150
…and 44 more.
Other Closed antineoplastic and hazardous drug reconstitution and transfer system devices
- BD PhaSeal 515003BD Switzerland Sàrl · Closed antineoplastic and hazardous drug reconstitution and transfer system1 recall183 reports
- BD PhaSeal 515056BD Switzerland Sàrl · Closed antineoplastic and hazardous drug reconstitution and transfer system1 recall159 reports
- BD PhaSeal Optima 515070BD Switzerland Sàrl · Closed antineoplastic and hazardous drug reconstitution and transfer system1 recall154 reports
- Alaris 10012241-0500BD Switzerland Sàrl · Closed antineoplastic and hazardous drug reconstitution and transfer system1 recall138 reports
- BD PhaSeal 515005BD Switzerland Sàrl · Closed antineoplastic and hazardous drug reconstitution and transfer system1 recall128 reports
- ICU Medical CL3011Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system3 recalls26 reports
- BD PhaSeal Optima 515052BD Switzerland Sàrl · Closed antineoplastic and hazardous drug reconstitution and transfer system1 recall114 reports
- ICU Medical CL3511Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system3 recalls6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page