ICU Medical CH3603Recalls & reported problems
Icu Medical, Inc. · 60” (152 cm) Appx 1.9 ml, Smallbore Ext Set w/MicroClave™ Clear, 0.2 Micron Low Protein Binding Filter, Spiros™ w/Red Cap, Clamp · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00887709081147), received between Dec 1, 2020 and Dec 1, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K070532); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fluid/Blood Leak2
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 1, 2020Fluid/Blood Leak; Fluid/Blood Leak
THE EVENT INVOLVED A 60¿ (152 CM) APPX 1.9 ML, SMALLBORE EXT SET W/MICROCLAVE® CLEAR, 0.2 MICRON LOW PROTEIN BINDING FILTER, SPIROS® W/RED CAP, CLAMP THAT THE CUSTOMER REPORTED AN UNSPECIFIED CHEMOTHERAPY MEDICATION LEAKING FROM THE FILTER DURING PATIENT USE. THERE WAS PATIENT INVOLVEMENT; HOWEVER, NO REPORT OF ADVERSE EVENT. THIS EVENT CAPTURES THE FIRST OF TWO INCIDENTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10931254 ↗MalfunctionDec 1, 2020Fluid/Blood Leak; Fluid/Blood Leak
THE EVENT INVOLVED A 60¿ (152 CM) APPX 1.9 ML, SMALLBORE EXT SET W/MICROCLAVE® CLEAR, 0.2 MICRON LOW PROTEIN BINDING FILTER, SPIROS® W/RED CAP, CLAMP THAT THE CUSTOMER REPORTED AN UNSPECIFIED CHEMOTHERAPY MEDICATION LEAKING FROM THE FILTER DURING PATIENT USE. THERE WAS PATIENT INVOLVEMENT, HOWEVER, NO REPORT OF ADVERSE EVENT. THIS EVENT CAPTURES THE SECOND OF TWO INCIDENTS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10931231 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K070532. One submission often covers many models, so check the product and lot details in each record.
- Z-0699-2021 · Aug 31, 2020 · Open, Classified42 IN (107 cm) Appx 5.0 ml, 20 Drop Admin Set, Spiros, REF: 20131-01; 31" (79 cm) Admin Set w/20 Drop Integrated Clave Drip Chamber, Spiros, Bag Hanger Drop-in Red Cap, 4 Units, RE
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page