Skip to content

ICU Medical MC330247Recalls & reported problems

Icu Medical, Inc. · 8" (20 cm) Appx 0.56 ml, Smallbore Bifuse Pressure Infusion (400 psig) Ext Set w/2 Remv MicroClave™ Clear, 2 Purple Clamps, Rotating Luer · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887709081512), received between Oct 28, 2025 and Oct 28, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “No Flow” (1 report).

8 other recalls cover devices cleared under the same FDA submission (K100576); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 28, 2025No Flow; No Flow

    EVENT OCCURRED REGARDING SMALLBORE BIFUSE PRESSURE INFUSION WHERE IT WAS REPORTED THAT THE SMALLBORE BIFUSE PRESSURE INFUSION WITH TWO REMOVABLE MICROCLAVE CLEAR 2 PURPLE CLAMPS WAS USED FOR INFUSION. ONE SIDE OF THE T CONNECTOR HAD BLOCKAGE, WHILE THE OTHER SIDE WORKED. THERE WAS NO LEAKING NOTED DURING THE EVENT. THERE WAS PATIENT INVOLVEMENT BUT NO REPORTED PATIENT HARM NOR DELAY IN THERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23406258 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K100576. One submission often covers many models, so check the product and lot details in each record.

  • Z-3160-2026 · May 12, 2026 · Open, Classified
    IV Sets with drip chambers
  • Z-1720-2022 · Aug 1, 2022 · Open, Classified
    8.5" (22 cm) Appx 0.85 ml, Bifuse Pressure Infusion (400 psig) Ext Set w/2 Remv MicroClave Clear, 2 Purple Clamps, Rotating Luer;
  • Z-0197-2021 · Jul 6, 2020 · Terminated
    7" (18 cm) Appx 0.31 ml, Smallbore Pressure Infusion (400psig) Ext Set w/Remv MicroClave Clear, Clamp, NanoClave T-Connector (Purple Ring), Rotating Luer; unit per pouch, 50 multi
  • Z-0193-2021 · Jul 6, 2020 · Terminated
    6.5" (17 cm) Appx 0.30 ml, Smallbore Pressure Infusion (400psig) Ext Set w/Remv MicroClave, NanoClave T-Connector, Purple Clamp, Rotating Luer. 1 unit per pouch, 50 multivac pouche
  • Z-0192-2021 · Jul 6, 2020 · Terminated
    9.5" (24 cm) Appx 0.45 ml, Smallbore Pressure Infusion (400 psig) Ext Set w/MicroClave, 2 NanoClave, NanoClave T-Connector, 3 Purple Clamps, Rotating Luer. 1 unit per pouch, 50 mul
  • Z-0201-2021 · Jul 6, 2020 · Terminated
    7" (18 cm) Appx 0.32 ml, Smallbore Pressure Infusion (400 psig) Ext Set w/Remv MicroClave Clear, NanoClave, Purple Clamp, Luer Lock; 1 unit per pouch, 50 multivac pouches per case.
  • Z-0204-2021 · Jul 6, 2020 · Terminated
    6.5" (17 cm) Appx 0.35 ml, Smallbore Pressure Infusion (400 psig) Ext Set w/Remv MicroClave Clear, NanoClave, Purple Clamp, Rotating Luer; 1 unit per pouch, 50 multivac pouches per
  • Z-1916-2020 · Oct 4, 2019 · Terminated
    REF: Z2560, 100" 10 Drop Admin Set w/Hand Pump, 2 Clave", Rotating Luer, 1 Pressure Infusion (400psig) Ext, Purple Clamp, UDI: (01)(01)00887709026889

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page