ICU Medical MC330623Recalls & reported problems
Icu Medical, Inc. · 9" (23 cm) Appx 0.67 ml, Ext Set Pressure Infusion (400 psig) w/2 MicroClave™ Clear, Purple Clamp, Rotating Luer · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887709097667), received between Apr 1, 2025 and Sep 15, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
8 other recalls cover devices cleared under the same FDA submission (K100576); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak3
- Crack1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50777905 ↗MalfunctionMay 23, 2026Fluid/Blood Leak
COMPLAINT EMAIL FROM (B)(6), SUPERVISOR OF (B)(6) HOSPITAL WAS RECEIVED WITH REGARDS TO A 9" (23 CM) APPX 0.67 ML, EXT SET PRESSURE INFUSION (400 PSIG) W/2 MICROCLAVE CLEAR, PURPLE CLAMP, ROTATING LUER WHERE THE REPORTER STATED THAT THE PATIENT WENT TO CT SCAN AND WHEN THEY CAME BACK, THE IV TUBING-RUBBER PLUG INSIDE THE MICROCLAVE HAD PROLAPSED OUTSIDE OF THE PLASTIC MICROCLAVE. THE EVENT OCCURRED IN EMERGENCY DEPT. AT (B)(6). THERE WAS PATIENT INVOLVEMENT BUT UNKNOWN ADVERSE EVENT. UNSURE OF HOW THIS OCCURRED. THERE WERE NO ADVERSE EVENT AND NO DELAY IN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25266291 ↗MalfunctionApr 1, 2025Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
THE EVENT INVOLVED A 9" (23 CM) APPX 0.67 ML, EXT SET PRESSURE INFUSION (400 PSIG) W/2 MICROCLAVE¿ CLEAR, PURPLE CLAMP, ROTATING LUER WHERE IT WAS REPORTED THAT DURING THE INTRAVENOUS (IV) CONTRAST INJECTION, THE CONTRAST FLOWED THROUGH THE CLAMP OFF SECOND LUMEN OUT ONTO THE PATIENT. THE VALVE BROKE AND THE RUBBER/SILICON PIECE STICKING OUTSIDE OF HUB. THE PURPLE CLAMP WAS CLOSED. THE INJECTION OF CONTRAST WAS GOING INTO THE UNCLAMPED LUMEN. THE PRESSURE CAUSED THE CONTRAST TO FLOW UP THROUGH THE OTHER CLAMPED OFF LUMEN AND FORCED THE RUBBER VALVE TO POP OUT OF IV. THE CONTRAST/LIQUID THEN FLOWED ALL OVER THE PATIENT INTO THE COMPUTED TOMOGRAPHY (CT) TABLE. THERE WAS PATIENT INVOLVEMENT AND UNKNOWN PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21742812 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K100576. One submission often covers many models, so check the product and lot details in each record.
- Z-3160-2026 · May 12, 2026 · Open, ClassifiedIV Sets with drip chambers
- Z-1720-2022 · Aug 1, 2022 · Open, Classified8.5" (22 cm) Appx 0.85 ml, Bifuse Pressure Infusion (400 psig) Ext Set w/2 Remv MicroClave Clear, 2 Purple Clamps, Rotating Luer;
- Z-0197-2021 · Jul 6, 2020 · Terminated7" (18 cm) Appx 0.31 ml, Smallbore Pressure Infusion (400psig) Ext Set w/Remv MicroClave Clear, Clamp, NanoClave T-Connector (Purple Ring), Rotating Luer; unit per pouch, 50 multi
- Z-0201-2021 · Jul 6, 2020 · Terminated7" (18 cm) Appx 0.32 ml, Smallbore Pressure Infusion (400 psig) Ext Set w/Remv MicroClave Clear, NanoClave, Purple Clamp, Luer Lock; 1 unit per pouch, 50 multivac pouches per case.
- Z-0204-2021 · Jul 6, 2020 · Terminated6.5" (17 cm) Appx 0.35 ml, Smallbore Pressure Infusion (400 psig) Ext Set w/Remv MicroClave Clear, NanoClave, Purple Clamp, Rotating Luer; 1 unit per pouch, 50 multivac pouches per
- Z-0193-2021 · Jul 6, 2020 · Terminated6.5" (17 cm) Appx 0.30 ml, Smallbore Pressure Infusion (400psig) Ext Set w/Remv MicroClave, NanoClave T-Connector, Purple Clamp, Rotating Luer. 1 unit per pouch, 50 multivac pouche
- Z-0192-2021 · Jul 6, 2020 · Terminated9.5" (24 cm) Appx 0.45 ml, Smallbore Pressure Infusion (400 psig) Ext Set w/MicroClave, 2 NanoClave, NanoClave T-Connector, 3 Purple Clamps, Rotating Luer. 1 unit per pouch, 50 mul
- Z-1916-2020 · Oct 4, 2019 · TerminatedREF: Z2560, 100" 10 Drop Admin Set w/Hand Pump, 2 Clave", Rotating Luer, 1 Pressure Infusion (400psig) Ext, Purple Clamp, UDI: (01)(01)00887709026889
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page