Skip to content

ICU Medical 42800-23Recalls & reported problems

Icu Medical, Inc. · Transpac™IT Monitoring Kit Neonatal, 12", 3 Stopcocks, 30ml Flush (For Use with Infusion Pump) · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00887709097964), received between Nov 14, 2024 and Nov 21, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K052828); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deformation Due to Compressive Stress2
  • Inaccurate Information2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 21, 2024Deformation Due to Compressive Stress; Inaccurate Information; Deformation Due to Compressive Stress; Inaccurate Information

    THE EVENT INVOLVED A TRANSPAC¿ IT MONITORING KIT NEONATAL, 12", 3 STOPCOCKS, 30ML FLUSH (FOR USE WITH INFUSION PUMP). THE REPORTER STATED THAT THE PRODUCT WAS INVOLVED IN A PATIENT SAFETY REPORT IN THE FACILITY. THE ISSUE WAS THAT THE IV TUBING WAS BENT CAUSING AN ALARM FOR HIGH PRESSURE. THE WAVEFORM WAS DAMPENED OR WOULD READ AS DISCONNECTED AND WOULD CAUSE ALARMS TO GO OFF. THERE WAS PATIENT INVOLVEMENT AND THERE WAS DELAY IN THERAPY; HOWEVER, THERE WAS NO HUMAN HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20752410 ↗
  • MalfunctionNov 14, 2024Deformation Due to Compressive Stress; Inaccurate Information; Deformation Due to Compressive Stress; Inaccurate Information

    THE EVENT INVOLVED A TRANSPAC¿ IT MONITORING KIT NEONATAL, 12", 3 STOPCOCKS, 30ML FLUSH (FOR USE WITH INFUSION PUMP) WHERE THE REPORTER STATED THAT THE IV TUBING IS BENT CAUSING AN ALARM FOR HIGH PRESSURE. IT WAS STATED THAT WHEN THEY TRIED TO ZERO THE LINE, THE WAVEFORM WAS DAMPENED OR WOULD READ AS DISCONNECTED. IT WOULD CAUSE ALARMS TO GO OFF. AFTER TROUBLESHOOTING WITH A FLEX NURSE, THEY NOTICED THAT THE LINE ENTERING THE TRANSDUCER WAS COMPRESSED DOWN. THEY STRAIGHTENED THE IV TUBING CONNECTING TO THE TRANSDUCER AND GOT AN ACCURATE READING, HOWEVER AS SOON AS RELEASED THE TUBING AGAIN WAS FACED WITH THE SAME ALARMS AND DAMPENED WAVE FORM. THERE WAS PATIENT INVOLVEMENT, AND NO HUMAN HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20684408 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K052828. One submission often covers many models, so check the product and lot details in each record.

  • Z-1071-2019 · Nov 27, 2018 · Terminated