Skip to content

ICU Medical 42800-27Recalls & reported problems

Icu Medical, Inc. · Transpac™IT Monitoring Kit, 60" Safeset™Reservoir, 2 Blood Sampling Port, 3ml Flush Device, Macrodrip · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00887709097995), received between Aug 4, 2023 and Feb 25, 2025 — classified as 1 death, 3 malfunctions.

The most frequently reported specific device problem is “Disconnection” (1 report).

1 other recall cover devices cleared under the same FDA submission (K052828); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Disconnection1
  • Fluid/Blood Leak1
  • Material Deformation1
  • Material Separation1
  • Separation Failure1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 25, 2025Material Separation; Material Separation

    THE EVENT INVOLVED A TRANSPAC¿ IT MONITORING KIT, 60" SAFESET RESERVOIR, 2 BLOOD SAMPLING PORT, 3ML FLUSH DEVICE, MACRODRIP WHERE IT WAS REPORTED THAT WHILE ATTEMPTING TO GET PATIENT READY TO TRANSPORT FOR CT, ARTERIAL LINE TUBING BECAME DISCONNECTED, CAUSING BLOOD TO FLOW FREELY. BEDSIDE MONITOR ALARMED AND ISSUE WAS DISCOVERED QUICKLY, SO BLOOD LOSS WAS MINIMAL. NO CLEAR EPISODE OF ANYTHING GETTING CAUGHT ON TUBING CAUSED THE SEPARATION OF TUBING. THERE WAS PATIENT INVOLVEMENT AND NO ADVERSE EVENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 21462005 ↗
  • DeathFeb 16, 2024Disconnection; Disconnection

    THE EVENT INVOLVED TRANSPAC¿ IT MONITORING KIT, 60" SAFESET RESERVOIR, 2 BLOOD SAMPLING PORT, 3ML FLUSH DEVICE, MACRODRIP WHERE A DISCONNECTION OF TUBING AT BLOOD SAMPLING PORT WAS REPORTED. THE PRODUCT WAS NOT DISCOLORED DURING THE ENTIRE USE. THE EVENT OCCURRED DURING INFUSION AND THERE WAS PATIENT HARM DUE TO BLEEDING THAT OCCURRED WHEN THE DEVICE MALFUNCTIONED AND BROKE APART. THE PATIENT NEEDED TO BE TURNED AND REPOSITIONED FOR CLEAN-UP. PATIENT SUSTAINED RESPIRATORY ARREST WHILE STAFF WAS ATTEMPTING TO CLEAN THE PATIENT UP. PATIENT THEN EXPIRED ONE DAY AFTER INITIAL ARREST AND PRODUCT ISSUE, AND AFTER BEING MADE COMFORT CARE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 18727280 ↗
  • MalfunctionJan 17, 2024Separation Failure; Separation Failure

    IT WAS REPORTED THAT A TRANSPAC¿ IT MONITORING KIT, 60" SAFESET RESERVOIR, 2 BLOOD SAMPLING PORT, 3ML FLUSH DEVICE, MACRODRIP DISCONNECTED DURING PATIENT USE ON PATIENT. THE PRODUCT APPEARS TO HAVE FAILED AT THE DISTAL CAD PORT. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18531911 ↗
  • MalfunctionAug 4, 2023Fluid/Blood Leak; Material Deformation; Fluid/Blood Leak; Material Deformation

    IT WAS REPORTED THAT A TRANSPAC¿ IT MONITORING KIT, 60" SAFESET RESERVOIR, 2 BLOOD SAMPLING PORT, 3ML FLUSH DEVICE, MACRODRIP SQUEEZE FLUSH MECHANISM APPEARS ¿TURNED¿ AND WHEN THE REGISTERED NURSE (RN) TRIED TO STRAIGHTEN IT, IT BEGAN LEAKING FROM THE FLUSH MECHANISM AREA. THE PRODUCT WA USED FOR ONE HOUR. THERE WAS PATIENT INVOLVED BUT NO HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17461109 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K052828. One submission often covers many models, so check the product and lot details in each record.

  • Z-1071-2019 · Nov 27, 2018 · Terminated
    ICU Medical, Transpac IV Monitoring Kit With Safeset Reservoir and Blood Sampling Port, 60" Tubing, Disposable Transducer, 3 ml Squeeze Flush, Macrodrip (Pole Mount), REF: 42640-06

Other Transducer, blood-pressure, extravascular devices

All Transducer, blood-pressure, extravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page