Skip to content

LifeShield Infusion Safety Software Suite - LifeShield DLM 2.1Recalls & reported problems

Icu Medical, Inc. · Infusion Safety Management Software · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. May 2, 2025Class IIOpen, Classified

    Firm has identified Drug Library Management defects in the software: 1) DLM software does not allow user to create percentage dose rate or rate change values with certain limits, potentially resulting in over-delivery to patient. 2) An unauthorized user may modify and approve a drug library, potentially leading to incorrect program parameters being used for therapy.

    Z-1763-2025 · Root cause: Software Design Change · Recall details · FDA record ↗

Other Infusion safety management software devices

All Infusion safety management software devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page