Spectrum IQ 3570009Recalls & reported problems

Baxter Healthcare Corporation · The Spectrum IQ Infusion System with Dose IQ Safety Software is an infusion pump that provides delivery of fluids into a patient in a controlled manner. The pumping mechanism is of linear peristaltic design and uses inlet and outlet valves for flow control. The Spectrum IQ infusion pump is specifically manufactured and calibrated for use with Baxter administration sets. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

9 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Baxter Spectrum IQ infusion pump appears in 8,979 adverse event reports, including 11 deaths and 187 injuries. The reports mostly describe false alarms, filling problems, and alarms or audible prompts that fail to sound. Of its 11 recalls, 9 are open and the other two are closed; the 9 open recalls cover missing screws, doors that may not close fully, and false occlusion alarms after software updates.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
9,086
Deaths reported
10
Injuries reported
187
Recalls naming it
11

Summary

The FDA's MAUDE database contains 9,086 adverse event reports that list this exact device (DI 00085412610900), received between Oct 3, 2018 and Sep 30, 2026 — classified as 10 deaths, 187 injuries, 8,889 malfunctions.

The most frequently reported specific device problem is “False Alarm” (1,983 reports).

11 FDA recall records name this device's identifier, 9 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 142 reportsDec 2023: 127 reportsJan 2024: 123 reportsFeb 2024: 116 reportsMar 2024: 136 reportsApr 2024: 99 reportsMay 2024: 158 reportsJun 2024: 192 reportsJul 2024: 183 reportsAug 2024: 103 reportsSep 2024: 74 reportsOct 2024: 105 reportsNov 2024: 131 reportsDec 2024: 69 reportsJan 2025: 129 reportsFeb 2025: 133 reportsMar 2025: 137 reportsApr 2025: 113 reportsMay 2025: 177 reportsJun 2025: 138 reportsJul 2025: 169 reportsAug 2025: 134 reportsSep 2025: 31 reportsOct 2025: 60 reportsNov 2025: 102 reportsDec 2025: 113 reportsJan 2026: 101 reportsFeb 2026: 106 reportsMar 2026: 130 reportsApr 2026: 173 reportsMay 2026: 145 reportsJun 2026: 80 reportsJul 2026: 83 reportsAug 2026: 126 reportsSep 2026: 127 reportsOct 2026: 0 reports202420252026192
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 13, 2026Class IIOpen, Classified

    Pumps were released without full testing being performed, including occlusion alarm testing.

    Z-1368-2026 · Root cause: Employee error · Recall details · FDA record ↗

  2. Jul 14, 2025Class IIOpen, Classified

    Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.

    Z-2415-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  3. Feb 5, 2025Class IOpen, Classified

    There is the potential for missing motor mounting screws, which may have occurred during the servicing process.

    Z-1254-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Oct 2, 2024Class IIOpen, Classified

    The door on the Spectrum IQ Infusion pump may not be able to fully close.

    Z-0348-2025 · Root cause: Process control · Recall details · FDA record ↗

  5. Apr 9, 2024Class IIOpen, Classified

    There is a potential for cracks on the mount of the front panel of the device.

    Z-1810-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  6. Mar 20, 2024Class IIOpen, Classified

    One device was improperly performed testing prior to release from a Service Center.

    Z-1775-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  7. Jun 15, 2023Class IOpen, Classified

    A medical device correction is being issued for SIGMA Spectrum Infusion System (V8 Platform) and Spectrum IQ Infusion System with Dose IQ Safety Software infusion pumps due to an increase in reported false upstream occlusion alarms following upgrades to software versions v8.01.01 and v9.02.01.

    Z-2104-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  8. Dec 29, 2021Class IOpen, Classified

    There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spectrum IQ infusion pumps.

    Z-0529-2022 · Root cause: Device Design · Recall details · FDA record ↗

  9. Nov 23, 2020Class IIOpen, Classified

    There is a potential software error during programming.

    Z-0715-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  10. Jun 30, 2020Class ITerminated

    Baxter is updating the device IFU for safety communication regarding cleaning practices.

    Z-0123-2021 · Root cause: Labeling design · Recall details · FDA record ↗

  11. Sep 18, 2018Class IITerminated

    If an operator attempts to manually program a Basic Infusion for a drug that is not included in the Drug Library using the Basic-Dose Mode, the operator will receive a false hard limit exceeded error, preventing him or her from programming the infusion.

    Z-0107-2019 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • False Alarm1,983
  • Filling Problem1,810
  • Failure to Sense1,693
  • Inaudible or Unclear Audible Prompt/Feedback798
  • No Audible Prompt/Feedback516
  • No Display/Image480
  • Defective Alarm383
  • Protective Measures Problem295

Patient problems

  • No Clinical Signs, Symptoms or Conditions7,951
  • No Consequences Or Impact To Patient573
  • No Patient Involvement353
  • Low Blood Pressure/ Hypotension75
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available43
  • High Blood Pressure/ Hypertension33
  • Insufficient Information10
  • Tachycardia9

Recent adverse event reports

The 25 most recent of 9,086 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026False Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951730 ↗
  • MalfunctionSep 30, 2026Protective Measures Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951384 ↗
  • MalfunctionSep 30, 2026Defective Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950931 ↗
  • MalfunctionSep 30, 2026False Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950482 ↗
  • MalfunctionSep 30, 2026Filling Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949664 ↗
  • MalfunctionSep 30, 2026Filling Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949663 ↗
  • MalfunctionSep 30, 2026False Alarm

    It was reported that a Spectrum IQ Infusion Pump alarmed upstream occlusion during preventive maintenance. There was no patient involvement. No additional information is available.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949339 ↗
  • MalfunctionSep 30, 2026Defective Alarm

    It was reported that a Spectrum IQ Infusion Pump failed downstream occlusion test during preventive maintenance. There was no patient involvement. No additional information is available.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948947 ↗
  • MalfunctionSep 30, 2026False Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941447 ↗
  • MalfunctionSep 29, 2026False Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937957 ↗
  • MalfunctionSep 29, 2026Intermittent Loss of Power

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935609 ↗
  • MalfunctionSep 29, 2026False Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926720 ↗
  • MalfunctionSep 28, 2026False Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926364 ↗
  • MalfunctionSep 28, 2026Protective Measures Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926123 ↗
  • MalfunctionSep 28, 2026False Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50925467 ↗
  • MalfunctionSep 28, 2026Defective Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923438 ↗
  • MalfunctionSep 28, 2026Defective Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923307 ↗
  • MalfunctionSep 28, 2026Defective Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923306 ↗
  • MalfunctionSep 28, 2026Filling Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923187 ↗
  • MalfunctionSep 28, 2026Filling Problem

    It was reported that a Spectrum IQ Infusion Pump had low flow rate (under infusion) during programming/setup. There was no patient involvement. No additional information is available.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920883 ↗
  • MalfunctionSep 25, 2026Filling Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50904641 ↗
  • MalfunctionSep 25, 2026Failure to Sense

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50904493 ↗
  • MalfunctionSep 25, 2026Failure to Sense

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50902548 ↗
  • MalfunctionSep 25, 2026Excess Flow or Over-Infusion; Insufficient Flow or Under Infusion

    It was reported that a Spectrum IQ Infusion Pump inaccurately delivered Parenteral Nutrition (PN), with an excessive amount remaining in the bag at the end of the infusion. The pump was reported to infuse PN (Parenteral Nutrition) 5-10% faster or slower than expected over extended periods during patient infusion. There was no report of patient injury or medical intervention associated with this event. No additional information is available.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50899800 ↗
  • MalfunctionSep 25, 2026Insufficient Flow or Under Infusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50899126 ↗