Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00887761001541), received between Nov 30, 2017 and Nov 7, 2025 — classified as 1 injury, 11 malfunctions.
The most frequently reported specific device problem is “Device Alarm System” (9 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Oct 14, 2021Class IITerminated
An unrecoverable error may occur where a component will fail to complete the reboot cycle, become non-functional and must be replaced.
Z-0402-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Alarm System9
- Activation Problem1
- Computer Software Problem1
- No Audible Alarm1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Bone Fracture(s)1
- Fall1
- No Information1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 7, 2025Device Alarm System; Device Alarm System
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE CODE CALL WAS NOT ALERTING . THERE WAS NO PATIENT/USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23498684 ↗MalfunctionMar 14, 2025Device Alarm System; Device Alarm System
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE CODE BLUE WAS NOT ANNUNCIATING. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21612105 ↗MalfunctionJul 24, 2024Device Alarm System; Device Alarm System
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE CODE BLUE FAILED. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19822649 ↗MalfunctionJun 1, 2024Device Alarm System; Device Alarm System
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE BED EXIT WAS NOT ANNUNCIATING. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19441809 ↗MalfunctionDec 22, 2023Device Alarm System; Device Alarm System
HILLROM RECEIVED A REPORT FROM A HILLROM TECHNICIAN STATING THE CODE CALLS WERE NOT ANNUNCIATING IN THE CORRECT LOCATION. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18387294 ↗MalfunctionSep 27, 2023Device Alarm System; Device Alarm System
THE CUSTOMER REPORTED THE BED ALARMS WERE NOT ROUTING TO THE NURSE'S STATION. A PATIENT FALL OCCURRED BUT THERE WAS NO INJURY TO THE PATIENT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17826431 ↗MalfunctionJul 20, 2023Activation Problem; Activation Problem
THE CUSTOMER REPORTED THE CODE BLUE CORD ON THE NAVICARE NURSE CALL SYSTEM WAS PULLED BUT THE CODE CALL DID NOT ANNUNCIATE. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17364634 ↗MalfunctionMay 15, 2023Device Alarm System; Device Alarm System
IT WAS REPORTED THAT THE CODE BLUE CALL FOR 4 SOUTH ROOMS 425 AND 424 DID NOT ROUTE, VIA THE NURSE CALL SYSTEM, TO THE SWITCHBOARD OPERATOR LOCATION. THE CUSTOMER CONFIRMED THAT THERE WAS NO PATIENT IMPACT OR INJURY ASSOCIATED WITH THIS EVENT, AND THE ALERT ROUTED AND ANNUNCIATED PROPERLY IN THE ASSOCIATED ROOMS AND ON THE ASSOCIATED UNIT. IT IS NOTED THAT THE ASSOCIATED ROOMS ARE "NEW INSTALLATION THAT OPENED RECENTLY." THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF #(B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16929425 ↗MalfunctionApr 5, 2023Device Alarm System; Device Alarm System
IT WAS REPORTED BY THE CUSTOMER THAT A CODE BLUE DOES NOTHING, DOES NOT SOUND, LIGHTS, SHOWS IN THE NURSE CALL GRAPHIC ROOM STATION WITH 10¿ SCREEN GRS 10. THERE WAS NO PATIENT IMPACT REPORTED AS A RESULT OF PRODUCT PERFORMANCE. THIS COMPLAINT WAS CAPTURED UNDER HILLROM REF # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16682011 ↗InjuryMar 31, 2023Device Alarm System; Device Alarm System
THE CUSTOMER CONTACTED HILLROM TO REPORT A FALL EVENT AND REQUESTED A NURSE CALL REPORT. THE CUSTOMER REPORTED THAT A PATIENT FELL THREE WEEKS PRIOR TO NOTIFYING HILLROM, AND THAT AT THE TIME OF THE EVENT, THE BED EXIT CALL DID NOT THE ANNUNCIATE AT THE CONSOLE OR AT THE DOME LIGHT CAUSING A DELAY IN RESPONSE. THE PATIENT SUSTAINED AN UNSPECIFIED FRACTURE WHICH REQUIRED SURGICAL INTERVENTION. IT IS NOTED THAT THE PATIENT HAS RECOVERED AND IS CURRENTLY IN ONGOING REHAB. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).
Patient problems: Bone Fracture(s); Fall
MAUDE report 16655232 ↗MalfunctionFeb 27, 2023No Audible Alarm; No Audible Alarm
IT WAS REPORTED THAT CODE BLUE CALLS WERE NOT ANNUNCIATING AT THE PBX OPERATOR FOR ROOMS ICU 01 AND ICU 03 ON THE 2ND FLOOR; HOWEVER, THE CALLS WERE ANNUNCIATING LOCALLY. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS REPORT. THIS EVENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16442118 ↗MalfunctionNov 30, 2017Computer Software Problem; Computer Software Problem
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE CODE BLUE LEVER WAS PULLED BUT THAT THE CALL WAS DELAYED IN RINGING AT THE PRIMARY NURSE STATION.
Patient problems: No Information
MAUDE report 7073275 ↗
Other System, communication, powered devices
- Voalte Nurse Call 4.0Hill-Rom, Inc. · System, communication, powered2 recalls2 deaths405 reports
- NaviCare® Nurse Call Systems 3.9Hill-Rom, Inc. · System, communication, powered2 recalls2 deaths87 reports
- NaviCare® Nurse Call Systems 3.7Hill-Rom, Inc. · System, communication, powered1 recall0 reports
- Tobii Dynavox TD I-110Tobii Dynavox LLC · System, Communication, Powered1 recall0 reports
- Voalte Nurse Call 4.1Hill-Rom, Inc. · System, communication, powered14 reports
- GEN4 DIRECT ACCESS 4C370U-P10424CCurbell Medical Products, Inc. · System, Communication, Powered3 reports
- GEN4 PLUS, RESPONDER 5, ENHANCED 5C410U-10134003ACurbell Medical Products, Inc. · System, Communication, Powered3 reports
- HomeSafe AutoAlert Pendant 300000219851Lifeline Systems Company · System, communication, powered2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page