These are communication systems, labeled as recallZ-0402-2022 · Hill-Rom Inc. · initiated Oct 14, 2021

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hill-Rom Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An unrecoverable error may occur where a component will fail to complete the reboot cycle, become non-functional and must be replaced.

FDA root cause category: Under Investigation by firm

What the company is doing

The firm disseminated a Medical Device Correction Notice to its consignees on 10/21/2021 by mail. Customer: Please share the information with potential users in your organization. Complete the attached response form and return to [email protected] within one month of receipt of this notice to confirm you have received this communication and understand the potential issue. In the event, this issue should occur please contact Hillrom Technical Support at 1-800-445-3720, option 3 or [email protected] for service and replacement. Hillrom has remediated this issue as part of the next software upgrade to version 4.0.300. Please contact Hillrom to discuss software upgrade. Distributor: Please share this information with your nurse call customer base. Complete the attached response form and return to [email protected] within one month of receipt of this notice to confirm you have received this communication, understand the potential issue and will take action to inform your customer base.

Recall details

Recall number
Z-0402-2022
Status
Terminated
Initiated by the company
Oct 14, 2021
Recalling firm
Hill-Rom Inc.
Quantity
1,050 units
Product code
ILQ
Distribution
Worldwide distribution - US Nationwide.

Product description

These are communication systems, labeled as: a. NaviCare Nurse Call, Part Numbers: 3.6, 3.7, 3.9 b. Voalte Nurse Call, Part Number 4.0 *Product was rebranded from NaviCare to Voalte in 2020

Code information (models, lots, serial numbers)

All software versions 3.5.400 and above: a. Part Number 3.6, UDI 00887761001541; Part Number 3.7, UDI 00887761021198; Part Number 3.9, UDI 00887761985209 b. Part Number 4.0, UDI 00887761995079

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.