Progressa+ Bed P7501ARecalls & reported problems
Hill-Rom, Inc. · Hospital Bed · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Hill-Rom Progressa+ hospital bed has 36 adverse event reports, with 2 injuries and no deaths. Most reports describe physical damage to the bed, such as torn or split materials, along with brakes that do not hold and the bed moving on its own, plus a power cord with exposed copper wiring. There is 1 open recall, started in October 2023, over static electricity that can build up on the casters and power drive and pass a mild shock to whoever pushes the bed.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 33
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 00887761966925), received between Aug 23, 2024 and Aug 12, 2026 — classified as 2 injuries, 31 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (24 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Oct 4, 2023Class IIOpen, Classified
Progressa+ beds have the potential for static buildup from the casters and/or power drive. This static electricity could be passed to the operator through the transport handle during transport, resulting in a mild static electric shock.
Z-0212-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Split, Cut or Torn24
- Unintended Movement3
- Activation Problem1
- Collapse1
- Disconnection1
- Flare or Flash1
- Improper Flow or Infusion1
- No Audible Alarm1
Patient problems
- No Clinical Signs, Symptoms or Conditions31
- Unspecified Tissue Injury2
Recent adverse event reports
The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 12, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50225519 ↗MalfunctionJul 27, 2026Unintended Movement
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50037659 ↗MalfunctionJul 17, 2026Unintended Movement
BAXTER RECEIVED A REPORT THAT PROGRESSA PLUS HAD BRAKES NOT HOLDING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25861711 ↗MalfunctionApr 17, 2026Material Split, Cut or Torn
BAXTER RECEIVED A REPORT THAT PROGRESSA PLUS HAD THE AUXILIARY POWER CORD DAMAGED WITH COPPER WIRE EXPOSED. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24912312 ↗MalfunctionMar 11, 2026Material Split, Cut or Torn
ON 11-FEB-2026, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT PROGRESSA PLUS (PRODUCT CODE P7501A000032, SERIAL NUMBER (B)(6)), HAD POWER CORD DAMAGED WITH COPPER WIRE EXPOSED. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24573415 ↗MalfunctionMar 6, 2026Collapse
ON 09-FEB-2026, A CUSTOMER CONTACTED TECHNICAL SERVICE TO REPORT THAT PROGRESSA PLUS (PRODUCT CODE P7501A000095, SERIAL NUMBER (B)(6)), HAD SIDERAIL WAS BROKEN AND FALLING OFF. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24532853 ↗MalfunctionFeb 24, 2026Material Split, Cut or Torn
ON 29-JAN-2026, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT PROGRESSA PLUS (PRODUCT CODE P7501A000032, SERIAL NUMBER (B)(6)), HAD AUXILIARY POWER CORD DAMAGED WITH COPPER WIRES EXPOSED. THIS OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24431730 ↗InjuryFeb 24, 2026Improper Flow or Infusion
IT WAS REPORTED PROGRESSA PLUS BED HAD AN ACB (AIR CONTROL BOARD) ERROR FAULT CAUSING THE MATTRESS TO DEFLATE WHICH RESULTED IN THE PATIENT SUSTAINING A DEEP TISSUE PRESSURE INJURY. THE PATIENT RECEIVED UNSPECIFIED ¿WOUND CARE TREATMENT.¿ MULTIPLE ATTEMPTS WERE MADE TO THE CUSTOMER TO OBTAIN SPECIFIC DETAILS OF THE REPORTED DEEP TISSUE PRESSURE INJURY, INCLUDING LOCATION, CHARACTERISTICS, AND WHAT MEDICAL INTERVENTION WAS PERFORMED, HOWEVER NO RESPONSE WAS RECEIVED. AN INSPECTION PERFORMED BY A BAXTER TECHNICIAN NOTED THE SUPPLY HOSE HAD DISCONNECTED. THE TECHNICIAN RECONNECTED THE SUPPLY HOSE AND NOTED THE BED WAS FUNCTIONING AS DESIGNED. NO FURTHER INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Unspecified Tissue Injury
MAUDE report 24431622 ↗MalfunctionFeb 11, 2026No Audible Alarm
ON 14-JAN-2026, A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT PROGRESSA PLUS (PRODUCT CODE P7501A000032, SERIAL NUMBER (B)(6)), HAD BED EXIT NOT SENDING SIGNAL TO NURSE CALL. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24314597 ↗MalfunctionFeb 3, 2026Material Split, Cut or Torn
ON (B)(6) 2026, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT PROGRESSA PLUS (PRODUCT CODE P7501A000032, SERIAL NUMBER (B)(6) , HAD AUXILIARY POWER CORD DAMAGED WITH COPPER WIRES EXPOSED. THIS OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24248467 ↗MalfunctionJan 21, 2026Material Split, Cut or Torn
ON 06-JAN-2026, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT PROGRESSA PLUS (PRODUCT CODE P7501A000033, SERIAL NUMBER (B)(6)), HAD AUXILIARY POWER CORD DAMAGED WITH EXPOSED COPPER WIRES. THIS OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24134074 ↗MalfunctionJan 15, 2026Flare or Flash
ON 30-DEC-2025, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT PROGRESSA PLUS (PRODUCT CODE P7501A000191, SERIAL NUMBER (B)(6)), A FIRE WAS OBSERVED INVOLVING THE POWER CORD. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24083772 ↗MalfunctionJan 7, 2026Material Split, Cut or Torn
ON 31-DEC-2025, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT PROGRESSA PLUS (PRODUCT CODE P7501A000032, SERIAL NUMBER (B)(6)), HAD AUXILIARY POWER CORD DAMAGED WITH COPPER WIRES EXPOSED. THIS OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23998115 ↗MalfunctionJan 7, 2026Material Split, Cut or Torn
ON (B)(6) 2025, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT PROGRESSA PLUS (PRODUCT CODE P7501A000032, SERIAL NUMBER (B)(6)), HAD AUXILIARY POWER CORD DAMAGED WITH COPPER WIRES EXPOSED. THIS OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23995594 ↗MalfunctionNov 6, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE PROGRESSA PLUS AUXILIARY POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED AFTER PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. IT WAS UNKNOWN IF ANY SECOND DEVICE WAS ALSO INVOLVED. THE CUSTOMER DID NOT KNOW IF THIS EVENT WAS ALREADY COMMUNICATED TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23486992 ↗MalfunctionNov 4, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE PROGRESSA PLUS AUXILIARY POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED AFTER PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. IT WAS UNKNOWN IF ANY SECOND DEVICE WAS ALSO INVOLVED. THE CUSTOMER DID NOT KNOW IF THIS EVENT WAS ALREADY COMMUNICATED TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23466351 ↗MalfunctionNov 3, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE PROGRESSA PLUS AUXILIARY POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT KNOW IF THIS EVENT WAS ALREADY COMMUNICATED TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23458760 ↗MalfunctionNov 3, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
ON 10-OCT-2025, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT PROGRESSA PLUS (PRODUCT CODE P7501A001971, (B)(6), HAD POWER CORD DAMAGED WITH EXPOSED COPPER WIRES. THIS OCCURRED AFTER PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23458245 ↗MalfunctionOct 30, 2025Activation Problem; Activation Problem
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE CPR FUNCTION WAS INOPERATIVE . THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23426012 ↗MalfunctionOct 15, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE AUX CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23286593 ↗MalfunctionAug 27, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE PROGRESSA PLUS AUXILIARY POWER CORD HAD COPPER WIRES EXPOSED. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22907828 ↗MalfunctionAug 21, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT STATING THAT PROGRESSA PLUS AUXILIARY POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT AND OCCURRED AFTER PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22861712 ↗MalfunctionAug 14, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN STATING THE POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22801607 ↗MalfunctionJul 21, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT PROGRESSA PLUS (PRODUCT CODE P7501A000032, SERIAL NUMBER (B)(6)), HAD AUXILIARY POWER CORD DAMAGED WITH COPPER WIRE EXPOSED. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22555242 ↗MalfunctionJun 10, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
BAXTER RECEIVED A REPORT STATING THAT PROGRESSA PLUS AUXILIARY POWER CORD HAD DAMAGE WITH EXPOSED COPPER WIRES. THE BED WAS LOCATED AT THE ACCOUNT AND OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22177932 ↗
FDA clearance and classification
The FDA regulates product code FNL (bed, ac-powered adjustable hospital) under 21 CFR 880.5100 as a Class II device. 21 CFR 880.5100 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Progressa+ Bed P7501A
Has the Progressa+ Bed P7501A been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0212-2024 (Class II), began on Oct 4, 2023: Progressa+ beds have the potential for static buildup from the casters and/or power drive. This static electricity could be passed to the operator through the transport handle during transport, resulting in a mild… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Progressa+ Bed P7501A?
The FDA's MAUDE database lists 33 adverse event reports for this device, received between Aug 23, 2024 and Aug 12, 2026. In those reports, the most common specific device problems are “Material Split, Cut or Torn” (24), “Unintended Movement” (3) and “Activation Problem” (1); the most common patient problem is “Unspecified Tissue Injury” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Progressa+ Bed P7501A?
None of the reports are classified as deaths, 2 as injuries and 31 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Progressa+ Bed P7501A FDA approved?
Devices of this type (product code FNL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Hill-Rom, Inc.
Is the Progressa+ Bed P7501A MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Progressa+ Bed P7501A, and is it still sold?
The FDA's device database lists Hill-Rom, Inc. as the company responsible for it (brand Progressa+ Bed, model P7501A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BED, AC-POWERED ADJUSTABLE HOSPITAL devices
- Centrella™ Bed P7900BHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL2 recalls5 deaths1,831 reports
- Progressa® Bed System P7500AHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death1,320 reports
- VersaCare® Med Surg Bed P3200KHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL3 deaths1,149 reports
- Medline MDR107003EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall3 deaths333 reports
- Medline MDR107002EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death330 reports
- InTouch 2141Stryker Corporation · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths361 reports
- Citadel Plus FX811B3B4AMABBArjohuntleigh AB · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths354 reports
- TotalCare® Bed P1900QHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL299 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page