Skip to content

iLED7 Surgical Light 4068210Recalls & reported problems

Baxter Medical Systems GmbH + Co. KG · iLED 7 Ceiling Duo · Class 2

All models of iLED7 Surgical Light (6)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00887761968295), received between Dec 9, 2024 and Feb 18, 2026 — classified as 3 injuries.

The most frequently reported specific device problem is “Excessive Heating” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 4, 2024Class IIOpen, Classified

    Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction due to customer reports of thermal skin injuries when using the iLED 7 surgical light system products listed below. Specific VidiaPort and Trulight product codes are also impacted as they are equipped with iLED 7 light-heads

    Z-2315-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Excessive Heating1
  • Sharp Edges1
  • Use of Device Problem1

Patient problems

  • Burn(s)1
  • Laceration(s)1
  • Partial thickness (Second Degree) Burn1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 18, 2026Use of Device Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED A PATIENT SUSTAINED A BURN WITH AN ILED7 SURGICAL LIGHT. THE PATIENT WAS UNDERGOING A THREE-HOUR THYROID SURGICAL PROCEDURE WHEN A 1.5X1CM PINK AREA ¿JUST BELOW THE SURGICAL AREA¿ WAS OBSERVED TO HAVE DEVELOPED. THE AREA EVOLVED DURING THE CONTINUED PROCEDURE, INTO AN AREA DESCRIBED AS ¿DOUBLING IN SIZE¿ AND ¿OOZING.¿ THE WOUND WAS LATER DESCRIBED A ¿3X2CM BURN THAT WAS PINK AND OOZING.¿ THE ¿PINK AREA¿ HAD AN ABSENT OUTER EPIDERMAL LAYER INDICATIVE OF A SECOND-DEGREE/PARTIAL THICKNESS BURN. NO SPECIFIC MEDICAL TREATMENT WAS REPORTED; HOWEVER, IT WAS REPORTED THAT THE BURN INJURY REQUIRED REPOSITIONING AND FOLLOW-UP WITH A HEALTHCARE PROVIDER. THE CUSTOMER ATTRIBUTES THE INJURY TO THE USE OF TWO LAMPS AT 80% BRIGHTNESS NOT BEING SEPARATED, BUT FIXED (OVERLAPPING). THE LIGHTS WERE NOT FOCUSED ON THE PATIENT BUT ON THE SURGICAL FIELD. ADDITIONAL INFORMATION RECEIVED NOTED THAT MULTIPLE NON-BAXTER HEADLIGHTS WERE USED IN CONJUNCTION WITH THE ILED LIGHTS. NO FURTHER INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 24381735 ↗
  • InjuryAug 26, 2025Excessive Heating; Excessive Heating

    IT WAS REPORTED AN ILED 7 CEILING DUO BURNED A PATIENT. DURING AN UNSPECIFIED SURGICAL PROCEDURE, THE SURGICAL TEAM FOUND THE PATIENT¿S SKIN WAS ¿LIKE PEELING AT BACK OF THE PATIENT¿ AND TEMPERATURE WHERE THE LIGHT WAS FOCUSED ON THE PATIENT¿S BACK WAS HOTTER COMPARED TO THE OTHER PARTS OF THE PATIENT¿S BODY. THE PATIENT WAS TREATED WITH COLD FLUIDS APPLIED TO THE AFFECTED AREA. THE BIOMEDICAL DEPARTMENT DID AN INVESTIGATION BY FOCUSING THE LIGHT ON ONE SPOT AND FOUND THAT THE AREA WHERE IT WAS FOCUSED ON WAS HOT AND REACHED A TEMPERATURE OF 43 DEGREES CELSIUS. THE LIGHT DEFAULT NORMAL SETTINGS ARE 50%; HOWEVER, THE OPERATING ROOM TEAM CAN INCREASE THEM DURING THE PROCEDURE IF NEEDED. THE BAXTER TECHNICIAN CHECKED THE SOFTWARE OF THE LIGHT AND FOUND THAT ALL ARE UP TO DATE. MULTIPLE ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION WERE UNSUCCESSFUL. NO FURTHER INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Burn(s)

    MAUDE report 22895897 ↗
  • InjuryDec 9, 2024Sharp Edges; Sharp Edges

    IT WAS REPORTED THAT AN EMPLOYEE WAS INJURED WHILE TRYING TO MOVE THE ILED 7 SPRING ARM. THE SPRING ARM ALLEGEDLY HAD MISSING AND LOOSE PARTS. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: Laceration(s)

    MAUDE report 20876973 ↗

FDA clearance and classification

The FDA regulates product code FSQ (light, surgical, instrument) under 21 CFR 878.4580 as a Class II device. 21 CFR 878.4580 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the iLED7 Surgical Light 4068210

Has the iLED7 Surgical Light 4068210 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2315-2024 (Class II), began on Jun 4, 2024: Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction due to customer reports of thermal skin injuries when using the iLED 7 surgical light system products listed below. Specific VidiaPort and… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the iLED7 Surgical Light 4068210?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Dec 9, 2024 and Feb 18, 2026. In those reports, the most common specific device problems are “Excessive Heating” (1), “Sharp Edges” (1) and “Use of Device Problem” (1); the most common patient problems are “Burn(s)” (1), “Laceration(s)” (1) and “Partial thickness (Second Degree) Burn” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the iLED7 Surgical Light 4068210?

None of the reports are classified as deaths, 3 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the iLED7 Surgical Light 4068210 FDA approved?

Devices of this type (product code FSQ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Baxter Medical Systems GmbH + Co. KG.

Who makes the iLED7 Surgical Light 4068210, and is it still sold?

The FDA's device database lists Baxter Medical Systems GmbH + Co. KG as the company responsible for it (brand iLED7 Surgical Light, model 4068210). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Light, surgical, instrument devices

All Light, surgical, instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page