Skip to content

TruLight 4038210Recalls & reported problems

Baxter Medical Systems GmbH + Co. KG · Hillrom™ TruLight 5000 / 3000 Ceiling Duo · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00887761995918), received between Mar 9, 2022 and Feb 13, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component5

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Bruise/Contusion1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 13, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    CUSTOMER REPORTED DAMAGE TO THE SKY LAMP DURING CLEANING. THE LAMP DETACHED FROM THE JOINT AND CAUSED DAMAGE TO THE CEILING AND FLOOR. NO HARM OR INJURY TO PATIENT OR CAREGIVER WAS REPORTED THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT#: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21376787 ↗
  • MalfunctionSep 18, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    CUSTOMER REPORTED THAT THEY SAW A SCREW OF THE MONITOR BRACKET LOOSE. WHEN THEY MOVED THE BRACKET, THE MONITOR BRACKET BROKE, THE MONITOR CAME LOOSE AND BURST THE FIBER OPTIC CABLES. NO HARM OR INJURY TO PATIENT OR CAREGIVER WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20245805 ↗
  • MalfunctionFeb 2, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    CUSTOMER REPORTED THE LIGHT HEAD FELL TO THE FLOOR AND THE SPRING ARM HIT THE CEILING AFTER ONE STAFF MEMBER PULLED THE CLIP OUT OF THE CONNECTION. NO HARM OR INJURY TO PATIENT OR CAREGIVER WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18629136 ↗
  • MalfunctionJul 24, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED THAT THE SPRING ARM FROM THE LIGHT SYSTEM DETACHED FROM THE CENTRAL AXIS. IT DID NOT FALL DOWN. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17377023 ↗
  • MalfunctionMar 9, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A NURSE MOVED THE SPRING ARM WITH RADIATION PROTECTION SHIELD WHICH WAS ATTACHED TO THE TRULIGHT SURGICAL LIGHT AS THE SPRING ARM SEPARATED FROM THE SUPPORT ARM SYSTEM AND FELL DOWN. THE FALLING COMPONENT HIT THE NURSE AT THE SHOULDER. NO SERIOUS INJURY OCCURRED IN RELATION TO THE EVENT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: Bruise/Contusion

    MAUDE report 13715002 ↗

FDA clearance and classification

The FDA regulates product code FSQ (light, surgical, instrument) under 21 CFR 878.4580 as a Class II device. 21 CFR 878.4580 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the TruLight 4038210

Has the TruLight 4038210 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00887761995918), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the TruLight 4038210?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Mar 9, 2022 and Feb 13, 2025. In those reports, the most common specific device problem is “Detachment of Device or Device Component” (5); the most common patient problem is “Bruise/Contusion” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the TruLight 4038210?

None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the TruLight 4038210 FDA approved?

Devices of this type (product code FSQ) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Baxter Medical Systems GmbH + Co. KG.

Who makes the TruLight 4038210, and is it still sold?

The FDA's device database lists Baxter Medical Systems GmbH + Co. KG as the company responsible for it (brand TruLight, model 4038210). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Light, surgical, instrument devices

All Light, surgical, instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page