Skip to content

Head Sections 1850955Recalls & reported problems

Baxter Medical Systems GmbH + Co. KG · Head section double joint H V · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887761970380), received between Jan 7, 2025 and Jan 7, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Unintended Movement” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Movement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 7, 2025Unintended Movement; Unintended Movement

    THE CUSTOMER REPORTED THAT THE ARTICULATED HEAD SECTION OF THE OPERATING TABLE BROKE, CAUSING THE PATIENT TO FALL DURING A PROCEDURE. NO INJURY WAS REPORTED IN RELATION TO THIS EVENT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21091075 ↗

Other TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED devices

All TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page