Skip to content

Trumpf Medical 4080100Recalls & reported problems

Baxter Medical Systems GmbH + Co. KG · PST 500 · Class 2

All models of TRUMPF MEDICAL (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00887761979390), received between Jun 23, 2021 and Aug 14, 2026 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Unintended Movement” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Movement4
  • Unstable4
  • Device Slipped1
  • Operating System Becomes Nonfunctional1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 14, 2026Unintended Movement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50261249 ↗
  • MalfunctionApr 21, 2026Unstable

    DURING COMPLAINT EVALUATION WAS FOUND THAT TABLE PST 500 HAD WOBBLING MOVEMENT WHEN THE TABLETOP WAS PUSHED. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24940021 ↗
  • MalfunctionDec 22, 2025Unintended Movement; Unintended Movement

    ON 25-NOV-2025, A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT PST 500 (PRODUCT CODE 4080100, SERIAL NUMBER (B)(6)), HAD AN ISSUE, THE BRAKES DO NOT EXTEND EVENLY, CAUSING THE OPERATING TABLE TO WOBBLE. ONE BRAKE RELEASES INTRAOPERATIVELY, CAUSING THE OPERATING TABLE TO WOBBLE. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT INJURY OR DEATH REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23872421 ↗
  • MalfunctionNov 21, 2025Unintended Movement; Unintended Movement

    ANOTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE PST 500 HAD AN ISSUE, THE HYDRAULIC STAMPS WERE DISENGAGING ON ITS OWN. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23619417 ↗
  • MalfunctionAug 9, 2024Operating System Becomes Nonfunctional; Operating System Becomes Nonfunctional

    CUSTOMER REPORTED THAT TABLE HAS NO FUNCTIONS. DISPLAYING ERROR 62. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19951035 ↗
  • MalfunctionFeb 20, 2024Unstable; Unstable

    CUSTOMER REPORTED THAT AFTER SOME TIME HYDRAULIC STAMPS DISENGAGE. IT WAS STATED THAT USUALLY ONLY ONE STAMP LOOSE PRESSURE AND TABLE WILL LOOSE LEVEL AND START MOVING ON THAT ONE STAMP. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED.THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18743751 ↗
  • MalfunctionMar 27, 2023Device Slipped; Device Slipped

    CUSTOMER REPORTED THAT HYDRAULIC UNIT AFTER SOME TIME DISENGAGES (LOSS OF PRESSURE) WITHOUT TABLE RECOGNIZING ITS UNLOCKED. STAFF THEN HAVE TO UNLOCK THE TABLE SO SOFTWARE WILL KNOW ITS UNLOCKED (WITHOUT ANY EFFECT BECAUSE BRAKES ARE ALREADY UNLOCKED) AND THEN LOCK IT AGAIN. ADDITIONALLY IT WAS CONFIRMED THAT ALL FOUR STAMPS RELEASED DURING PROCEDURES AND THEN THE TABLE MOVES WHEN SOMEONE IS LEANING ON IT. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16618636 ↗
  • MalfunctionSep 8, 2021Unstable; Unstable

    THE CUSTOMER REPORTED THE OPERATING TABLE WAS ON THE BRAKES WHEN STARTING THE SURGICAL PROCEDURE. AFTER TEN MINUTES TWO BRAKES RELEASED AND THE DEVICE WAS ROCKING. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12437992 ↗
  • MalfunctionJun 23, 2021Unstable; Unintended Movement; Unstable; Unintended Movement

    IT WAS REPORTED, DURING A VERY PRECISE SURGICAL PROCEDURE THE OPERATING TABLE BECAME UNSTABLE (TILTING). NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED.THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12048378 ↗

Other Table, radiologic devices

All Table, radiologic devices →

Sources and updates