Likorall™ overhead lift 3122010Recalls & reported problems
Liko AB · Overhead track patient lifting/transfer system · Class 1
All models of Likorall overhead lift (25)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00887761979819), received between Feb 7, 2023 and Jun 16, 2026 — classified as 1 injury, 12 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Sep 14, 2026Class IIOpen, Classified
Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use. Patients are at some risk for injury if a significant fall/drop were to occur due to this issue.
Z-0151-2027 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Insufficient Device Problem Information4
- Break3
- Detachment of Device or Device Component3
- Fracture2
- Adverse Event Without Identified Device or Use Problem1
- Device Fell1
- Material Split, Cut or Torn1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Fall3
- Head Injury1
- Hematoma1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 16, 2026Fracture
BAXTER RECEIVED A REPORT THAT LIKORALL 242 S, WHITE HAD MOTOR HOOK BROKEN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25522550 ↗MalfunctionDec 18, 2025Insufficient Device Problem Information; Device Fell; Insufficient Device Problem Information; Device Fell
A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE LIKORALL 242 S, WHITE HAD AN ISSUE, AN ACCIDENT OCCURRED DUE TO ENGINE FAILURE. THE PERSON HAD A FALL. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23839561 ↗MalfunctionOct 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE LIKORALL 242 S, WHITE HAD AN ISSUE, IT WAS REPORTED THE SLING BAR HOOK DETACHED FROM THE ADAPTER 22 DURING A TRANSFER LEADING TO A PATIENT FALL. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23227147 ↗MalfunctionSep 29, 2025Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information
A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT DURING THE TRANSFER OF A RESIDENT, THE LOAD-BEARING STRAP FAILED. LUCKILY, THEY WERE ABOVE THE BED AT THE TIME OF THE INCIDENT, WHICH PREVENTED ANY INJURIES. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY. THE DEVICE WAS REQUESTED FOR SERVICE/INSPECTION BY BAXTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23165962 ↗MalfunctionSep 15, 2025Mechanical Problem; Insufficient Device Problem Information; Mechanical Problem; Insufficient Device Problem Information
THE CUSTOMER REPORTED THAT A LIKORALL 242 LIFT WAS INVOLVED IN A PATIENT FALL. THE SEQUENCE OF EVENTS SURROUNDING THE FALL WAS NOT PROVIDED. HOWEVER, THE CUSTOMER REPORTED THAT THE PATIENT SUSTAINED A ¿BUMP ON THE HEAD¿ AND ¿FELL ON HIS BUTTOCKS.¿ MEDICAL INTERVENTION WAS NOT REPORTED.
Patient problems: Fall; Hematoma
MAUDE report 23051156 ↗MalfunctionSep 11, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
DURING SERVICING BY BAXTER, TECHNICAL SERVICE IDENTIFIED THAT THE LIKORALL 242 S, CHARGER WAS DAMAGED WITH EXPOSED COPPER.THIS OCCURRED DURING BAXTER SERVICING/TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY. THE DEVICE WAS REQUESTED FOR SERVICE/INSPECTION BY BAXTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23027647 ↗MalfunctionAug 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT LIKORALL 242 S, WHITE, HAD AN ISSUE, SUSPICION OF THE SLINGBAR HARNESS COMING OUT OF THIS LIKORALL 242S ENGINE. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY. THE DEVICE WAS REQUESTED FOR SERVICE/INSPECTION BY BAXTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22786994 ↗MalfunctionMay 14, 2025Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE LIKORALL 242 S, WHITE HAD AN ISSUE, WHEN INSPECTING A LIKORAL 242'S LIFT, THE LIFTSTRAP WAS CHANGED AND WHEN TESTING THE MANUAL EMERGENCY LOWERING, THE LIFT STRAP SLID UNCONTROLLABLY STRAIGHT THROUGH AT FIRST ¿HAND PUMPING¿. USUALLY YOU GET TO ¿PUMP¿ DOWN TYPE 6-7 CM AT A TIME AT 60 KG SAMPLE WEIGHT. THIS OCCURRED DURING BIOMED TESTING. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22016856 ↗InjuryMar 14, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED A PATIENT FELL WHEN BEING TRANSFERRED WITH A LIKORALL OVERHEAD LIFT. THE PATIENT WENT TO THE HOSPITAL AND RECEIVED 30 SUTURES TO THE RIGHT SIDE OF THEIR HEAD. NO FURTHER INFORMATION WAS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Fall; Head Injury
MAUDE report 21603712 ↗MalfunctionJan 9, 2025Break; Break
IT WAS REPORTED THAT WHEN A PATIENT WAS BEING MOVED FROM THE BED WITH A LIKORALL 242 LIFT, THE ¿SUSPENSION EYELET ON THE LIFTING JUMPER BROKE¿ CAUSING THE PATIENT TO FALL. THE CUSTOMER REPORTED ¿LIGHTER PHYSICAL INJURIES¿ BECAUSE THE PATIENT LANDED ON THE BED, BUT THE FALL CAUSED FEARS. THE LOCAL BAXTER TEAM IN SWEDEN OBTAINED THE FOLLOWING ADDITIONAL INFORMATION. THE FEMALE PATIENT WAS REPORTED TO BE BETWEEN 65-85 YEARS OLD AND WEIGHED 46-90 KG. AT THE TIME OF THE REPORTED INCIDENT, THE LIKORALL 242 WAS BEING USED WITH A UNIVERSAL SLINGBAR 450 (PROD. NO. 3156085) AND A GULDMANN BASIC BASIC L SLING. THE COMPOSITE HOOK ON THE UNIVERSAL SLINGBAR BROKE WHEN THE PATIENT WAS RAISED 20 CM ABOVE THE BED AND THE PATIENT¿S FOREHEAD HIT THE SIDERAIL OF THE BED. THE NURSE ASSESSED THE PATIENT¿S VITAL SIGNS AND SKIN STATUS, AND NO TREATMENT WAS REQUIRED.
Patient problems: Fall
MAUDE report 21110061 ↗MalfunctionNov 20, 2024Break; Break
A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE LIKORALL 242 S, COMPOSITE HOOK BROKE DURING A PATIENT TRANSFER. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20736550 ↗MalfunctionOct 2, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER ALLEGED THE LIFT¿S MOTOR WAS NONFUNCTIONAL AFTER A BATTERY CHANGE THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20350886 ↗MalfunctionFeb 7, 2023Fracture; Fracture
IT WAS REPORTED THAT ONE OF THE SLINGBAR'S HOOKS BROKE DURING A PATIENT TRANSFER. THERE WAS NO PATIENT/CAREGIVER INJURY REPORTED AS THE PATIENT FELL BACK ONTO THE BED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16311823 ↗
FDA clearance and classification
The FDA regulates product code FSA (lift, patient, non-ac-powered) under 21 CFR 880.5510 as a Class I device. 21 CFR 880.5510 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Likorall™ overhead lift 3122010
Has the Likorall™ overhead lift 3122010 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0151-2027 (Class II), began on Sep 14, 2026: Potential issue involving motor shaft breakage combined with reduced performance of the Single Fault Safety (SFS) system may lead to instances where the lift strap lowers more quickly than expected during use.… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Likorall™ overhead lift 3122010?
The FDA's MAUDE database lists 13 adverse event reports for this device, received between Feb 7, 2023 and Jun 16, 2026. In those reports, the most common specific device problems are “Break” (3), “Detachment of Device or Device Component” (3) and “Fracture” (2); the most common patient problems are “Fall” (3), “Head Injury” (1) and “Hematoma” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Likorall™ overhead lift 3122010?
None of the reports are classified as deaths, 1 as an injury and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Likorall™ overhead lift 3122010 FDA approved?
Devices of this type (product code FSA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Liko AB.
Who makes the Likorall™ overhead lift 3122010, and is it still sold?
The FDA's device database lists Liko AB as the company responsible for it (brand Likorall™ overhead lift, model 3122010). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Lift, patient, non-ac-powered devices
- Likorall™ overhead lift 3123001Liko AB · Lift, patient, non-ac-powered1 recall1 death32 reports
- Likorall™ overhead lift 3121001Liko AB · Lift, patient, non-ac-powered1 recall22 reports
- Likorall™ overhead lift 3124050Liko AB · Lift, patient, non-ac-powered1 recall17 reports
- Viking™ XL mobile lift 2040043Liko AB · Lift, patient, non-ac-powered1 recall14 reports
- Medline MDS400ELBMedline Industries, Inc. · Lift, patient, non-ac-powered59 reports
- Uno™ mobile lift 2010004Liko AB · Lift, patient, non-ac-powered1 recall6 reports
- Likorall™ overhead lift 3122502Liko AB · Lift, patient, non-ac-powered1 recall5 reports
- Medline MDS600ELBMedline Industries, Inc. · Lift, patient, non-ac-powered55 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page