Skip to content

Uno™ mobile lift 2010004Recalls & reported problems

Liko AB · Mobile patient lifting system, battery-powered · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00887761981775), received between Nov 5, 2024 and Feb 24, 2026 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 30, 2025Class IOpen, Classified

    The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

    Z-2031-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Device Fell1
  • Device Tipped Over1
  • Disconnection1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Pain2
  • Laceration(s)1
  • Vertebral Fracture1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2026Device Fell

    ON 29-JAN-2026, A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT UNO 102 ES (PRODUCT CODE P2010005, SERIAL NUMBER (B)(6)), HAD IMPROPERLY POSITIONED HOOK, RESULTING IN A PATIENT FALL. THE FALL WAS ESTIMATED TO BE A FEW CENTIMETERS ABOVE THE BED. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24431376 ↗
  • InjuryDec 31, 2025Disconnection; Disconnection

    DURING THE TRANSFER OF A PATIENT TO A BED USING AN UNO MOBILE PATIENT LIFTING SYSTEM, THE HOOK (CONNECTING THE SLINGBAR TO THE LIFT ARM) CAME OFF. THE HOOK HIT THE PATIENT ON THE FOREARM AS IT DROPPED AND CAUSED THE PATIENT TO FALL A SHORT DISTANCE ON TO THE MATTRESS OF THE BED. THE PATIENT WAS NOT INJURED FROM THE FALL, HOWEVER, SUSTAINED A 15CM LACERATION TO THE FOREARM FROM THE HOOK. THE INJURY WAS CLEANED AND THEN CLOSED WITH SUTURES, AND A SLING WAS APPLIED. AN X-RAY CONFIRMED NO FRACTURE. NO ADDITIONAL INFORMATION WAS AVAILABLE.

    Patient problems: Laceration(s)

    MAUDE report 23934131 ↗
  • MalfunctionApr 3, 2025Device Tipped Over; Device Tipped Over

    A CUSTOMER CONTACTED TECHNICAL SERVICE TO REPORT THAT THE UNO 102 EE LIFT HAD AN ISSUE, CUSTOMER ALLEGES THAT THE LIFT IS TIPPING TO THE RIGHT. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED. THERE WERE NO OTHER DEVICES REPORTED TO BE INVOLVED. THE CUSTOMER DID NOT COMMUNICATE THIS EVENT TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY. THE DEVICE WAS REQUESTED FOR SERVICE/INSPECTION BY BAXTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21761069 ↗
  • MalfunctionMar 10, 2025Material Twisted/Bent; Material Twisted/Bent

    A OTHER HEALTH CARE PROFESSIONAL CONTACTED TECHNICAL SERVICE TO REPORT THAT THE UNO 102 EE HAD AN ISSUE, THE LEFT LEG WAS BENT OFF THE FLOOR. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21558903 ↗
  • MalfunctionJan 27, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT WITH DURING A TRANSFER FROM WHEELCHAIR TO BED USING AN UNO 102 LIFT AND NON-LIKO SLING (HUMAN CARE), THE SLING BAR DETACHED, AND THE PATIENT FELL FIRST ONTO THE BED, AND THEN ONTO THE FLOOR. THE PATIENT REPORTED PAIN AFTER THE FALL, AND A RED MARK WAS OBSERVED ON THEIR BACK. NO MEDICAL INTERVENTION WAS REQUIRED. THE STAFF INVOLVED IN THE EVENT DID NOT REPORT ANY SING OF DEVICE MALFUNCTION.

    Patient problems: Pain

    MAUDE report 21241289 ↗
  • InjuryNov 5, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT AS A PATIENT WAS BEING TRANSFERRED FROM HIS BED TO A WHEELCHAIR USING THE UNO 102 LIFT, THE LIFT¿S ¿JUMPER¿ (SLINGBAR) DETACHED AT THE QUICK RELEASE WHILE THE PATIENT WAS HANGING IN THE LIFT. THE PATIENT FELL DOWN TO THE EDGE OF THE BED AND THEN SLID DOWN THE SIDE OF THE BED TO THE FLOOR. AS THE PATIENT SLID DOWN, HE HIT HIS HEAD ON THE EDGE OF THE BED. IT IS NOTED THAT THE PATIENT WAS ALERT AND TALKATIVE FOLLOWING THE FALL BUT WAS SUBSEQUENTLY TAKEN TO THE HOSPITAL BY AMBULANCE DUE TO ¿PAIN.¿ THE PATIENT WAS PROVIDED OXYGEN AND A CT OF THE BRAIN, CERVICAL SPINE, AND LUMBAR SPINE WAS PERFORMED WHICH FOUND A ¿SLIGHT COMPRESSION IN THE L3 VERTEBRAE, THE BRAIN AND CERVICAL SPINE WERE NOTED TO BE UNREMARKABLE. IT IS NOTED THE OXYGEN WAS ¿PHASED OUT¿ AND THE PATIENT¿S PAIN WAS TREATED WITH BUPRENORPHINE PATCH WHICH WAS LATER CHANGED TO ¿OXYNORM AND ULTIMATELY NORGESIC AND ALVEDON.¿ THE PATIENT WAS REPORTED TO HAVE FREE MOBILIZATION WHEN DISCHARGED HOME TO HIS HEALTHCARE CENTER.

    Patient problems: Pain; Vertebral Fracture

    MAUDE report 20614700 ↗

FDA clearance and classification

The FDA regulates product code FSA (lift, patient, non-ac-powered) under 21 CFR 880.5510 as a Class I device. 21 CFR 880.5510 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Uno™ mobile lift 2010004

Has the Uno™ mobile lift 2010004 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2031-2025 (Class I), began on May 30, 2025: The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Uno™ mobile lift 2010004?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Nov 5, 2024 and Feb 24, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (2), “Device Fell” (1) and “Device Tipped Over” (1); the most common patient problems are “Pain” (2), “Laceration(s)” (1) and “Vertebral Fracture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Uno™ mobile lift 2010004?

None of the reports are classified as deaths, 2 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Uno™ mobile lift 2010004 FDA approved?

Devices of this type (product code FSA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Liko AB.

Who makes the Uno™ mobile lift 2010004, and is it still sold?

The FDA's device database lists Liko AB as the company responsible for it (brand Uno™ mobile lift, model 2010004). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Lift, patient, non-ac-powered devices

All Lift, patient, non-ac-powered devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page