Centrella Bed recallZ-0141-2023 · Baxter Healthcare Corporation · initiated Sep 30, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).
FDA root cause category: Device Design
What the company is doing
Baxter issued an Urgent Medical Device Correction on 09/30/2022. An updated letter was sent on 10/21/2022 to request that users temporarily stop using the WatchCare system. Hillrom, a Baxter company, has received reports of Radio Frequency (RF) interference with other devices near beds that are installed with WatchCare. Specifically, RF interference has been reported with telemetry devices, a bladder scanner, a fetal monitor/doppler, an infusion pump, and a personal use insulin pump. Other third-party medical devices may also be affected. While the system was developed to comply with the most-recent RF standards, it radiates RF that might affect other devices in the vicinity, including devices on both patients and staff members. Given the types of devices that may be affected on both patients and staff, serious harm or death may occur. No serious injuries or deaths reported. Actions to be taken: 1. Immediately locate all affected devices and remove them from clinical care areas (where possible) until the WatchCare functionality is disabled. Affected devices are those Centrella, Progressa and Versacare beds which have WatchCare functionality. Baxter will contact you to arrange for WatchCare to be disabled. Please contact Baxter at 1(800) 445- 3720 or [email protected] for assistance or discuss alternative actions while you wait for the WatchCare feature to be disabled. 2. Be aware that RF emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to, telemetry devices, bladder scanners, fetal monitor/dopplers, infusion pumps, and insulin pumps/blood glucose sensors). Please double-check all unexpected or atypical results and monitor infusions closely. 3. While Baxter continues to work to determine the etiology of this problem and the issue is resolved, use standard, non-RF-based, incontinence management pads. On April 06, 2023, an updated Medical Device Correction letter was sent to customers to infor
Recall details
- Recall number
- Z-0141-2023
- Status
- Open, Classified
- Initiated by the company
- Sep 30, 2022
- Posted by the FDA
- Oct 31, 2022
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 6407 units
- Product code
- FNL
- Distribution
- US
Product description
Hillrom Centrella Hospital Bed with WatchCare Incontinence Management System, Product code/Part numbers: P7900B100193, P7900B100195, P7900B100230, P7900B100236, P7900B100239, P7900B100244, P7900B100256, P7900B100257, P7900B100259, P7900B100264, P7900B100268, P7900B100269, P7900B100292, P7900B100296, P7900B100302, P7900B100303, P7900B100304, P7900B100305, P7900B100312, P7900B100316, P7900B100318, P7900B100335, P7900B100341, P7900B100342, P7900B100354, P7900B100361, P7900B100368, P7900B100372, P7900B100381, P7900B100382, P7900B100384, P7900B100385, P7900B100387, P7900B100388, P7900B100392, P7900B100395, P7900B100396, P7900B100407, P7900B100408, P7900B100438, P7900B100441, P7900B1PL270, P7900B1PLT71, P7900B1SPL05, P7900B200193, P7900B200195, P7900B200230, P7900B200236, P7900B200264, P7900B200268, P7900B200292, P7900B200302, P7900B200318, P7900B200342, P7900B200361, P7900B200372, P7900B200381, P7900B200382, P7900B200392, P7900B200395, P7900B200396, P7900B200408, P7900B200439, P7900B200441, P7900B200443, P7900B200445, P7900B200452, P7900B200453, P7900B200464, P7900B200485, P7900B2SPL12, P7900B300441, P7900B300453, P7900B300518
Code information (models, lots, serial numbers)
UDI/DI 887761985162, all serial numbers
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Centrella™ Bed P7900BHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL2 recalls5 deaths1,937 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 90865).
- Z-0142-2023 · Open, ClassifiedHillrom Hospital Bed Accessory, Watch Care Incontinence Management System for Centrella Bed, Product Code/Part Number P00697905
- Z-0143-2023 · Open, ClassifiedHillrom Hospital Bed Accessory, Watch Care Incontinence Management System for Progressa Bed, Product Code/Part Number P00697903.
- Z-0144-2023 · Open, ClassifiedHillrom Hospital Bed Accessory, WatchCare Incontinence Management System for VersaCare Bed Rev. K, Product Code/Part Number P00697902
- Z-0145-2023 · Open, ClassifiedHillrom Hospital Bed Accessory, Watch Care Incontinence Management System for VersaCare Bed Rev. A-J, Product Code/Part Number P00697901
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page