- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00887761996717), received between Mar 23, 2022 and Sep 3, 2025 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Improper or Incorrect Procedure or Method1
- Inflation Problem1
- Sparking1
Patient problems
- Unspecified Tissue Injury2
- No Clinical Signs, Symptoms or Conditions1
- Pressure Sore/Ulcer1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjurySep 3, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
IT WAS REPORTED A PATIENT SUSTAINED AN UNSTAGEABLE PRESSURE ULCER IN CONJUNCTION WITH A COMPELLA MATTRESS. THE PATIENT STATED WHEN THEY ¿RAISED THEIR HEAD AND KNEES THEY COULD FEEL THE METAL PART OF THE BED.¿ THE PATIENT WAS FOUND TO HAVE A NEW OPEN AREA ON THE PATIENT¿S BUTTOCKS. THE NURSE WAS UNABLE TO STAGE THE WOUND DUE TO ¿MINIMAL SLOUGH NOTED.¿ ADDITIONALLY, THE WOUND WAS MEASURED AT 2.2 CM X 1.5 CM X 0.1 CM (ESTIMATED). THE PATIENT WAS TREATED WITH MEDIHONEY AND COLLAGEN APPLICATION TO THE WOUND AND COVERED WITH A DRY PROTECTIVE DRESSING. THE BAXTER SERVICE TECHNICIAN INSPECTED THE BED. THE TECHNICIAN NOTED THE BED¿S CPR VALVE WAS OPEN, PUTTING THE BED INTO CPR MODE AND CAUSING THE MATTRESS TO DEFLATE. THE TECHNICIAN CLOSED THE CPR VALVE TO RESOLVE THE ISSUE, AND THE BED IS FUNCTIONING AS DESIGNED. THE CAUSE OF THE EVENT IS LIKELY DUE TO THE BED¿S CPR MODE BEING ENGAGED WHICH DID NOT ALLOW THE BED¿S MATTRESS TO REINFLATE (USER ERROR). WHEN THE CPR FUNCTION IS ENGAGED/ACTIVATED AIR IS RELEASED FROM THE BED¿S SURFACE UNTIL THE SURFACE IS COMPLETELY DEFLATED. THE DEVICE INSTRUCTIONS FOR USE STATE THE FOLLOWING WHEN USING THE BED¿S CPR CONTROLS. ¿TO INFLATE THE AIR SURFACE AFTER
Patient problems: Pressure Sore/Ulcer
MAUDE report 22954436 ↗MalfunctionDec 2, 2024Sparking; Sparking
BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE COMPELLA AIR MATTRESS, NA, R POWER CORD SPARKED AND SMOKED DUE TO SHORT. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED DURING PATIENT USE. THE REPORTER DID NOT KNOW IF THERE WAS ANY PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. IT WAS UNKNOWN IF ANY SECOND DEVICE WAS ALSO INVOLVED. THE CUSTOMER DID NOT KNOW IF THIS EVENT WAS ALREADY COMMUNICATED TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20821279 ↗InjurySep 1, 2022Inflation Problem; Inflation Problem
IT WAS REPORTED THAT A COMPELLA BED HAD NO AIR IN THE CENTER AND THE PATIENT DEVELOPED A DEEP TISSUE INJURY (DTI). THE BED WAS LOCATED AT THE ACCOUNT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: Unspecified Tissue Injury
MAUDE report 15335601 ↗InjuryMar 23, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
THE CUSTOMER REPORTED THE COMPELLA THE MATTRESS WAS NOT INFLATING, AND THE BED DISPLAYED AN ERROR CODE 6. THE CUSTOMER ALSO REPORTED THE PATIENT HAD A NEW DEEP TISSUE INJURY (DTI) POTENTIALLY CAUSED BY THE BED NOT INFLATING PROPERLY. THE BED WAS LOCATED AT THE ACCOUNT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Patient problems: Unspecified Tissue Injury
MAUDE report 13862956 ↗
Other BED, AC-POWERED ADJUSTABLE HOSPITAL devices
- Centrella™ Bed P7900BHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL2 recalls5 deaths1,937 reports
- Progressa® Bed System P7500AHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death1,377 reports
- VersaCare® Med Surg Bed P3200KHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL3 deaths1,193 reports
- Medline MDR107002EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death374 reports
- Medline MDR107003EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall3 deaths362 reports
- InTouch 2141Stryker Corporation · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths361 reports
- Citadel Plus FX811B3B4AMABBArjohuntleigh AB · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths358 reports
- TotalCare® Bed P1900QHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL317 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page