Skip to content

Compella™Surface P642AREN0100000Recalls & reported problems

Hill-Rom, Inc. · Hospital Bed Surface · Class 2

All models of CompellaSurface (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00887761996717), received between Mar 23, 2022 and Sep 3, 2025 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available1
  • Improper or Incorrect Procedure or Method1
  • Inflation Problem1
  • Sparking1

Patient problems

  • Unspecified Tissue Injury2
  • No Clinical Signs, Symptoms or Conditions1
  • Pressure Sore/Ulcer1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjurySep 3, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    IT WAS REPORTED A PATIENT SUSTAINED AN UNSTAGEABLE PRESSURE ULCER IN CONJUNCTION WITH A COMPELLA MATTRESS. THE PATIENT STATED WHEN THEY ¿RAISED THEIR HEAD AND KNEES THEY COULD FEEL THE METAL PART OF THE BED.¿ THE PATIENT WAS FOUND TO HAVE A NEW OPEN AREA ON THE PATIENT¿S BUTTOCKS. THE NURSE WAS UNABLE TO STAGE THE WOUND DUE TO ¿MINIMAL SLOUGH NOTED.¿ ADDITIONALLY, THE WOUND WAS MEASURED AT 2.2 CM X 1.5 CM X 0.1 CM (ESTIMATED). THE PATIENT WAS TREATED WITH MEDIHONEY AND COLLAGEN APPLICATION TO THE WOUND AND COVERED WITH A DRY PROTECTIVE DRESSING. THE BAXTER SERVICE TECHNICIAN INSPECTED THE BED. THE TECHNICIAN NOTED THE BED¿S CPR VALVE WAS OPEN, PUTTING THE BED INTO CPR MODE AND CAUSING THE MATTRESS TO DEFLATE. THE TECHNICIAN CLOSED THE CPR VALVE TO RESOLVE THE ISSUE, AND THE BED IS FUNCTIONING AS DESIGNED. THE CAUSE OF THE EVENT IS LIKELY DUE TO THE BED¿S CPR MODE BEING ENGAGED WHICH DID NOT ALLOW THE BED¿S MATTRESS TO REINFLATE (USER ERROR). WHEN THE CPR FUNCTION IS ENGAGED/ACTIVATED AIR IS RELEASED FROM THE BED¿S SURFACE UNTIL THE SURFACE IS COMPLETELY DEFLATED. THE DEVICE INSTRUCTIONS FOR USE STATE THE FOLLOWING WHEN USING THE BED¿S CPR CONTROLS. ¿TO INFLATE THE AIR SURFACE AFTER

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 22954436 ↗
  • MalfunctionDec 2, 2024Sparking; Sparking

    BAXTER RECEIVED A REPORT FROM A BAXTER TECHNICIAN TO REPORT THE COMPELLA AIR MATTRESS, NA, R POWER CORD SPARKED AND SMOKED DUE TO SHORT. THE BED WAS LOCATED AT THE ACCOUNT. THIS OCCURRED DURING PATIENT USE. THE REPORTER DID NOT KNOW IF THERE WAS ANY PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. IT WAS UNKNOWN IF ANY SECOND DEVICE WAS ALSO INVOLVED. THE CUSTOMER DID NOT KNOW IF THIS EVENT WAS ALREADY COMMUNICATED TO ANOTHER BAXTER DEPARTMENT OR AUTHORITY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20821279 ↗
  • InjurySep 1, 2022Inflation Problem; Inflation Problem

    IT WAS REPORTED THAT A COMPELLA BED HAD NO AIR IN THE CENTER AND THE PATIENT DEVELOPED A DEEP TISSUE INJURY (DTI). THE BED WAS LOCATED AT THE ACCOUNT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

    Patient problems: Unspecified Tissue Injury

    MAUDE report 15335601 ↗
  • InjuryMar 23, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE CUSTOMER REPORTED THE COMPELLA THE MATTRESS WAS NOT INFLATING, AND THE BED DISPLAYED AN ERROR CODE 6. THE CUSTOMER ALSO REPORTED THE PATIENT HAD A NEW DEEP TISSUE INJURY (DTI) POTENTIALLY CAUSED BY THE BED NOT INFLATING PROPERLY. THE BED WAS LOCATED AT THE ACCOUNT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

    Patient problems: Unspecified Tissue Injury

    MAUDE report 13862956 ↗

Other BED, AC-POWERED ADJUSTABLE HOSPITAL devices

All BED, AC-POWERED ADJUSTABLE HOSPITAL devices →