Skip to content

Compella™Surface P642ACAP0100000Recalls & reported problems

Hill-Rom, Inc. · Hospital Bed Surface · Class 2

All models of CompellaSurface (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00887761996779), received between Feb 22, 2023 and Feb 22, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Use of Device Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Use of Device Problem1

Patient problems

  • Pressure Sore/Ulcer1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 22, 2023Use of Device Problem; Use of Device Problem

    THE CUSTOMER INITIALLY REPORTED THE COMPELLA ¿CONNECTING TUBE¿ (THAT KEEPS THE CENTER SECTION OF THE DEVICE INFLATED) EASILY BECOMES UNDONE, LEAVING THE PATIENT LYING ON A FLAT MATTRESS AND ALLEGED THE PATIENT HAD A PRESSURE ULCER. FOLLOW UP WITH THE CUSTOMER CONFIRMED THE PATIENT DEVELOPED A DEEP TISSUE INJURY TO BOTH THEIR SACRUM AND LEFT BUTTOCKS DUE TO USING THE DEVICE WITH HOSES DISCONNECTED (DEFLATED) AND ADDITIONALLY STATING THE PATIENT WAS NOT TURNED PER CUSTOMER POLICY. MEDICAL INTERVENTION TO BOTH THE SACRUM AND BUTTOCKS INCLUDED: CLEANING WITH NORMAL SALINE, APPLICATION OF Z GUARD AND SANTYLE TO WOUND. THE BED WAS LOCATED AT THE ACCOUNT. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 16417361 ↗

Other BED, AC-POWERED ADJUSTABLE HOSPITAL devices

All BED, AC-POWERED ADJUSTABLE HOSPITAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page