Skip to content

Comprehensive® Shoulder System 113606Recalls & reported problems

Biomet Orthopedics, LLC · Class 2 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

6 other recalls cover devices cleared under the same FDA submission (K060692, K193038); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Mar 3, 2025Open, Classified

    Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.

    Z-1403-2025 · Root cause: Nonconforming Material/Component · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K060692, K193038. One submission often covers many models, so check the product and lot details in each record.

  • Z-1543-2020 · Feb 17, 2020 · Terminated
    Comprehensive Shoulder System Mini Humeral Stem 9mm,, Model No. 113629 - Product Usage: The intended use of the comprehensive humeral stem mini is to act as the humeral component o
  • Z-0591-2019 · Oct 10, 2018 · Terminated
    Comprehensive Mini Stem, Item No. 113631:
  • Z-0979-2018 · Dec 8, 2017 · Terminated
    Humeral Stem 83mm(Size 17) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
  • Z-0980-2018 · Dec 8, 2017 · Terminated
    Humeral Stem 55mm(Size 13) Product Usage: Intended for primary and revision total shoulder arthroplasty as well as hemi-arthroplasty
  • Z-0299-2018 · Jun 5, 2017 · Terminated
    COMP PRIMARY STEM 8MM MINI, model # 113628
  • Z-1755-2015 · Apr 16, 2015 · Terminated
    Comprehensive Shoulder System, 4 mm, 122 mm long, Primary Shoulder Stem, Ti-6Al-4V, Standard Length Porous Plasma. Shoulder joint replacement components include humeral stems, hum

Other PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED devices →