Skip to content

Comprehensive Shoulder System recallZ-1403-2025 · Zimmer Orthopedics Manfacturing Limited-Galway · initiated Mar 3, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer Orthopedics Manfacturing Limited-Galway or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On March 3, 2025, the firm notified affected customers via Urgent Medical Device Recall letters. Customers were informed of the excess metal on the inner taper of the stem due to a chipped manufacturing tool. Customers were asked to notify all affected persons and assist their Zimmer Biomet sales representative in quarantining all affected product on hand. The products will be removed from your facility.

Recall details

Recall number
Z-1403-2025
Status
Open, Classified
Initiated by the company
Mar 3, 2025
Posted by the FDA
Mar 20, 2025
Recalling firm
Zimmer Orthopedics Manfacturing Limited-Galway
Quantity
90 US; 136 OUS
Product code
MBF
FDA submissions
K193038
Distribution
US distribution nationwide. International distribution to Dominican Republic, Canada, India, Australia, Japan, EMEA, Korea, Taiwan, and Hong Kong.

Product description

Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627

Code information (models, lots, serial numbers)

Item No. 113605, UDI-DI 00887868267550 Lots 66173214, 66173215; Item No. 113606, UDI-DI 00887868267567 Lots 66173216, 66173218, 66173219; Item No. 113607, UDI-DI 00887868267574 Lots 66173220, 66173221; Item No. 113627, UDI-DI 00887868267741 Lots 65901697, 65909480, 65915809

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

38 devices cleared under K193038 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates