Vital™ Navigation System 854M0021Recalls & reported problems
Zimmer Biomet Inc · Vital™ Navigation Driver · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00887868310232), received between Mar 8, 2022 and Jan 12, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Material Twisted/Bent3
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Insufficient Information1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 12, 2026Material Twisted/Bent
IT WAS REPORTED THAT DURING A REVISION SURGERY TO EXTEND A CONSTRUCT, THE TIP OF A VITALITY NAVIGATION DRIVER STRIPPED. THE SURGERY WAS COMPLETED USING THE NAVIGATION DRIVER, AND THERE WAS NO PATIENT HARM OR PROCEDURAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24038652 ↗MalfunctionJun 25, 2025Fracture; Fracture
IT WAS REPORTED THAT TWO NAVIGATED VITAL DRIVERS BROKE DURING SURGERY WHEN PLACING A SCREW; THE STAR TIP BROKE ON BOTH AND BOTH WERE RETRIEVED. THE SCREW WAS ALL THE WAY IN AND FINAL TIGHTENED WHEN THE EVENT HAPPENED; THERE WAS NO PATIENT IMPACT. THIS IS REPORT TWO OF TWO FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22329946 ↗MalfunctionJun 25, 2025Fracture; Fracture
IT WAS REPORTED THAT TWO NAVIGATED VITAL DRIVERS BROKE DURING SURGERY WHEN PLACING A SCREW; THE STAR TIP BROKE ON BOTH AND BOTH WERE RETRIEVED. THE SCREW WAS ALL THE WAY IN AND FINAL TIGHTENED WHEN THE EVENT HAPPENED; THERE WAS NO PATIENT IMPACT. THIS IS REPORT ONE OF TWO FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22329834 ↗MalfunctionJan 25, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGERY, THREE VITALITY DRIVERS FRACTURED WHEN TRYING TO REMOVE A STUCK ILIAC SCREW. THE SCREW WAS ULTIMATELY ABLE TO BE REMOVED AND REPLACED WITH A SHORTER SCREW. THIS RESULTED IN A 20 MINUTE DELAY TO THE PROCEDURE BUT NO OTHER PATIENT IMPACT. THIS IS REPORT FOUR OF FOUR FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18584934 ↗MalfunctionFeb 13, 2023Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT TWO VITALITY CONNECTORS BROKE, THE TIP OF ONE VITALITY T20 FINAL DRIVER BROKE, THE TIP OF A VITALITY T27 DRIVER TWISTED, AND THE TIPS OF TWO VITAL NAVIGATION T27 DRIVERS TWISTED INTRA-OPERATIVELY. OTHER CONNECTORS AND DRIVERS WERE USED TO COMPLETE THE PROCEDURE. THERE WAS A DELAY TO THE PROCEDURE THAT DID NOT LEAD TO PATIENT IMPACTS. THIS IS REPORT SIX OF SIX FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16362892 ↗MalfunctionFeb 13, 2023Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT TWO VITALITY CONNECTORS BROKE, THE TIP OF ONE VITALITY T20 FINAL DRIVER BROKE, THE TIP OF A VITALITY T27 DRIVER TWISTED, AND THE TIPS OF TWO VITAL NAVIGATION T27 DRIVERS TWISTED INTRA-OPERATIVELY. OTHER CONNECTORS AND DRIVERS WERE USED TO COMPLETE THE PROCEDURE. THERE WAS A DELAY TO THE PROCEDURE THAT DID NOT LEAD TO PATIENT IMPACTS. THIS IS REPORT FIVE OF SIX FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16362884 ↗MalfunctionMar 8, 2022Material Deformation; Material Deformation
THE TIP OF A SCREW INSERTER WAS FOUND TO BE TWISTED DURING A ROUTINE INSPECTION. THE SPECIFIC SURGICAL OR PATIENT INFORMATION IS UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 13690653 ↗
Other Orthopedic Stereotaxic Instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page