Skip to content

G7® 110003450Recalls & reported problems

Biomet Orthopedics, LLC · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This is a Biomet hip replacement cup used in hip arthroplasty. The 29 reports mostly describe inserter handles and shells that would not thread properly, stripped or broken threads, and parts deforming or fracturing during surgery, sometimes forcing the team to switch to another handle. Two injuries and no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
28
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 00887868462481), received between Aug 2, 2023 and Dec 30, 2025 — classified as 2 injuries, 26 malfunctions.

The most frequently reported specific device problem is “Damaged Thread” (7 reports).

3 other recalls cover devices cleared under the same FDA submission (K121874); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 5 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 3 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Damaged Thread7
  • Difficult to Insert6
  • Material Deformation5
  • Fracture4
  • Physical Resistance/Sticking3
  • Separation Failure3
  • Difficult to Remove2
  • Mechanical Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions28

Recent adverse event reports

The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 30, 2025Damaged Thread; Damaged Thread

    IT WAS REPORTED THE THREADS ON A MONOBLOCK CUP INSERTER BROKE. THERE WAS NO PATIENT INVOLVEMENT. NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23927889 ↗
  • MalfunctionDec 18, 2025Difficult to Insert; Damaged Thread; Difficult to Insert; Damaged Thread

    IT WAS REPORTED THAT THE INSERTER HANDLE TIP IS STRIPPED AND CANNOT THREAD ONTO THE SHELL PROPERLY. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23838276 ↗
  • MalfunctionSep 4, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22967336 ↗
  • InjurySep 3, 2025Separation Failure; Physical Resistance/Sticking; Damaged Thread; Separation Failure; Physical Resistance/Sticking; Damaged Thread

    UPON RECEIPT OF ADDITIONAL INFORMATION, THE PREVIOUSLY REPORTED EVENT WAS FOUND TO BE A DUPLICATE COMPLAINT ENTRY. THE EVENT HAS BEEN REPORTED UNDER 0001825034-2025-02673.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22956738 ↗
  • MalfunctionAug 25, 2025Separation Failure; Physical Resistance/Sticking; Damaged Thread; Separation Failure; Physical Resistance/Sticking; Damaged Thread

    NO FURTHER EVENT INFORMATION AT TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22885594 ↗
  • MalfunctionJul 15, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE SHELL INSERTER WAS FOUND BROKEN IN THE TRAY AND COULD NOT BE SCREWED TO THE CUP. ANOTHER HANDLE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. IT WAS REPORTED NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22506009 ↗
  • MalfunctionMay 12, 2025Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THAT THE INSERTER HANDLE WOULD NOT THREAD ONTO THE NEW SHELL. THE SHELL WAS SUCCESSFULLY INSERTED WITH ANOTHER HANDLE. THERE WAS NO KNOWN IMPACT OR CONSEQUENCES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22005321 ↗
  • InjuryApr 15, 2025Damaged Thread; Damaged Thread

    IT WAS REPORTED THAT THE THREADS ON THE G7 MONOBLOCK STRAIGHT IMPACTOR WERE NOTICED TO BE DAMAGED. NO KNOWN IMPACT OR CONSEQUENCE TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21842305 ↗
  • MalfunctionFeb 5, 2025Fracture; Fracture

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21315752 ↗
  • MalfunctionSep 3, 2024Difficult to Insert; Difficult to Insert

    IT WAS REPORTED THE CUP INSERTER WOULD NOT THREAD PROPERLY. ANOTHER INSTRUMENT WAS USED TO COMPLETE THE CASE. THERE WAS NO HARM OR HEALTH CONSEQUENCES TO THE PATIENT. IT WAS REPORTED THAT NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20127582 ↗
  • MalfunctionAug 26, 2024Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE SURGEON COULD NOT DISENGAGE NON-MODULAR INSERTER FROM THE G7 SHELL, SURGEON CLAIMS THE THREADS OF THE INSERTER ARE SLIGHTLY DAMAGED. THE SURGERY WAS COMPLETED WITH A NEW SHELL AND INSERTER. THE SURGICAL TECHNIQUE OF THE PRODUCTS WAS UTILIZED. A MINOR DELAY TOOK PLACE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20065715 ↗
  • MalfunctionJul 16, 2024Damaged Thread; Damaged Thread

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19752774 ↗
  • MalfunctionJul 10, 2024Damaged Thread; Damaged Thread

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE INSERTER HANDLE WAS CROSS THREADED WITH THE SHELL. THE SHELL WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO HARM OR HEALTH CONSEQUENCES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19710857 ↗
  • MalfunctionJul 5, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THE IMPACTOR THREADS WERE DAMAGED WHICH LED TO DAMAGED MONOBLOCK G7 IMPACTOR. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19679179 ↗
  • MalfunctionApr 17, 2024Difficult to Insert; Naturally Worn; Difficult to Insert; Naturally Worn

    THERE IS NO UPDATE TO THE PRIOR EVENT DESCRIPTION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19124842 ↗
  • MalfunctionApr 9, 2024Separation Failure; Separation Failure

    IT WAS REPORTED THAT WHEN THE SURGEON WAS IMPLANTING THE ACETABULAR SHELL INTO THE PATIENT, THE STRAIGHT HANDLE CUP INSERTER WOULD NOT UNSCREW FROM THE SHELL. THE PROCEDURE WAS COMPLETED USING A NEW INSERTER AND IMPLANT. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19075599 ↗
  • MalfunctionApr 9, 2024Mechanical Problem; Mechanical Problem

    NO ADDITIONAL EVEN INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19070557 ↗
  • MalfunctionMar 28, 2024Fracture; Material Deformation; Fracture; Material Deformation

    IT WAS REPORTED DURING SURGERY THAT THE LAST THREAD ON THE G7 INSERTER BROKE OFF FROM WEAR. THERE WAS NO REPORTED HARM OR IMPACT TO THE PATIENT. THE SURGERY WAS COMPLETED WITH ANOTHER DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18994751 ↗
  • MalfunctionFeb 5, 2024Difficult to Remove; Difficult to Remove

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18640471 ↗
  • MalfunctionDec 29, 2023Mechanical Jam; Mechanical Jam

    NO FURTHER INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18418144 ↗
  • MalfunctionDec 18, 2023Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED DURING THE HAMMERING OF THE SHELL, IT WAS IMPOSSIBLE TO UNSCREW THE SHELL FROM THE IMPACTOR. NO KNOWN CONSEQUENCES OR IMPACT TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18344534 ↗
  • MalfunctionDec 16, 2023Difficult to Remove; Difficult to Remove

    NO FURTHER INFORMATION AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18341600 ↗
  • MalfunctionDec 15, 2023Material Deformation; Material Deformation

    IT WAS REPORTED THAT WHILE PREPARING FOR THE CASE, THE THREADS ON THE CUP INSERTER WERE FOUND TO BE DAMAGED AND WOULD NOT WORK. ANOTHER INSERTER WAS USED TO COMPLETE THE CASE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18334649 ↗
  • MalfunctionDec 1, 2023Difficult to Insert; Difficult to Insert

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18247467 ↗
  • MalfunctionNov 9, 2023Material Deformation; Material Deformation

    NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18100569 ↗

FDA clearance and classification

Cleared through 510(k) K121874 on Nov 23, 2012 (decision: Substantially Equivalent). Submitted by Biomet Manufacturing Corp as “G7 Acetabular System”. The FDA review took 149 days. FDA record ↗

The FDA regulates product code PBI (prosthesis, hip, constrained, cemented or uncemented, metal/polymer, + additive) under 21 CFR 888.3310 as a Class II device. 21 CFR 888.3310 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K121874. One submission often covers many models, so check the product and lot details in each record.

  • Z-0007-2023 · Aug 25, 2022 · Open, Classified
    G7 Acetabular System, Acetabular Shell, Item numbers Item Numbers: 010000663 (52 MM E), 010000667 (60 MM G). used in hip arthroplasty
  • Z-1838-2014 · May 7, 2014 · Terminated
    G7 Acetabular System Positioning Guide Post, orthopedic surgical instrument for hip prosthesis. PN:110003500.
  • Z-1837-2014 · May 7, 2014 · Terminated
    G7 Acetabular System Positioning Guide Rod, orthopedic surgical instrument for hip prosthesis PN 110003458.

Common questions about the G7® 110003450

Has the G7® 110003450 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00887868462481), based on the FDA recall file dated Nov 27, 2026. However, 3 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the G7® 110003450?

The FDA's MAUDE database lists 28 adverse event reports for this device, received between Aug 2, 2023 and Dec 30, 2025. In those reports, the most common specific device problems are “Damaged Thread” (7), “Difficult to Insert” (6) and “Material Deformation” (5). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the G7® 110003450?

None of the reports are classified as deaths, 2 as injuries and 26 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the G7® 110003450 FDA approved?

It was cleared through 510(k) K121874 on Nov 23, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the G7® 110003450, and is it still sold?

The FDA's device database lists Biomet Orthopedics, LLC as the company responsible for it (brand G7®, model 110003450). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other prosthesis, hip, constrained, cemented or uncemented, metal/polymer, + additive devices

All prosthesis, hip, constrained, cemented or uncemented, metal/polymer, + additive devices →

Sources and updates