Skip to content

Comprehensive® Instrumentation 110028879Recalls & reported problems

Biomet Orthopedics, LLC · Class 2

All models of COMPREHENSIVE INSTRUMENTATION (18)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Biomet Comprehensive shoulder prosthesis is a cemented, semi-constrained metal and polymer implant used to replace the shoulder joint. The 28 reports describe fractures breaking off or breaking apart during surgery procedures, most often involving the glenosphere inserter tip, with no patient harm reported. There have been no deaths and no injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
28
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 00887868470011), received between Apr 1, 2021 and Jan 2, 2026 — classified as 28 malfunctions.

The most frequently reported specific device problem is “Fracture” (28 reports).

20 other recalls cover devices cleared under the same FDA submission (K080642, K132239, K193373); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 3 reportsApr 2025: 1 reportsMay 2025: 2 reportsJun 2025: 2 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture28
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions28

Recent adverse event reports

The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 2, 2026Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23945748 ↗
  • MalfunctionDec 15, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23802217 ↗
  • MalfunctionDec 9, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23761105 ↗
  • MalfunctionNov 26, 2025Break; Fracture; Break; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23657877 ↗
  • MalfunctionNov 11, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A PROCEDURE, THE BLACK PART OF THE GLENOSPHERE INSERTER BROKE OFF. NO PATIENT CONSEQUENCES WERE REPORTED AS A RESULT OF THE EVENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23525612 ↗
  • MalfunctionSep 4, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22973937 ↗
  • MalfunctionAug 15, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22815376 ↗
  • MalfunctionJul 23, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIP OF THE INSTRUMENT WAS FRACTURED. THERE WAS NO REPORT OF FOREIGN BODY RETAINED OR HARM TO A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22575822 ↗
  • MalfunctionJul 10, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A PROCEDURE, THE PLASTIC TIP OF THE LOCKING GLENOID IMPACTOR FRACTURED DURING IMPACTION. ANOTHER IMPACTOR WAS USED TO COMPLETE THE CASE AND NO PATIENT CONSEQUENCES WERE REPORTED AS A RESULT OF THE EVENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22472044 ↗
  • MalfunctionJun 17, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING AN INITIAL PROCEDURE, THE IMPACTOR TIP FRACTURED DURING IMPACTION TO THE GLENOSPHERE WHILE FOLLOWING THE CORRECT SURGICAL TECHNIQUE. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE INSTRUMENT. NO COMPLICATIONS, INJURIES, OR SURGICAL INTERVENTIONS WERE REQUIRED, AND NO FOREIGN BODIES WERE RETAINED DUE TO THIS MALFUNCTION. THE CASE WAS COMPLETED WITHOUT ANY ADVERSE OUTCOMES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22232858 ↗
  • MalfunctionJun 2, 2025Fracture; Fracture

    IT WAS REPORTED THAT WHILE THE SURGEON WAS ATTEMPTING TO IMPACT THE IMPLANT THE TIP FRACTURED IN HALF. THE SURGEON REMOVED THE FRACTURED PIECE FROM THE PATIENT. A SECOND INSTRUMENT WAS USED TO COMPLETE THE SURGERY. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22125934 ↗
  • MalfunctionMay 28, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22098904 ↗
  • MalfunctionMay 19, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING SURGERY THAT THE IMPACTOR TIP WAS FRACTURED. THE PATIENT DID NOT RETAIN ANY FOREIGN MATERIAL FROM THE FRACTURED DEVICE. THERE WAS NO HARM, INJURY, SURGICAL/MEDICAL INTERVENTION TO PATIENT AND NO REPORT OF A DELAY. A BACK UP TRAY WAS USED TO COMPLETE THE PROCEDURE. DUE DILIGENCE IS COMPLETE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22044503 ↗
  • MalfunctionApr 29, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21929939 ↗
  • MalfunctionMar 28, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING AN INITIAL PROCEDURE, THE IMPACTOR TIP FRACTURED DURING IMPACTION TO THE GLENOSPHERE WHILE FOLLOWING THE CORRECT SURGICAL TECHNIQUE. THE SURGEON NOTED THE DEVICE TIP WAS BROKEN AFTER COMPLETING ITS INTENDED USE. NO COMPLICATIONS, INJURIES, OR SURGICAL INTERVENTIONS WERE REQUIRED, AND NO FOREIGN BODIES WERE RETAINED DUE TO THIS MALFUNCTION. THE CASE WAS COMPLETED WITHOUT ANY ADVERSE OUTCOMES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21721490 ↗
  • MalfunctionMar 27, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING AN INITIAL PROCEDURE, THE IMPACTOR TIP FRACTURED DURING IMPACTION ON THE GLENOSPHERE WHILE FOLLOWING THE CORRECT SURGICAL TECHNIQUE. DURING THE FIRST USE OF THE DEVICE, THE PLASTIC PIECE OF THE DEVICE BROKE OFF. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE INSTRUMENT. NO COMPLICATIONS, INJURIES, OR SURGICAL INTERVENTIONS WERE REQUIRED, AND NO FOREIGN BODIES WERE RETAINED DUE TO THIS MALFUNCTION. THE CASE WAS COMPLETED WITHOUT ANY ADVERSE OUTCOMES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21713981 ↗
  • MalfunctionMar 21, 2025Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21663124 ↗
  • MalfunctionFeb 18, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIP OF THE IMPACTOR BROKE. NO ADVERSE EVENT HAS BEEN REPORTED AS A RESULT OF THE MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21402705 ↗
  • MalfunctionJul 22, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19794902 ↗
  • MalfunctionJun 4, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19461710 ↗
  • MalfunctionMay 22, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19371160 ↗
  • MalfunctionMay 21, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19366349 ↗
  • MalfunctionMar 6, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING AN INITIAL PROCEDURE, THE BLACK PLASTIC TIP WAS FRACTURED AS THE SURGEON IMPACTED THE GLENOSPHERE. ALL PIECES WERE RETRIEVED, AND A SECONDARY IMPACTOR WAS USED TO FINISH THE SURGERY. NO COMPLICATIONS, INJURIES, SURGICAL INTERVENTIONS, OR FOREIGN BODIES WERE RETAINED DUE TO THIS MALFUNCTION. THE CASE WAS COMPLETED WITHOUT ANY ADVERSE OUTCOMES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18851033 ↗
  • MalfunctionFeb 1, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING INITIAL SHOULDER SURGERY, THE IMPACTOR FRACTURED DURING GLENOSPHERE IMPACTION. THE IMPLANT WAS ALREADY INSERTED AND SEATED. THE SURGEON RETRIEVED THE BROKEN PIECES AND IMPACTED THE IMPLANT AGAIN WITH A SECONDARY IMPACTOR. THE CORRECT SURGICAL TECHNIQUE WAS USED. NO COMPLICATIONS, INJURIES, OR SURGICAL INTERVENTIONS WERE REQUIRED, AND NO FOREIGN BODIES WERE RETAINED DUE TO THIS MALFUNCTION. THE CASE WAS COMPLETED WITHOUT ANY ADVERSE OUTCOMES TO THE PATIENT. ATTEMPTS HAVE BEEN MADE, AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18627804 ↗
  • MalfunctionDec 4, 2023Fracture; Fracture

    NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18258871 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K080642, K132239, K193373. One submission often covers many models, so check the product and lot details in each record.

  • Z-1131-2021 · Jan 15, 2021 · Terminated
    Central Screw 6.5x25mm - Product Usage: used as a sterile screw that is intended for implantable shoulder prosthesis assembly and/or fixation.
  • Z-0619-2021 · Nov 25, 2020 · Terminated
    Comprehensive Reverse Shoulder System, Item numbers 115394, 115395, 115396, 115398, 180550, 180551, 180552, 180553, 180554, 180555, 180556, 180557, 180558, 180559, 180560, 1805
  • Z-0093-2021 · Sep 2, 2020 · Terminated
    Biomet Comprehensive Reverse Shoulder Glenosphere Mini Baseplate with Taper Adapter-Shoulder Prosthesis item: 010000589 - Product Usage: intended to serve as the foundation for t
  • Z-2144-2020 · Apr 17, 2020 · Terminated
    Extremities Item Number: 1)113954 Hybrid Glenoid Glenoid Base, 4 MM 2) XL-115366 Comprehensive Reverse Shoulder Humeral Bearing, ArComXL, 44 MM X 41 MM, Standard 3) XL-115364 Comp
  • Z-1750-2020 · Mar 13, 2020 · Terminated
    Comprehensive Humeral Tray Taper Extraction Pliers - Replacement Tips, Model Number 110028522 - Product Usage: Intended to be used in surgery to relieve pain and to replace the gle
  • Z-1859-2017 · Feb 17, 2017 · Terminated
    various polyethylene implants Extremities, Shoulders and Elbows BioModular Shoulder System; Comprehensive Shoulder System; Comprehensive Reverse Shoulder System Product Usage: Sh
  • Z-1103-2017 · Dec 20, 2016 · Terminated
    Comprehensive Reverse Shoulder System Humeral Tray Model 115340
  • Z-1548-2013 · May 15, 2013 · Terminated
    Superior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain.
  • Z-2548-2011 · Apr 22, 2011 · Terminated
    Custom biomodular humeral tray 44 mm sterile, REF CP561686, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder products are indicated for use in patients who
  • Z-2549-2011 · Apr 22, 2011 · Terminated
    Custom comprehensive reverse humeral tray set w/ lock rings 44 mm + 15 mm & + 18 mm sterile, REF CP561705, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder
  • Z-2547-2011 · Apr 22, 2011 · Terminated
    Custom solar comprehensive reverse humeral tray with lock ring 44 mm +5mm, sterile, REF CP561659, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder product
  • Z-2550-2011 · Apr 22, 2011 · Terminated
    Custom comprehensive reverse shoulder bio-modular base 44 mm humeral tray standard, REF CP561510, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder product
  • Z-2541-2011 · Apr 22, 2011 · Terminated
    Bio-modular reverse shoulder humeral tray with lock ring, 44 mm 10 degree angle, sterile, single use, REF 11-113685, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse S
  • Z-2544-2011 · Apr 22, 2011 · Terminated
    Comprehensive reverse shoulder humeral tray with locking ring, 44 mm +5 mm, sterile, single use, REF 115345, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulde
  • Z-2542-2011 · Apr 22, 2011 · Terminated
    Bio-modular reverse shoulder humeral tray with lock ring, 44 mm standard, sterile, single use, REF 11-113680, Biomet Orthopedics, Warsaw, IN. Biomet Comprehensive Reverse Shoulder

…and 5 more.

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page