LEVEL ONE CMF 25-872-07-91Recalls & reported problems
KLS-Martin L.P. · SCREW, MAXDRIVE, TI-6AL-4V · Class 2 · Implant
All models of LEVEL ONE CMF (150)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00888118046734), received between Dec 6, 2016 and Mar 12, 2025 — classified as 7 injuries.
The most frequently reported specific device problem is “Osseointegration Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Osseointegration Problem4
- Adverse Event Without Identified Device or Use Problem2
- Break1
- Detachment Of Device Component1
- Fracture1
- Patient-Device Incompatibility1
Patient problems
- Impaired Healing3
- Device Embedded In Tissue or Plaque2
- Inadequate Osseointegration2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryMar 12, 2025Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED A CMF SCREW BECAME LOOSE FROM A CONSOLIDATING MANDIBULAR OSTEOTOMY SITE 10 WEEKS AFTER IMPLANT. IT WAS REMOVED AND REPLACED.
Patient problems: Inadequate Osseointegration
MAUDE report 21588166 ↗InjuryJul 3, 2024Osseointegration Problem; Osseointegration Problem
IT WAS REPORTED THAT NON-LOCKING FIXATION SCREWS IMPLANTED IN A MANDIBLE HAD BECOME LOOSENED FROM COMPROMISED BONE 17 MONTHS AFTER IMPLANTATION. THEY WERE REMOVED, DISCARDED, AND REPLACED.
Patient problems: Inadequate Osseointegration
MAUDE report 19670596 ↗InjuryNov 11, 2021Patient-Device Incompatibility; Patient-Device Incompatibility
IT WAS REPORTED THAT SCREWS WERE OBSERVED TO HAVE LOOSENED FROM RAMUS DUE TO PATIENT'S ANATOMICAL ANOMALY.
Patient problems: Impaired Healing
MAUDE report 12800470 ↗InjuryAug 20, 2021Adverse Event Without Identified Device or Use Problem; Osseointegration Problem; Adverse Event Without Identified Device or Use Problem; Osseointegration Problem
IT WAS REPORTED A SCREW BECAME LOOSE LIKELY DUE TO POOR BONE QUALITY; IT WAS REMOVED.
Patient problems: Impaired Healing
MAUDE report 12348170 ↗InjuryApr 8, 2021Adverse Event Without Identified Device or Use Problem; Osseointegration Problem; Adverse Event Without Identified Device or Use Problem; Osseointegration Problem
PER THE DOCTOR IT WAS REPORTED SCREWS BECAME LOOSE DUE TO PATIENT'S NON-COMPLIANCE AND WERE REMOVED.
Patient problems: Impaired Healing
MAUDE report 11637658 ↗InjurySep 7, 2018Break; Break
IT WAS REPORTED THAT A SCREW BROKE DURING IMPLANTATION. A PORTION OF THE SCREW REMAINED IN THE PATIENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 7856079 ↗InjuryDec 6, 2016Detachment Of Device Component; Fracture
THE SURGEON WAS PLATING A LEFT ANGLE FRACTURE WHEN THE SCREW HEAD FRACTURED OFF. THE REMAINDER OF THE SCREW IS STILL IN THE PATIENTS MANDIBLE. THERE WERE NO CONSEQUENCES TO THE PATIENT BESIDES THE REST OF THE SCREW BEING LEFT IN THE MANDIBLE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6147379 ↗
Other PLATE, BONE devices
- MatrixNEURO 04.503.104.01Synthes GmbH · PLATE, BONE252 reports
- RapidSorb 806.004.02SSynthes GmbH · PLATE, BONE146 reports
- TruMatch SD980.001Materialise NV · Plate, Bone1 recall21 reports
- Matrix 04.511.225.01Synthes GmbH · PLATE, BONE62 reports
- NA 400.834Synthes GmbH · PLATE, BONE58 reports
- TruMatch SD980.016Materialise NV · Plate, Bone55 reports
- ProPlan CMF SD900.101Materialise NV · Plate, Bone1 recall1 report
- TruMatch SD980.013Materialise NV · Plate, Bone1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page