LEVEL ONE CMF 50-760-26-09Recalls & reported problems
KLS-Martin L.P. · PLATE, RECON, TLTS, ANGLE, RIGHT, 2.0-2.5 MM SCREW, CP TITANIUM · Class 2 · Implant
All models of LEVEL ONE CMF (151)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 17
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00888118078216), received between Dec 20, 2016 and May 20, 2026 — classified as 17 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (13 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break13
- Fracture4
- Adverse Event Without Identified Device or Use Problem1
- Material Deformation1
Patient problems
- Impaired Healing10
- Insufficient Information4
- Bruxism1
- No Clinical Signs, Symptoms or Conditions1
- Twiddlers Syndrome1
- Unspecified Infection1
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
InjuryMay 20, 2026Fracture
IT WAS REPORTED THAT A PLATE WAS IDENTIFIED AS FRACTURED, DURING A PATIENT FOLLOW UP APPOINTMENT. IT WAS REMOVED AND REPLACED.
Patient problems: Insufficient Information
MAUDE report 25234211 ↗InjuryFeb 27, 2026Fracture
IT WAS REPORTED THAT A PLATE FRACTURED APPROXIMATELY FIVE MONTHS AFTER IMPLANTATION. A PORTION OF IT REMAINS IMPLANTED.
Patient problems: Insufficient Information
MAUDE report 24469037 ↗InjurySep 12, 2025Fracture; Fracture
IT WAS REPORTED A PLATE BECAME FRACTURED DUE TO THE PATIENT GRINDING THEIR TEETH. IT WAS REMOVED AND NOT REPLACED.
Patient problems: Bruxism
MAUDE report 23037854 ↗InjuryApr 30, 2025Fracture; Fracture
IT WAS REPORTED FOLLOWING APPROXIMATELY 1 YEAR AND 8 MONTHS OF IMPLANTATION A CMF PLATE WAS FOUND TO HAVE BECOME FRACTURED. IT WAS REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21938474 ↗InjuryOct 26, 2022Break; Break
IT WAS REPORTED THAT A PLATE BROKE A PORTION WAS REMOVED.
Patient problems: Impaired Healing
MAUDE report 15676906 ↗InjuryMay 21, 2022Material Deformation; Material Deformation
IT WAS REPORTED THAT A PLATE BECAME DEFORMED AFTER SUCCESSFUL CONSOLIDATION OF UNDERLYING BONE. THE PLATE WAS REMOVED.
Patient problems: Insufficient Information
MAUDE report 14465907 ↗InjuryMar 18, 2022Break; Break
IT WAS REPORTED THAT A PLATE BROKE. IT WAS REMOVED AND REPLACED.
Patient problems: Impaired Healing
MAUDE report 13804744 ↗InjuryOct 26, 2021Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PLATE WAS FOUND FRACTURED AFTER THE BONE WAS CONSOLIDATED. A PORTION OF THE FRACTURED PLATE WAS REMOVED.
Patient problems: Insufficient Information
MAUDE report 12698909 ↗InjuryMay 28, 2021Break; Break
IT WAS REPORTED A PLATE BROKE AND A PORTION OF IT WAS REMOVED.
Patient problems: Impaired Healing
MAUDE report 11901237 ↗InjuryJul 29, 2020Break; Break
IT WAS REPORTED A PLATE BROKE DUE TO POSSIBLE PATIENT NONCOMPLIANCE AND WAS REMOVED.
Patient problems: Impaired Healing
MAUDE report 10341287 ↗InjuryMar 26, 2020Break; Break
IT WAS REPORTED THAT A PLATE BROKE AND WAS REMOVED FROM A PATIENT WITH SECONDARY OSTEORADIONECROSIS.
Patient problems: Impaired Healing
MAUDE report 9885080 ↗InjuryDec 19, 2019Break; Break
IT WAS REPORTED THE PLATE BROKE. IT WAS REMOVED AND REPLACED.
Patient problems: Impaired Healing
MAUDE report 9498390 ↗InjuryNov 8, 2019Break; Break
IT WAS REPORTED THAT THE PRODUCT WAS REMOVED DUE TO PATIENT CONDITION.
Patient problems: Unspecified Infection
MAUDE report 9298660 ↗InjuryMay 23, 2019Break; Break
IT WAS REPORTED THE PLATE BROKE. IT WAS REMOVED AND REPLACED.
Patient problems: Impaired Healing
MAUDE report 8635477 ↗InjuryApr 9, 2019Break; Break
IT WAS REPORTED THAT A PLATE BROKE. IT WAS REMOVED AND REPLACED.
Patient problems: Impaired Healing
MAUDE report 8496401 ↗InjuryJul 13, 2018Break; Break
IT WAS REPORTED A RECONSTRUCTION PLATE BROKE. DEVICE WILL BE REMOVED AT A LATER DATE.
Patient problems: Impaired Healing
MAUDE report 7686252 ↗MalfunctionJun 29, 2018Break; Break
IT WAS REPORTED THE PATIENT DEVELOPED OSTEONECROSIS AND SOME POINT THEREAFTER THE PLATE BROKE.
Patient problems: Impaired Healing
MAUDE report 7651571 ↗InjuryDec 20, 2016Break
PLATE BROKE INSIDE PATIENTS MOUTH. PATIENT CLAIMS TO HAVE BEEN MOVING HER JAW FROM SIDE TO SIDE A LOT AND WAS DOING THIS AT THE TIME THE PLATE BROKE. THE BREAK WAS AT NON UNION BETWEEN FIBULA FREE FLAP AND NATIVE BONE.
Patient problems: Twiddlers Syndrome
MAUDE report 6191459 ↗
FDA clearance and classification
The FDA regulates product code MQN (external mandibular fixator and/or distractor) under 21 CFR 872.4760 as a Class II device. 21 CFR 872.4760 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the LEVEL ONE CMF 50-760-26-09
Has the LEVEL ONE CMF 50-760-26-09 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00888118078216), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the LEVEL ONE CMF 50-760-26-09?
The FDA's MAUDE database lists 18 adverse event reports for this device, received between Dec 20, 2016 and May 20, 2026. In those reports, the most common specific device problems are “Break” (13), “Fracture” (4) and “Material Deformation” (1); the most common patient problems are “Impaired Healing” (10), “Bruxism” (1) and “Twiddlers Syndrome” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the LEVEL ONE CMF 50-760-26-09?
None of the reports are classified as deaths, 17 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the LEVEL ONE CMF 50-760-26-09 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the LEVEL ONE CMF 50-760-26-09, and is it still sold?
The FDA's device database lists KLS-Martin L.P. as the company responsible for it (brand LEVEL ONE CMF, model 50-760-26-09). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR devices
- DISTRACTION, INTERNAL 51-405-42-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR20 reports
- LEVEL ONE CMF 50-797-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR13 reports
- DISTRACTION, INTERNAL 51-563-30-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR13 reports
- LEVEL ONE CMF 50-777-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR11 reports
- NA 04.315.125Synthes GmbH · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- DISTRACTION, INTERNAL 51-526-30-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- DISTRACTION, INTERNAL 51-401-91-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR10 reports
- LEVEL ONE CMF 50-795-26-09KLS-Martin L.P. · EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR9 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page