Skip to content

DISTRACTION, INTERNAL 51-526-30-09Recalls & reported problems

KLS-Martin L.P. · DIST, MANDIBLE, HYPERDRIVE, RTCH, 1.5-1.8 MM SCREW, TI-6AL-4V · Class 2 · Implant

All models of DISTRACTION, INTERNAL (67)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00888118099617), received between Mar 16, 2017 and Jun 12, 2025 — classified as 10 injuries.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Detachment of Device or Device Component2
  • Detachment Of Device Component1
  • Difficult or Delayed Activation1
  • Entrapment of Device1
  • Fracture1

Patient problems

  • Impaired Healing5
  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 12, 2025Fracture; Fracture

    IT WAS REPORTED THAT A DISTRACTOR BECAME FRACTURED FOLLOWING 15 DAYS OF IMPLANTATION. IT WAS REMOVED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22193871 ↗
  • InjuryFeb 21, 2024Break; Break

    IT WAS REPORTED A DISTRACTOR BROKE. IT WAS REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 18748205 ↗
  • InjuryDec 8, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DEVICE ATTACHMENT SOCKET BECAME SEPARATED FROM RATCHET HOUSING.

    Patient problems: Insufficient Information

    MAUDE report 12964308 ↗
  • InjuryDec 1, 2021Break; Break

    A PATIENT CONDITION POSSIBLY CONTRIBUTED TO A DISTRACTOR BREAKAGE, IT WAS REMOVED AND REPLACED.

    Patient problems: Impaired Healing

    MAUDE report 12908173 ↗
  • InjuryNov 3, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DEVICE ATTACHMENT SOCKET BECAME SEPARATED FROM RATCHET HOUSING.

    Patient problems: Insufficient Information

    MAUDE report 12749003 ↗
  • InjuryJul 19, 2021Break; Break

    IT WAS REPORTED THAT A DISTRACTOR BROKE. IT WAS REMOVED.

    Patient problems: Impaired Healing

    MAUDE report 12192965 ↗
  • InjuryMay 11, 2020Entrapment of Device; Difficult or Delayed Activation; Entrapment of Device; Difficult or Delayed Activation

    IT WAS REPORTED THE DISTRACTOR WOULD NOT TURN. DOCTOR'S OPINION IS MAYBE TISSUE WAS STUCK IN THE TRACK. THE PRODUCT WAS REMOVED PRIOR TO COMPLETION OF DISTRACTION.

    Patient problems: Impaired Healing

    MAUDE report 10042901 ↗
  • InjurySep 12, 2019Break; Break

    IT WAS REPORTED A DISTRACTOR BROKE AND WAS REMOVED.

    Patient problems: Impaired Healing

    MAUDE report 9007983 ↗
  • InjuryAug 24, 2018Break; Break

    IT WAS REPORTED THAT A DISTRACTOR BROKE DURING IMPLANTATION. IT WAS REMOVED AND REPLACED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7812535 ↗
  • InjuryMar 16, 2017Break; Detachment Of Device Component

    IT WAS REPORTED THE DISTRACTOR WAS OPEN TO ABOUT 25MM AND WAS USED TO RETRACT THE PATIENTS SKULL. AFTER ABOUT A WEEK THE END RATCHET SNAPPED OFF AND THE DEVICE WAS NOT ABLE TO RETRACT ANY FURTHER. IT IS BELIEVED TISSUE AND OR BONE GOT CAUGHT IN THE TRACK, WHICH CAUSED THE DEVICE TO BREAK. DEVICE WAS REMOVED ON (B)(6) 2017. DISTRACTOR WAS USED AS A RETRACTOR. THE USE OF THE DEVICE IN THIS MANNER IS CONSIDERED OFF INDICATIONS FOR USE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6411143 ↗

Other EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR devices

All EXTERNAL MANDIBULAR FIXATOR AND/OR DISTRACTOR devices →

Sources and updates