Skip to content

LEVEL ONE NEURO 99-576-76-07Recalls & reported problems

KLS-Martin L.P. · SCREW, ONEDRIVE, DRILL FREE, V2, TI-6AL-4V · Class 2 · Implant

All models of LEVEL ONE NEURO (27)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888118101297), received between Mar 12, 2018 and Nov 26, 2018 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Break1

Patient problems

  • Cerebrospinal Fluid Leakage1
  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 26, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED DEVICES WERE REMOVED DUE TO PATIENT CONDITION.

    Patient problems: Cerebrospinal Fluid Leakage

    MAUDE report 8103558 ↗
  • InjuryMar 12, 2018Break; Break

    IT WAS REPORTED THAT A SCREW HEAD BROKE OFF DURING IMPLANTATION. THE THREADS OF THE SCREW REMAIN IN THE PATIENT. IFU (B)(4) STATES "CARE IS TO BE USED WHEN SEATING MINI OR MICRO CRANIO-MAXILLO BONE SCREWS. INTRAOPERATIVE FRACTURE OF SCREWS CAN OCCUR IF EXCESSIVE FORCE (TORQUE) OR MECHANICAL OVERLOAD IS APPLIED WHILE SEATING BONE SCREWS INTO POSITION."

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7330126 ↗