Skip to content

Medline MDS9698LRecalls & reported problems

Medline Industries, Inc. · THERMOMETER,TEMPLE,6-10 SECOND · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00888277409937), received between Sep 22, 2026 and Sep 30, 2026 — classified as 13 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 13 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Computer Software Problem8
  • Unable to Obtain Readings4
  • Calibration Problem1
  • Display or Visual Feedback Problem1
  • Incorrect Measurement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions13

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Computer Software Problem

    It was reported that "We have 2 that have already stopped working giving us an error 1 message. When reviewing the user guide, this is due to it being outside of the normal temperature."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948247 ↗
  • MalfunctionSep 30, 2026Computer Software Problem

    It was reported that "THEY HAVE 2 THERMOMETERS THAT CONSISTENLY ARE RECEIVING ERRORS THEY HAVE TRIED CHANGING BATTERIES BUT STILL GET THE SAME ERROR."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946598 ↗
  • MalfunctionSep 30, 2026Computer Software Problem

    This VMSR report summarizes 8 malfunction events that occurred between April 1 and June 30, 2023. The account reported that temple thermometers (MDS9698L) displayed Error 1 codes during operation. The customer reported that 8 thermometers had exhibited the issue over a period of several months. No patient information was available. No patient or user injuries were reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946482 ↗
  • MalfunctionSep 28, 2026Incorrect Measurement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918320 ↗
  • MalfunctionSep 28, 2026Computer Software Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917035 ↗
  • MalfunctionSep 28, 2026Display or Visual Feedback Problem

    It was reported that they do not work. every time we go to use them they error out.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50913218 ↗
  • MalfunctionSep 25, 2026Unable to Obtain Readings

    It was reported that "Customer is getting an error message of Error 1 which would indicate that this item is not at room temp. Customer believe this is defective as she has another one of this and has not recevived this error message before".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50898547 ↗
  • MalfunctionSep 23, 2026Computer Software Problem; Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50874941 ↗
  • MalfunctionSep 23, 2026Computer Software Problem; Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873920 ↗
  • MalfunctionSep 23, 2026Computer Software Problem

    It was reported that SCREEN IS DISPLAYING ERROR 1 MESSAGE IN RED BACKGROUND, CUSTOMER TRIED TROUBLESHOOTING WITH MANUAL AND IT STILL GIVES ERROR MESSAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50871857 ↗
  • MalfunctionSep 23, 2026Calibration Problem

    It was reported that "THEY HAVE 1 THERMOMETER THAT CONSISTENLY IS RECEIVING ERRORS THEY HAVE TRIED CHANGING BATTERIES BUT STILL GET THE SAME ERROR."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50869618 ↗
  • MalfunctionSep 23, 2026Computer Software Problem

    It was reported that both of them give her a red error message. The second one started doing this within 2 weeks and she cannot use it.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50867203 ↗
  • MalfunctionSep 22, 2026Unable to Obtain Readings

    It was reported that NONE OF THESE ARE REGISTERING, NOT GIVING ANY SORT OF READING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50854299 ↗

Other Thermometer, electronic, clinical devices

All Thermometer, electronic, clinical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page