Why it was recalled

It has been determined that certain Blood Pressure Monitors containing a faulty microchip have failed to power on or have shown display screen issues after the initial few uses, as well as overheating at times making the unit warm to the touch.

FDA root cause category: Under Investigation by firm

What the company is doing

Medline issued a MEDICAL DEVICE RECALL notice to its consignees on 08/30/2024 via email and USPS first class mail. The notice explained the problem, risk to user and requested that units demonstrating the failure be discarded. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected]. Medline issued a follow up notification to distributors on 09/13/2024. The notification contained the same information as the initial letter, but also instructed distributors to notify customers if the product was further distributed.

Recall details

Recall number
Z-0032-2025
Status
Open, Classified
Initiated by the company
Aug 30, 2024
Posted by the FDA
Oct 8, 2024
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
52,521 units
Product code
DXN
FDA submissions
K193627
Distribution
Worldwide distribution - US Nationwide and the countries of Panama and Jamaica.

Product description

MEDLINE AUTOMATIC DIGITAL BLOOD PRESSURE UNIT, REF MDS1001

Code information (models, lots, serial numbers)

GTIN 00888277629656, Lot Numbers: L230380005, L230480005, L230580005, L230680005, L230780005, L230980005, L231080005, L231180005, L231280005, L240280005