Skip to content

HeartWare® Battery 1650DE (UDI 00888707005708)Recalls & reported problems

Heartware Inc. · Class 3 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This HeartWare battery powers a heart pump used to support failing hearts. Nearly all of its 487 reports describe battery problems, many involving batteries that would not charge. Most patients had no clinical signs, and the reports include 1 injury and no deaths. One open recall from June 2022 covers battery faults that can stop it from charging or powering the controller. Reports fell from 27 the year before to none in the latest year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
487
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 487 adverse event reports that list this exact device (DI 00888707005708), received between Dec 6, 2018 and Jun 25, 2025 — classified as 1 injury, 486 malfunctions.

The most frequently reported specific device problem is “Battery Problem” (445 reports).

1 FDA recall record name this device's identifier, 1 still open.

59 other recalls cover devices cleared under the same FDA submission (P100047); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 27 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202627
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 28, 2022Open, Classified

    Battery performance issues. Battery electrical faults render it unable to power the controller, unable to accept charge from the battery charger, and/or result in the battery to appear to remain charged while discharging.

    Z-1512-2022 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Battery Problem445
  • Material Integrity Problem23
  • Connection Problem17
  • Break2
  • Display or Visual Feedback Problem2
  • Power Problem2
  • Complete Loss of Power1
  • Crack1

Patient problems

  • No Clinical Signs, Symptoms or Conditions362
  • No Known Impact Or Consequence To Patient90
  • No Consequences Or Impact To Patient61
  • No Patient Involvement4
  • Anxiety2
  • Discomfort1

Recent adverse event reports

The 25 most recent of 487 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WAS NOT CHARGING. THE BATTERY WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330475 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WAS NOT CHARGING. THE BATTERY WILL BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330473 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WAS NOT CHARGING. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330466 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WAS NOT CHARGING. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330461 ↗
  • MalfunctionJun 25, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT LEDS ON THE BATTERY WERE NOT WORKING AS INTENDED. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330459 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WAS NOT CHARGING. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330458 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT WHEN THE BATTERY WAS CONNECTED TO THE BATTERY CHARGER, THE INDICATOR LIGHT SHOWED A FLASHING RED LIGHT. ALL FOUR PORTS OF THE BATTERY CHARGER WERE TRIED, AND THERE WAS ALWAYS A FLASHING RED LIGHT. THE BATTERY WAS LEFT CONNECTED, BUT THE BATTERY NEVER CHARGED. THE BATTERY WILL BE TAKEN OUT OF SERVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330457 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY DID NOT CHARGE AS INTENDED ON ANY PORT OF THE BATTERY CHARGER. THE BATTERY WAS PLACED ON CHARGER OVERNIGHT, HOWEVER WAS NOT FULLY CHARGED. THE BATTERY WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330456 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WAS NOT CHARGING. THE BATTERY WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330455 ↗
  • MalfunctionJun 25, 2025Power Problem; Power Problem

    IT WAS REPORTED THAT THE BATTERY EXHIBITED A RED LIGHT EMITTING DIODE ON THE BATTERY CHARGE CAPACITY DISPLAY. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330452 ↗
  • MalfunctionJun 25, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE LIGHT EMITTING DIODE (LED) ON THE BATTERY WAS NOT WORKING PROPERLY. THE BATTERY WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330451 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WOULD NOT CHARGE. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330449 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WAS NOT CHARGING AND PROVIDED RED LIGHT ON THE CHARGER. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330448 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WAS NOT CHARGING. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330447 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WAS NOT CHARGING. THE BATTERY WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330446 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WOULD NOT CHARGE AND A RED LIGHT WOULD DISPLAY ON THE CHARGER. THE BATTERY WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330445 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY EXHIBITED AN INABILITY TO CHARGE. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330444 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WAS NOT CHARGING AND SHOWED A RED LIGHT. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330443 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WAS NOT CHARGING ON ANY BATTERY CHARGER PORTS. THE PATIENT NOTED THE BATTERY WOULD APPEAR TO BE CHARGING WITH A YELLOW INDICATOR LIGHT BUT WOULD NOT CHARGE BEYOND ONE BAR ON THE POWER INDICATOR. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330440 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WAS NOT CHARGING. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330439 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT, WHEN CONNECTED TO THE BATTERY CHARGER, THE BATTERIES WOULD NOT CHARGE AND ONE OF THE BATTERY CHARGER STATUS LIGHTS FLASHED RED. THE BATTERIES WERE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330438 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WOULD NOT CHARGE "PAST TWO BARS ANYMORE" WHEN TESTED ON MULTIPLE CHARGER PORTS. THE BATTERY WAS EX CHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330434 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT WHEN THE BATTERY WAS CONNECTED TO THE BATTERY CHARGER THERE WAS A RED FLASHING INDICATOR LIGHT. THE BATTERY WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330433 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WOULD NOT CHARGE ON ANY OF THE CHARGER PORTS AND IT DISPLAYED A RED LIGHT WHEN ATTACHED. THE BATTERY WAS EXCHANGED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330430 ↗
  • MalfunctionJun 25, 2025Battery Problem; Battery Problem

    IT WAS REPORTED THAT THE BATTERY WAS NOT CHARGING. THE BATTERY WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22330429 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P100047. One submission often covers many models, so check the product and lot details in each record.

  • Z-0475-2024 · Oct 31, 2023 · Open, Classified
    HVAD (HeartWare Ventricular Assist Device) STERILE IMPLANT KIT, Model Number 1100
  • Z-1993-2023 · May 16, 2023 · Open, Classified
    HeartWare Ventricular Assist Device System. a) Monitor 1521US Explore Tech, b) Monitor 1521GB Explore Tech, c) Monitor 1521IL Explore Tech, d) Monitor 1521DE Explore Tech
  • Z-1266-2022 · May 5, 2022 · Open, Classified
    HeartWare Battery, Model #1650DE, a component of the HeartWare Ventricular Assist Device (HVAD) System.
  • Z-1207-2022 · Apr 14, 2022 · Open, Classified
    HeartWare HVAD Pump Implant Kit, a) Model Number 1104JP, b) Model Number MCS1705PU
  • Z-1103-2022 · Mar 30, 2022 · Open, Classified
    HeartWare Ventricular Assist Device (HVAD) System Controller, Model Number 1407DE Kit
  • Z-1107-2022 · Mar 30, 2022 · Open, Classified
    HeartWare Ventricular Assist Device (HVAD) System, individual pamphlets: HVAD System Emergency Responder Guide (FI-Finnish) HVAD System Emergency Responder Guide (FI-Finnish) H
  • Z-1104-2022 · Mar 30, 2022 · Open, Classified
    HeartWare Ventricular Assist Device (HVAD) System Controller, Model Number 1420
  • Z-1105-2022 · Mar 30, 2022 · Open, Classified
    HeartWare Ventricular Assist Device (HVAD) System Controller, Model Number 1420-Controller 2.0
  • Z-1106-2022 · Mar 30, 2022 · Open, Classified
    HeartWare Ventricular Assist Device (HVAD) System, Model Number 1521DE, MONITOR 1521DE EXPLORE TECH
  • Z-1110-2022 · Mar 30, 2022 · Open, Classified
    Instructions for Use, Patient Manuals, and Emergency Responder Guides for HeartWare Ventricular Assist Device (HVAD) System. HVAD System. IFUs are distributed as part of the Pump
  • Z-1102-2022 · Mar 30, 2022 · Open, Classified
    HeartWare Ventricular Assist Device (HVAD) System, Pump Implant Kit, Model Number1104
  • Z-0726-2022 · Jan 27, 2022 · Open, Classified
    HeartWare Ventricular Assist Device (HVAD) System
  • Z-2478-2021 · Aug 6, 2021 · Open, Classified
    Medtronic HVAD Pump Implant Kits
  • Z-2188-2021 · Jun 3, 2021 · Open, Classified
    HeartWare HVAD Pump Implant Kit, REF 1103
  • Z-2190-2021 · Jun 3, 2021 · Open, Classified
    HeartWare HVAD Pump Implant Kit, REF 1125

…and 44 more.

Other Ventricular (Assist) Bypass devices

All Ventricular (Assist) Bypass devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.