Skip to content

HeartWare BatteryRecalls & reported problems across all models

Heartware Inc. · 11 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

5 open recalls name at least one HeartWare Battery model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
4,241
Deaths reported
0
Injuries reported
16
Recalls
5

We found 11 models of HeartWare Battery by Heartware Inc. in the FDA's device registry (GUDID). Together they are named in 4,241 adverse event reports in the FDA's MAUDE database, received between Jan 12, 2016 and Sep 30, 2026 — classified as 16 injuries, 4,225 malfunctions.

The most frequently reported specific device problem is “Battery Problem” (3,920 reports).

5 FDA recall records name at least one of these models by its device identifier, 5 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 4 reportsJan 2024: 11 reportsFeb 2024: 7 reportsMar 2024: 8 reportsApr 2024: 11 reportsMay 2024: 5 reportsJun 2024: 6 reportsJul 2024: 4 reportsAug 2024: 5 reportsSep 2024: 4 reportsOct 2024: 7 reportsNov 2024: 6 reportsDec 2024: 1 reportsJan 2025: 4 reportsFeb 2025: 6 reportsMar 2025: 7 reportsApr 2025: 2 reportsMay 2025: 2 reportsJun 2025: 44 reportsJul 2025: 6 reportsAug 2025: 5 reportsSep 2025: 2 reportsOct 2025: 2 reportsNov 2025: 2 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 2 reportsOct 2026: 0 reports20242025202644
Last 36 months. The most recent month may be incomplete.

HeartWare Battery recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 28, 2022Open, Classified

    Battery performance issues. Battery electrical faults render it unable to power the controller, unable to accept charge from the battery charger, and/or result in the battery to appear to remain charged while discharging.

    Z-1512-2022 · Root cause: Device Design · names 6 of these models · Recall details · FDA record ↗

  2. May 5, 2022Open, Classified

    Batteries for the HVAD system have a weld defect affecting internal components within the HVAD battery, resulting in the battery to malfunction and no longer provide power or prevent the battery from holding a complete charge or properly recharging.

    Z-1266-2022 · Root cause: Process control · Recall details · FDA record ↗

  3. Feb 26, 2021Open, Classified

    IFU and Patient Manual updated to describe HVAD Controller Connection Verification. The labeling is being corrected to reflect the actual verified behavior of a successful connector to the controller, which is a "beep" and not "audible click", to clarify under what alarm conditions a beep will not be heard, and identify alternate confirmation methods (visual indicator) for successful connection. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87603

    Z-1438-2021 · Root cause: Device Design · Recall details · FDA record ↗

  4. Feb 26, 2021Open, Classified

    IFU and Patient Manual updated to describe Expected Alarm Duration. The Instructions for Use (IFU) is being updated to advise clinicians that alarms that resolve after a very short period may mean there is an intermittent problem and the cause of the alarm can be evaluated by analysis of the controller log files. The Patient Manual (PM) is being updated to instruct patients to contact their clinicians if they cannot identify an alarm before it is cleared as it might mean there is an intermittent problem. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87604

    Z-1469-2021 · Root cause: Device Design · Recall details · FDA record ↗

  5. May 2, 2018Open, Classified

    Possible transient electrical connection interruption between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the Controller, while the power source is connected, that may cause unintended switching to the secondary power source and/or unexpected audible tones (beeping). The interruption typically lasts 1-2 seconds, due to oxidation between power source connector and socket.

    Z-1903-2018 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

Most reported HeartWare Battery problems

Device problems

  • Battery Problem3,920
  • Connection Problem166
  • Material Integrity Problem150
  • Power Problem94
  • Display or Visual Feedback Problem27
  • Material Deformation19
  • Failure to Charge14
  • Break13

Patient problems

  • No Clinical Signs, Symptoms or Conditions2,191
  • No Known Impact Or Consequence To Patient1,543
  • No Consequences Or Impact To Patient559
  • No Patient Involvement35
  • Anxiety16
  • Discomfort2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Bacterial Infection1

HeartWare Battery models

11 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent HeartWare Battery adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Heartware Inc. products

All Heartware Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page