Skip to content

Arthrex® AR-11210Recalls & reported problems

Arthrex, Inc. · PUNCH, LARGE 2.75MM, 15 DEG UP CURVED · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867002968), received between Mar 2, 2021 and May 13, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Component Missing” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Component Missing2
  • Difficult to Open or Close2
  • Break1
  • Dull, Blunt1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 13, 2024Break; Break

    ON 04/17/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA SEMS-06541976 THAT AN AR-11210 LARGE PUNCH LEFT A PIECE OF METAL IN THE PATIENT. THIS OCCURRED DURING A PROCEDURE WHERE THE METAL WAS TAKEN OUT, AND THE PATIENT WAS OKAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19297522 ↗
  • MalfunctionApr 13, 2021Component Missing; Dull, Blunt; Difficult to Open or Close

    THIS IS A CONVERSION FROM CC183280, LINE 281969. AN CUSTOMER UPDATE WAS RECEIVED WHICH MADE A REPAIR TO A COMPLAINT. THE CUSTOMER STATED THE INSTRUMENT DOES NOT CLOSE CORRECTLY AND IS BLUNT. NO ADDITIONAL INFORMATION AS TO WHAT FAILURE MODE OCCURRED WAS PROVIDED. THERE WAS NO HARM OR ADVERSE EVENT FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED BY THE CUSTOMER. UPDATE 08-APR-2021: FURTHER INFORMATION WERE PROVIDED THAT INTRA-OPERATIVELY IT WAS FOUND THAT THE INSTRUMENT DID NOT GRIP. THE INSTRUMENT WAS REMOVED FROM THE OPERATING AREA, CLEANED AND SENT FOR REPAIR. THE CUSTOMER REPORTED THAT THE MISSING PIN WAS NOT NOTICED EITHER IN THE OPERATING DEPARTMENT OR IN THE STERILIZATION DEPARTMENT. THIS IS NOW REPORTABLE. DURING RETURNED DEVICE EVALUATION, A REPORTABLE MALFUNCTION WAS DISCOVERED. DEVICE HAS A MISSING TIP PIN. THE PRECISE CAUSE FOR A MISSING TIP PIN COULD NOT BE DETERMINED. TIP PIN WAS NOT RETURNED OR ACCOUNTED FOR DURING EVALUATION/INVESTIGATION PROCESS. FUNCTION TEST COULD NOT BE PERFORMED DUE TO THE RELATED CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11655796 ↗
  • MalfunctionMar 2, 2021Component Missing; Difficult to Open or Close

    IT WAS REPORTED, THAT DURING KNEE PROCEDURE JAW ON THE AR-11210, WOULD NOT OPEN. THE CASE WAS COMPLETES USING A DIFFERENT PUNCH. ADDITIONAL INFORMATION 2/10/2021 PER THE SALES REP- AR-11210 / BATCH 82412 ¿ DEVICE HAS A MISSING TIP PIN ON THE DISTAL END OF THE TIP (UNDETERMINED). ADDITIONAL INFORMATION 2/11/2021 PER THE SALES REP THE PIN CAME OFF AFTER WE REMOVED THE INSTRUMENT FROM THE KNEE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11398295 ↗

Other Accessories, arthroscopic devices

All Accessories, arthroscopic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page