Skip to content

Arthrex® AR-11910NRRecalls & reported problems

Arthrex, Inc. · GRSPR,MINI,2.75 MM 15°UP CRV W/NR HNDL · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867003842), received between Jun 21, 2017 and Jun 21, 2017 — classified as 1 injury.

The most frequently reported specific device problem is “Component Missing” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Component Missing1
  • Device Or Device Fragments Location Unknown1
  • Difficult to Open or Close1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 21, 2017Component Missing; Device Or Device Fragments Location Unknown; Difficult to Open or Close

    IT WAS REPORTED VIA MEDWATCH (B)(4) THAT DURING A NON-UNION OF BONE AFTER ORIF (OPEN REDUCTION INTERNAL FIXATION) OF ANKLE A SCREW OR PIN FROM THE 2.75 MM 15° CURVED MINI GRASPER WAS MISSING. DURING THE PROCEDURE THE SURGEON WAS ATTEMPTING TO OPEN THE GRASPER IN THE PATIENT'S ANKLE JOINT, BUT THE JAWS OF THE INSTRUMENT WOULD NOT OPEN. THE INSTRUMENT WAS TAGGED AND SENT TO SPD (STERILE PROCESSING DEPARTMENT). AFTER SPD CLEANED AND INSPECTED THE INSTRUMENT, IT WAS FOUND THAT EITHER A SET PIN OR TINY SCREW WAS MISSING FROM THE JAWS OF THE INSTRUMENT. THE PATIENT HAD AN X-RAY PERFORMED AT THE NEXT VISIT AND NO METAL OR RETAINED FOREIGN OBJECT WAS NOTED. UNCLEAR WHAT HAPPENED WITH THE SCREW.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6656256 ↗

Other Accessories, arthroscopic devices

All Accessories, arthroscopic devices →