Arthrex® AR-1204AF-105Recalls & reported problems

Arthrex, Inc. · 10.5 MM FLIP CUTTER II · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00888867004122), received between Jun 26, 2019 and Sep 19, 2025 — classified as 1 injury, 9 malfunctions.

The most frequently reported specific device problem is “Break” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break9
  • Material Disintegration2
  • Difficult to Remove1
  • Material Separation1
  • Output Problem1

Patient problems

  • Foreign Body In Patient5
  • No Clinical Signs, Symptoms or Conditions3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Insufficient Information2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 19, 2025Break; Break

    ON 2ND SEP 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-1204AF-105, FLIPCUTTER® II, 10.5 MM, THE TIP BROKE WHILE DRILLING. THIS WAS DETECTED DURING A PCL RECONSTRUCTION WITH ALLOGRAFT AND INTERNAL BRACE ON (B)(6) 2025. THE CASE WAS COMPLETED SUCCESSFULLY BY USING A NEW AR-1204AF-105. THE TIP WAS BREAK OFF INSIDE THE PATIENT AND THE BROKEN PIECE WAS NOT RETRIEVED FROM THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 23094975 ↗
  • MalfunctionFeb 19, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ACL SURGERY THE FLIPCUTTER 10.5 HAS BEEN USED AND INSTANTLY BROKE WHEN THEY HAVE STARTED DRILLING. THE BROKEN OFF PIECE HAS BEEN RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21412036 ↗
  • MalfunctionFeb 10, 2025Break; Break

    ON 01/16/2025, IT WAS REPORTED BY A DISTRIBUTOR VIA (B)(4) THAT AN AR-1204AF-105 FLIPCUTTERS TIP BROKE. THIS OCCURRED DURING AN ACL RECONSTRUCTION WHERE THE RETROGRADE DRILL ENTERED THE FEMORAL CONDYLE 5 MM AND BROKE AT ITS TIP. THERE WAS NO ADDITIONAL INFORMATION PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: Insufficient Information

    MAUDE report 21343635 ↗
  • MalfunctionSep 19, 2022Break; Break; Material Disintegration

    THE SURGEON HAD THREE FLIPCUTTER II DISINTEGRATE IN THE KNEE AS DRILLING FEMORAL TUNNEL. THE FIRST TIME THEY STARTED DRILLING WITH BLADE ON BONE AND THE REP INSTRUCTED THEM TO START THE DRILLING AT FULL SPEED BEFORE PULLING BACK ONTO THE BONE. BUT THE SAME THING HAPPENED ON THE SECOND AND THE THIRD TIME. THE PATIENT WAS A 17 YEAR OLD MALE. ON ALL THREE OCCASIONS, THE METAL PIECES OF THE DEVICE BROKE OFF AND NEEDED TO BE RETRIEVED BY THE SURGEON.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Foreign Body In Patient

    MAUDE report 15445916 ↗
  • MalfunctionSep 19, 2022Break; Break; Material Disintegration

    THE SURGEON HAD THREE FLIPCUTTER II DISINTEGRATE IN THE KNEE AS DRILLING FEMORAL TUNNEL. THE FIRST TIME THEY STARTED DRILLING WITH BLADE ON BONE AND THE REP INSTRUCTED THEM TO START THE DRILLING AT FULL SPEED BEFORE PULLING BACK ONTO THE BONE. BUT THE SAME THING HAPPENED ON THE SECOND AND THE THIRD TIME. THE PATIENT WAS A 17YEAR OLD MALE. ON ALL THREE OCCASIONS, THE METAL PIECES OF THE DEVICE BROKE OFF AND NEEDED TO BE RETRIEVED BY THE SURGEON.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Foreign Body In Patient

    MAUDE report 15445904 ↗
  • MalfunctionJun 6, 2022Break

    IT WAS REPORTED THAT DURING A CRUCIATE LIGAMENT RESTORATION THE DEVICES AR-1204AF-105 (LOT: 1335117583) BROKE DURING DRILLING INWARD INTO THE JOINT. THE BREAKAGE POINT WAS OUTSIDE OF THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14603537 ↗
  • MalfunctionDec 27, 2021Break; Output Problem

    IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY SURGERY THE MECHANISM OF THE DEVICE DID NOT WORK. WHEN THE SURGEON USED THE FLIPCUTTER IN THE PATIENT'S KNEE, THE SPINNER AT THE END OF THE FLIPCUTTER OPENED BUT COULD NOT BE CLOSED TO PULL THE METAL END OUT. WHAT TURNS AT THE END OF THE FLIPCUTTER DID NOT TURN WELL AND ALMOST GOT STUCK INSIDE THE KNEE. PART OF THE DIVIDE BROKE OFF INSIDE THE PATIENT AND THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE 17-DEC-2021: IT WAS CONFIRMED THAT A PART OF THE MECHANISM BROKE OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13087950 ↗
  • MalfunctionNov 2, 2021Material Separation

    ON (B)(6) 2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-1204AF-105 FLIPCUTTER II BROKE. THIS OCCURRED DURING A REVISION ACL RECONSTRUCTION PROCEDURE ON (B)(6) 2021. SURGEON WAS DRILLING THE FEMORAL TUNNEL, MAINLY THROUGH SOFT TISSUE AND SOME BONE, WHEN THE TIP OF THE FLIPCUTTER BROKE INSIDE PATIENT. THE TIP WAS REMOVED USING A GRASPER AND A NEW FLIPCUTTER DEVICE WAS OPENED WITH NO MORE ISSUES.

    Patient problems: Foreign Body In Patient

    MAUDE report 12737840 ↗
  • MalfunctionApr 1, 2021Break

    IT WAS REPORTED THAT DURING AN ACL PROCEDURE, AT THE MOMENT THAT THE AR-1204AF-105, MAKES CONTACT WITH THE TIBIAL BONE, THE TIP OF THE AR-1204AF-105, BROKE WITHIN THE JOINT SPACE. THE FRAGMENT WAS REMOVED AND THE CASE WAS COMPLETED BY USING A NEW AR-1204AF-105 WITHOUT FURTHER ISSUE.

    Patient problems: Insufficient Information

    MAUDE report 11602653 ↗
  • InjuryJun 26, 2019Break; Difficult to Remove

    IT WAS REPORTED THAT PATIENT UNDERWENT AN ACL PROCEDURE ON (B)(6) 2019 WHERE A FLIPCUTTER, AR-1204AF-105, WAS USED. DURING RETRO-DRILLING OF THE FEMORAL TUNNEL, THE CUTTER BROKE OFF. THE CUTTER WAS UNABLE TO BE RETRIEVED AND THE PATIENT WAS SCHEDULED FOR A SECOND SURGERY TO HAVE THE CUTTER REMOVED. THE PATIENT UNDERWENT A SECOND SURGERY ON (B)(6) 2019 AND HAD THE CUTTER REMOVED WITH NO INCIDENT. ADDITIONAL INFORMATION REQUESTED. ADDITIONAL INFORMATION PROVIDED (B)(6) 2019: THE FACILITY DISCARDED BOTH PARTS OF THE FLIPCUTTER AFTER THE CASE AND NO X-RAYS WERE TAKEN.

    Patient problems: Foreign Body In Patient

    MAUDE report 8734620 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page