Arthrex® AR-1204AF-60Recalls & reported problems
Arthrex, Inc. · 6.0 MM FLIP CUTTER II · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867004207), received between Jan 31, 2022 and Jan 31, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Dull, Blunt” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Dull, Blunt1
- Output Problem1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJan 31, 2022Dull, Blunt; Output Problem
ON (B)(6) 2022 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA THE PHONE THAT AN AR-1204AF-60 FLIPCUTTER LOCKED AND WOULD NOT RETURN TO CLOSING POSITION. THIS WAS DISCOVERED DURING USE IN A KNEE PROCEDURE W/ MENISCAL ROOT REPAIR ON (B)(6) 2022. THE SURGEON USE THE FLIPCUTTER TO CREATE THE FEMORAL SOCKET. AFTER REAMING THE BLADE COLD-WELDED AND WOULD NOT CLOSE. THE SURGEON REMOVED THE AR-1204F-60 WITH THE BLADE IN THE OPEN POSITION, USING THE ENTIRE FEMORAL TUNNEL TO BE LARGER THAN EXPECTED. THE KNEE SCORPION SUTURE PASSER AR-12990, WAS USED TO PASS THE SUTURES FOR THE POSTERIOR REPAIR. THE SURGEON USED THREE DIFFERENT AR-12990 KNEE SCORPION WITH THREE AR-12990N'S 2 OF (LOT 1345569) BOTH NEEDLES DELAMINATING INSIDE THE AR-12990 WHILE THE SURGEON WAS TRYING TO PASS THROUGH (LOT 14109312) PUSHED LATERAL WHILE PASSING THROUGH JAMMING INSIDE THE TRAP DOOR. THE CASE WAS COMPLETED USING A 4TH AR-12990N WITH OUT ISSUE. THE SALES REPRESENTATIVE STATED 30-35 MINS WAS ADDED TO THE PROCEDURE, REQUIRING ADDITIONAL ANESTHESIA.
Patient problems: Insufficient Information
MAUDE report 13397818 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page