Arthrex® AR-1204AF-75Recalls & reported problems
Arthrex, Inc. · 7.5 MM FLIP CUTTER II · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867004238), received between Dec 21, 2021 and Mar 5, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Disconnection1
- Material Integrity Problem1
- Retraction Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 5, 2024Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY PLUS OSTEOTOMY SURGERY AFTER INSERTION INTO THE TISSUE, THE DRILLING TIP OF THE RETROGRADE DRILL BROKE OFF. ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18839178 ↗InjuryMay 30, 2023Retraction Problem; Retraction Problem
ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A AR-1204AF-75 (7.5 MM FLIP CUTTER II) WAS UNABLE TO BE UN-FLIPPED AFTER IT HAS BEEN FLIPPED IN THE PATIENT'S KNEE. THIS OCCURRED DURING USE IN AN ACL PROCEDURE ON (B)(6) 2023. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO PATIENT HARM, AFTER REAMING FULL TIBIA TUNNEL INSTEAD OF RETROGRADE DRILLING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17022970 ↗MalfunctionNov 9, 2022Disconnection; Material Integrity Problem
IT WAS REPORTED THAT DURING AN ACL RECONSTRUCTION ALL-INSIDE SURGERY THE OUTSIDE PART OF THE FLIPCUTTER CAME OFF THE DEVICE DURING DRILLING. THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE ANYWAY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE 26-OCT-2022: FURTHER INFORMATION REVEALED THAT THE FLIPCUTTER HAS 2 PARTS: AN INSIDE WIRE AND AN OUTSIDE WIRE. THE OUTSIDE PART GOT LOOSE FROM THE INSIDE PART AT THE END OF THE BLUE PART AND IT SLID OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15758810 ↗MalfunctionDec 21, 2021Break
IT WAS REPORTED THAT DURING AN ALL-INSIDE ANTERIOR CRUCIATE LIGAMENT ARTHROSCOPIC KNEE SURGERY THE TIP OF THE DEVICE BROKE OFF INSIDE THE PATIENT DURING DRILLING. NO BROKEN PARTS REMAINED INSIDE THE PATIENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13053876 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page