Arthrex® AR-1204AS-80Recalls & reported problems
Arthrex, Inc. · FLIPCUTTER II, SHORT 8MM · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00888867004412), received between Dec 20, 2019 and May 28, 2025 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Adverse Event Without Identified Device or Use Problem1
- Failure to Cut1
- Inadequacy of Device Shape and/or Size1
- Mechanical Jam1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Code Available1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 28, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT DURING AN ALL-INSIDE ANTERIOR CRUCIATE LIGAMENT SURGERY THE SIZE INDICATED ON THE PACKAGING WAS 8 BUT THE FLIP CUTTER WAS ACTUALLY A 9. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE AVOE 14-MAY-2025: IT WAS CONFIRMED THAT THE ERROR WAS DETECTED WHILE WORKING ON THE PATIENT. THERE WERE NO EFFECTS ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22098288 ↗MalfunctionMay 24, 2024Failure to Cut; Failure to Cut
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY WHEN THEY OPENED THE FLIPCUTTER AND TRIED TO DRILL, IT CLOSED. THE DEVICE DID NOT STAY OPEN FOR DRILLING. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19390471 ↗MalfunctionMay 23, 2024Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGERY THE DEVICE COULD NOT BE FLIPPED AND GOT STUCK IN THE SLEEVE. NO PART OF THE DEVICE BROKE OFF. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19385901 ↗MalfunctionJan 8, 2024Break; Break
IT WAS REPORTED THAT DURING A CRUCIATE LIGAMENT RECONSTRUCTION SURGERY THE BLADE OF THE DEVICE BROKE OFF. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE AVOE 14-DEC-2023 IT WAS CONFIRMED THAT THE TIP OF THE DEVICE BROKE OFF IN THE PATIENT'S JOINT AND WAS THEN REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18466848 ↗MalfunctionJul 20, 2021Break
IT WAS REPORTED THAT DURING ANTERIOR CRUCIATE LIGAMENT SURGERY THE TIP BROKE OFF. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12198032 ↗InjuryOct 30, 2020Mechanical Problem
IT WAS REPORTED THAT DURING AN ACL SURGERY, THE FLIPCUTTER LOCKED IN TILTED POSITION. THE SURGEON HAD TO CREATE A COMPLETE FEMORAL TUNNEL. CHANGE OF FIXATION MODE WITH FEMORAL SCREW INSTEAD OF CORTICAL BUTTON. THE SURGERY WAS COMPLETED SUCCESSFULLY, NO HARM FOR PATIENT, USER OR THIRD PARTY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10764233 ↗InjuryDec 20, 2019Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED DURING AN ACL POST OP X-RAY THE TIP OF THE FLIP CUTTER WAS DISCOVERED INSIDE THE PATIENT. A REVISION IS SCHEDULED FOR (B)(6) 2019 TO REMOVE THE FOREIGN BODY. THE PATIENT IS A 13 YEAR OLD MALE. THE DOCTOR STATED THE OUTCOME OF THE PROCEDURE WAS SUCCESSFUL AND THE REVISION IS SOLELY TO REMOVE THE TIP FROM THE KNEE. ADDITIONAL INFORMATION OBTAINED (12/10/2019): THE SECOND PROCEDURE TO REMOVE THE TIP OF THE FLIP CUTTER WAS PERFORMED ON (B)(6) 2019 AS SCHEDULED AND IT WAS RETRIEVED SUCCESSFULLY WITH NO COMPLICATIONS.
Patient problems: No Code Available
MAUDE report 9506122 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page