Arthrex® AR-1250LTRecalls & reported problems
Arthrex, Inc. · STEP DRL,2.4MM/3.0MM TISSTak/Bio-S-TAK · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00888867006287), received between Apr 18, 2022 and Aug 6, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Foreign Body In Patient2
- Insufficient Information1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50158547 ↗MalfunctionAug 8, 2024Break; Break
ON 24TH JULY 2024, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT AN AR-1250LT, STEP DRILL FOR 3.0MM BIO-SUTURETAK®, WAS USED TO PUNCH HOLES INTO THE BONE TO FACILITATE SCREW INSERTION HOWEVER THE DRILL BIT BROKE DURING DRILLING AND ABOUT 15MM OF THE BROKEN DRILL BIT REMAINED IN THE SHOULDER BLADE. THIS WAS DETECTED DURING USE IN A REPAIR OF THE RIGHT SHOULDER JOINT CAPSULE AND LIP SURGERY ON (B)(6) 2019. THERE WAS NO ADVERSE EFFECT ON THE PATIENT'S ORIGINAL CONDITION, BUT IT MAY NOT BE POSSIBLE TO PERFORM MAGNETIC RESONANCE EXAMINATION IN THE FUTURE. IT WAS REPORTED TO NMPA AS ADVERSE EVENTS BY HOSPITAL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE ONLY ISSUE BEING THAT THE BROKEN PIECE WAS LEFT IN THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 19943932 ↗MalfunctionJul 11, 2024Break; Break
ON 06/28/2024 ADDITIONAL INFORMATION WAS PROVIDED BY THE ARTHREX SUBSIDIARY EMPLOYEE VIA EMAIL WHO STATED THAT THE PROCEDURE BEING PERFORMED WAS A PATELLOFEMORAL. ALL FRAGMENTS WERE REMOVED FROM THE PATIENT. THE HOSPITALS 2.0 GUIDE WIRE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 19721053 ↗MalfunctionMar 5, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
ON 02/19/2024, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA (B)(4) THAT AN AR-1250LT STEP DRILL GOT STUCK IN THE DRILL GUIDE WHEN MAKING THE PRE-HOLE. THIS OCCURRED DURING A CASE WHERE ANOTHER AR-1250LT WAS USED TO COMPLETE THE CASE. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18838913 ↗MalfunctionMar 1, 2024Break; Break
ON 02/15/2024, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA EMAIL THAT AN AR-1250LT STEP DRILL FOR 3.0MM BIO-SUTURETAK BROKE AND GOT STUCK IN THE PATIENT'S BONE. THE BROKEN FRAGMENTS WERE NOT RETRIEVED AND REMAIN IN THE PATIENT. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2024. NO FURTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION REQUESTED.
Patient problems: Foreign Body In Patient
MAUDE report 18818191 ↗MalfunctionSep 26, 2023Break; Break
ON 8/28/2023, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA SEMS-06016980 THAT AN AR-1250LT STEP DRILL BROKE DURING A BROSTROM PROCEDURE ON (B)(6) 2023. THE BONE TUNNEL WAS COMPLETED BEFORE THE DRILL BROKE. THE BROKEN PART WAS RECOVERED, AND THERE WAS NO HARM TO THE PATIENT. IT WAS POSSIBLE TO INSERT THE ANCHOR, AND THE SURGERY WAS COMPLETED SUCCESSFULLY. THE DRILL FRAGMENTS WERE DISPOSED OF AT THE HOSPITAL. THIS OCCURRED DURING USE WITH NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ON (B)(6) 2023, THE ARTHREX SUBSIDIARY EMPLOYEE PROVIDED THE FOLLOWING INFORMATION VIA EMAIL: THERE WAS NO CASE DELAY REPORTED. THIS EVENT WAS NOT AN ADVERSE REPORTABLE EVENT AND DID NOT MEET THE REPORTABILITY REQUIREMENTS OF ARTICLE 8 OF RDC 67/2009.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17816964 ↗MalfunctionApr 18, 2022Break
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY THE DRILL BROKE OFF AT THE BACK PART AND INSIDE THE PATIENT. THE BROKEN PIECE WAS RETRIEVED FROM PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH THE SAME DEVICE WAS USED ANYWAY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14135454 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.