Arthrex® AR-13975SRRecalls & reported problems
Arthrex, Inc. · FIBERWIRE GRASPER W/ SR HANDLE · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867015838), received between Apr 21, 2017 and Nov 18, 2025 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Detachment Of Device Component1
- Device Or Device Fragments Location Unknown1
- Entrapment of Device1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Foreign Body In Patient2
- Insufficient Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 18, 2025Break; Break
ON 11/03/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT THE AR-13975SR FIBERWIRE GRASPER TIP BROKE OFF. THIS WAS DISCOVERED DURING A CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23584324 ↗MalfunctionNov 22, 2024Break; Break
ON 10/30/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-13975SR FIBERWIRE® GRASPER BROKE. THIS OCCURRED DURING A SHOULDER ARTHROSCOPY ON (B)(6) 2024. THE BOTTOM JAW BROKE OFF WHILE INSERTING THE INSTRUMENT INTO THE JOINT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20758995 ↗MalfunctionDec 27, 2022Material Twisted/Bent; Break; Material Twisted/Bent
ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT A AR-13975SR FIBERWIRE® GRASPER W/SR HANDLE IS BROKEN AT THE DISTAL END OF THE TIP. THIS OCCURRED DURING A CASE, NO PIECES WERE LEFT IN THE PATIENT, AND THE CASE WAS COMPLETED. THERE WAS NO PATIENT EFFECT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16051575 ↗MalfunctionNov 16, 2021Break
ON (B)(6) 2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-13975SR FIBERWIRE GRASPER, DISTAL END BROKE. THIS WAS DISCOVERED DURING USE IN PROCEDURE ON (B)(6) 2021. ALL FRAGMENT WERE REMOVED FROM THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 12819165 ↗MalfunctionMay 24, 2021Break
IT WAS REPORTED THAT DURING A RCR/ BANKART PROCEDURE, THE AR-13975SR BROKE. THE FRAGMENT WAS REMOVED AND THE CASE WAS COMPLETED BY USING A NEW AR-13975SR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11872439 ↗InjuryApr 21, 2017Break; Detachment Of Device Component; Entrapment of Device; Device Or Device Fragments Location Unknown
IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, THE SURGEON WAS USING THE FIBERWIRE GRASPER WITH SR HANDLE TO GRASP THE BICEPS TENDON. THE UPPER JAW OF THE GRASPER BROKE-OFF INTO THE PATIENT'S JOINT. THE SURGEON ATTEMPTED TO LOCATE THE JAW FRAGMENT WITH THE SCOPE HOWEVER, COULD NOT LOCATE IT. THE SURGEON THEN HAD AN X-RAY BROUGHT IN AND WAS ABLE TO CONFIRM THE LOCATION OF THE JAW FRAGMENT BUT WAS UNABLE TO RETRIEVE IT. A SECOND SURGERY IS UNDETERMINED AT THE MOMENT AND THE SURGEON WILL WATCH FOR ANY SYMPTOMS BEING CAUSED BY THE DEVICE FRAGMENT.
Patient problems: Foreign Body In Patient
MAUDE report 6511012 ↗
Other Accessories, arthroscopic devices
- Arthrex® AR-12990Arthrex, Inc. · Accessories, arthroscopic409 reports
- NA 228302Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC287 reports
- FIRSTPASS 72290128Smith & Nephew, Inc. · ACCESSORIES,ARTHROSCOPIC232 reports
- NA 251723Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC145 reports
- COBRA 3910900093Stryker Corporation · Accessories, arthroscopic138 reports
- FirstPass 72290129Smith & Nephew, Inc. · Accessories, arthroscopic82 reports
- FirstPass 72290130Smith & Nephew, Inc. · Accessories, arthroscopic80 reports
- MITEK 242025Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC61 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page