Skip to content

Arthrex® AR-1934-24DSRecalls & reported problems

Arthrex, Inc. · 2.4MM BIO-SUTURETAK DISPOSABLES KIT · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00888867023864), received between Mar 21, 2022 and Jan 20, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Physical Resistance/Sticking2
  • Device Dislodged or Dislocated1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 20, 2026Physical Resistance/Sticking

    IT WAS REPORTED THAT THE DRILL BECAME STUCK INSIDE THE GUIDE. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THIS WAS NOTICED BEFORE THE SURGERY. THERE WAS NO CASE INVOLVEMENT REPORTED. NO FURTHER INFORMATION RECEIVED. UPDATE 15-JAN-2025: FURTHER INFORMATION WAS RECEIVED, ON 05-JAN-2026, THAT THE CASE OCCURRED DURING A SURGERY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW ANCHOR. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24114343 ↗
  • MalfunctionJun 27, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON 06/05/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-1934-24DS BIO-SUTURETAK DISPOSABLE KIT GUIDES CAME OUT BUT THE OBTURATORS WERE STUCK INSIDE OF THEM AND THE DRILLS DID NOT PASS THROUGH THE GUIDES. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22349482 ↗
  • MalfunctionMay 1, 2025Break; Break

    IT WAS REPORTED THAT DURING A SHOULDER INSTABILITY REPAIR SURGERY AN ISSUE WITH THE REPORTED DEVICE HAS OCCURRED. THE SURGEON HAS DRILLED THE HOLE AND WHEN HE TRIED TO REMOVE THE DRILL, IT THAT GOT STUCK. THE TIP OF THE DRILL FUSED WITH PART OF THE SLEEVE, WHILE THE LOWER PART, WHERE THE DRILL IS ATTACHED, BROKE OFF. THE PIECE WAS RECOVERED AND DID NOT REMAIN INSIDE THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21942701 ↗
  • MalfunctionJun 19, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT AFTER USING THE AR-1934-24DS THE DRILL BIT COULD NOT BE REMOVED FROM THE DRILL GUIDE. NO ADVERSE EFFECT TO THE PATIENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19567099 ↗
  • MalfunctionMay 21, 2024Break; Break

    IT WAS REPORTED ON (B)(6) 2023 BY AN ARTHREX EMPLOYEE VIA SEMS THAT AN AR-1934-24DS DRILL BIT GOT STUCK IN THE AR-1934BCF-24-2 DRILL SLEEVE. IT WAS USED AT THE ELBOW HEAD. USED A NEW PRODUCT, AND THE CASE WAS COMPLETED. CASE INVOLVEMENT, NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19359468 ↗
  • MalfunctionJan 3, 2023Break; Mechanical Jam

    ON (B)(6) 2022, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT A DRILL FROM AN AR-1934-24DS BIO-SUTURETAK DISPOSABLE KIT, GOT STUCK INSIDE THE GUIDE AND COULD NOT BE REMOVED. WHILE ATTEMPTING TO REMOVE IT, THE DRILL BROKE. A NEW PRODUCT WAS USED, AND CASE WAS COMPLETED. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2022.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16086969 ↗
  • MalfunctionMar 21, 2022Break

    ON 01/06/2022 IT WAS REPORTED BY A ARTHREX REPRESENTATIVE VIA SEMS THAT AN AR-1934-24DS, THE DRILL BIT BREAKAGE STACKED IN THE DRILL GUIDE, THE TIP OF THE DRILL WAS DAMAGED. THE CASE WAS COMPLETED USING THE NEW PRODUCT IN THE SAME LOT. NO PATIENT HARMS. ADDITIONAL INFORMATION 1/13/2022: ON 01/13/2022 IT WAS REPORTED BY A ARTHREX REPRESENTATIVE VIA SEMS THAT AN AR-1934-24DS, DURING A PROCEDURE THE DRILL BUT WAS STUCK INSIDE THE DRILL GUIDE. THE DRILL TIP WAS DAMAGED DURING DRILLING, NO DEBRIS CAME OFF THE DEVICE. THE CASE WAS COMPETED USING A NEW AR-1934-24DS. DURING RETURNED DEVICE EVALUATION, A REPORTABLE MALFUNCTION WAS DISCOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13839475 ↗

Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices

All INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices →