Arthrex® AR-1938DRecalls & reported problems
Arthrex, Inc. · STEP DRL,FOR 3.0MM KNOTLESS S-TAK · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867024212), received between Feb 17, 2017 and Feb 17, 2017 — classified as 2 injuries.
The most frequently reported specific device problem is “Material Fragmentation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Material Fragmentation1
- Metal Shedding Debris1
Patient problems
- Device Embedded In Tissue or Plaque1
- No Information1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 17, 2017Metal Shedding Debris
IT WAS REPORTED THAT TWO 2.6MM DRILLS FOR 3MM SUTURETAK (AR-1938D) ALONG WITH ONE 3MM SUTURETAK PERCUTANEOUS INSERTION KIT (CC104986 LINE181875 AR-1934PI-30) AND ONE TROCAR TIP OBTURATOR FOR 3MM SUTURETAK DISPOSABLE SPEAR (CC104986 LINE 181877 AR-1949S) WERE BEING USED IN A SHOULDER LABRAL REPAIR PROCEDURE. WHILE USING THE FIRST DRILL (CC104986 LINE 181872) WITH A COMPETITORS DEVICE SET AT FULL SPEED ALONG WITH THE DRILL GUIDE FROM AR-1934PI-30 (CC104986 LINE 181875), IT WAS NOTICED THAT THERE WERE METAL SHAVINGS BEING PRODUCED INSIDE THE PATIENT. AN ATTEMPT WAS MADE TO REMOVE THE METAL SHAVINGS FROM THE PATIENT BY USING A FRAZIER TIP SUCTION DEVICE AND HIGH FLOW SHAVER. THE SURGEON WAS UNABLE TO RETRIEVE ALL SHAVINGS. THE SECOND DRILL (CC104986 LINE181874) WAS USED WITH A COMPETITORS DEVICE, ALONG WITH THE DISPOSABLE SPEAR (CC104986 LINE 181877) AND ALSO PRODUCED METAL SHAVINGS. THE SAME TECHNIQUE WAS USED AGAIN TO TRY AND REMOVE THE METAL SHAVINGS, AND AGAIN ALL OF THE SHAVINGS WERE UNABLE TO BE RETRIEVED FROM THE PATIENT. SALES REP STATED THE SURGEON MENTIONED THAT SHE BELIEVES THE METAL SHAVINGS WERE CREATED BY THE DRILL AND DRILL GUIDE COMING IN CONTACT WITH EACH OTHER. A REUS
Patient problems: No Information
MAUDE report 6338830 ↗InjuryFeb 17, 2017Material Fragmentation
IT WAS REPORTED THAT TWO 2.6MM DRILLS FOR 3MM SUTURETAK (AR-1938D) ALONG WITH ONE 3MM SUTURETAK PERCUTANEOUS INSERTION KIT (CC104986 LINE181875 AR-1934PI-30) AND ONE TROCAR TIP OBTURATOR FOR 3MM SUTURETAK DISPOSABLE SPEAR (CC104986 LINE 181877 AR-1949S) WERE BEING USED IN A SHOULDER LABRAL REPAIR PROCEDURE. WHILE USING THE FIRST DRILL (CC104986 LINE 181872) WITH A COMPETITORS DEVICE SET AT FULL SPEED ALONG WITH THE DRILL GUIDE FROM AR-1934PI-30 (CC104986 LINE 181875), IT WAS NOTICED THAT THERE WERE METAL SHAVINGS BEING PRODUCED INSIDE THE PATIENT. AN ATTEMPT WAS MADE TO REMOVE THE METAL SHAVINGS FROM THE PATIENT BY USING A FRAZIER TIP SUCTION DEVICE AND HIGH FLOW SHAVER. THE SURGEON WAS UNABLE TO RETRIEVE ALL SHAVINGS. THE SECOND DRILL (CC104986 LINE181874) WAS USED WITH A COMPETITORS DEVICE, ALONG WITH THE DISPOSABLE SPEAR (CC104986 LINE 181877) AND ALSO PRODUCED METAL SHAVINGS. THE SAME TECHNIQUE WAS USED AGAIN TO TRY AND REMOVE THE METAL SHAVINGS, AND AGAIN ALL OF THE SHAVINGS WERE UNABLE TO BE RETRIEVED FROM THE PATIENT. SALES REP STATED THE SURGEON MENTIONED THAT SHE BELIEVES THE METAL SHAVINGS WERE CREATED BY THE DRILL AND DRILL GUIDE COMING IN CONTACT WITH EACH OTHER. A REUS
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6338826 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page