Arthrex® AR-21020Recalls & reported problems
Arthrex, Inc. · CURETTE, RING 3.4MM, CUTTING BOTH SIDE · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867026056), received between Apr 11, 2018 and Jul 8, 2025 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Crack1
- Migration or Expulsion of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 8, 2025Break; Break
IT WAS REPORTED THAT DURING A SURGERY THE DEVICE BROKE. ALL FRAGMENTS WERE RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH AN EQUIVALENT DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22440510 ↗MalfunctionMay 15, 2025Break; Break
IT WAS REPORTED THAT DURING A SURGERY THE DEVICE BROKE OFF. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT CURETTE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22026452 ↗MalfunctionOct 26, 2023Crack; Crack
ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-21020 CURETTE RING IS CRACKED. THE ISSUE OCCURRED DURING THE CASE AND WAS TAKEN OUT OF SERVICE IMMEDIATELY. THE PATIENT WAS NOT AFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18013429 ↗MalfunctionAug 3, 2023Break; Break
ON 7/5/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-21020 CURETTE HAD AN ISSUE. DURING THE SURGERY, THE SURGEON WAS PREPARING THE GLENOID FOR A LABRAL REPAIR, AND WHILE USING THE CURETTE, IT BROKE WHEN IT MADE CONTACT WITH THE BONE AND SNAPPED IN HALF. THE BROKEN PIECE WAS RETRIEVED FROM THE JOINT, AND THE CASE WAS CONTINUED WITHOUT THE PATIENT BEING AFFECTED. THIS OCCURRED DURING USE IN A LABRAL REPAIR PROCEDURE WITH NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ON 7/18/2023, THE SALES REPRESENTATIVE STATED THE FOLLOWING INFORMATION VIA PHONE: THIS EVENT OCCURRED DURING A LABRAL REPAIR PROCEDURE ON (B)(6) 2023. AFTER THE BROKEN PIECE HAD BEEN RETRIEVED, A NEW AR-21020 CURETTE WITH AN UNKNOWN LOT NUMBER WAS USED TO COMPLETE THE CASE SUCCESSFULLY WITH NO PATIENT HARM. THERE WAS NO CASE DELAY OR ANY ADVERSE EFFECT TO THE PATIENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17450172 ↗MalfunctionDec 27, 2022Break
ON 12/06/2022 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-21020 RING CURETTE BROKE OFF IN THE PATIENT. ALL WAS RETRIEVED AND THE PATIENT WAS NOT AFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16051439 ↗InjuryApr 11, 2018Break; Migration or Expulsion of Device
IT WAS REPORTED THAT DURING AN ANKLE ARTHROSCOPIC ARTHRODESIS, THE ANKLE WAS DECORTICATED WITH A 4MM BURR, A 3.5MM FULL RADIUS RESECTOR AND RING CURETTES. DURING DECORTICATION WITH RING CURETTES, IT WAS NOTED THAT A SMALL FRAGMENT OF THE RING CURETTE BROKE AND ENDED UP IN THE POSTERIOR RECESS OF THE ANKLE. ATTEMPTS TO RETRIEVE IT WERE UNSUCCESSFUL; SURGEON OPTED TO LEAVE IT IN INSTEAD OF POSSIBLY CAUSING MORE HARM. PATIENT WILL BE MONITORED BY SURGEON. PATIENT IS A (B)(6) MALE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 7422027 ↗
Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices
- Arthrex® AR-13995NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE430 reports
- Arthrex® AR-12990NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE197 reports
- Arthrex® AR-13991NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE128 reports
- Arthrex® AR-8981-08SArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall3 reports
- Arthrex® AR-8981-06SArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall1 report
- Centurion EB16Centurion Medical Products Corporation · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
- NA Lighthandle CoverSteris Corporation · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
- Centurion Medical Products Corp SUT15830AMedline Industries, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page