Skip to content

Arthrex® AR-300CRecalls & reported problems

Arthrex, Inc. · 300 POWER SYSTEM INSTRUMENT CASE · Class 1

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867028661), received between Jun 10, 2024 and Jun 19, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Corroded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Corroded1
  • Flaked1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 19, 2024Flaked; Flaked

    IT WAS REPORTED THAT THE COATING CHIPPED OFF. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19569794 ↗
  • MalfunctionJun 10, 2024Corroded; Corroded

    IT WAS REPORTED THAT THERE WAS A CHANGING OF THE SURFACE. THERE WAS NO HARM OR ADVERSE EVENT FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19497570 ↗

Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices

All INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.