Skip to content

Arthrex® AR-501-TTLDRecalls & reported problems

Arthrex, Inc. · IBAL UKA TIB TRAY SZ 4,LFT-MEDIAL/RL · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867034150), received between Oct 20, 2023 and Oct 20, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Unspecified Infection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 20, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    ON (B)(6) 2022, IT WAS REPORTED BY THROUGH A PATIENT'S LEGAL FILING THAT ON (B)(6) 2017 THE PATIENT UNDERWENT A LEFT KNEE UNI-COMPARTMENTAL KNEE ARTHROPLASTY. DURING THE PROCEDURE THE SURGEON IMPLANTED AN ARTHREX IBALANCE UNI-COMPARTMENTAL KNEE SYSTEM WITH A SIZE 3 FEMORAL COMPONENT, SIZE 4 TIBIAL TRAY, AND AN 8 MM TIBIAL BEARING INSERT INTO THE LEFT KNEE MEDIAL PARAPATELLAR FACETECTOMY. THEREAFTER THE PATIENT WAS HAVING PAIN IN THE LEFT KNEE AND ON (B)(6) 2019 AN MRI OF THE KNEE WAS TAKEN WHICH REVEALED ¿METALLIC ARTIFACT IS PRESENT INVOLVING THE MEDIAL COMPARTMENT OBSCURING AND DISTORTING THE UNDERLYING AND ADJACENT STRUCTURES.¿ ON (B)(6) 2019 THE PATIENT REPORTED A CRUNCHING SENSATION ON THE MEDIAL ASPECT OF THE KNEE. ON (B)(6) 2019 AND AGAIN ON (B)(6) 2020 THE PATIENT COMPLAINED OF ACHES, PAINS AND WEAKNESS IN THE KNEE. ON (B)(6) 2020 THE PATIENT PRESENTED TO ANOTHER SURGEON AND A BONE SCAN WAS ORDERED WHICH REVEALED A LOOSE IMPLANT. THE PATIENT WAS REFERRED TO ANOTHER SURGEON TO DISCUSS REVISION OPTIONS DUE TO LOOSENING OF THE IMPLANT AND ON (B)(6), 2020 THAT SURGEON PERFORMED AN EXPLANTATION AND ANTIBIOTIC SPACER PLACEMENT, LEFT TOTAL KNEE ARTHROPLASTY AND SAUCERIZATION OF FE

    Patient problems: Unspecified Infection

    MAUDE report 17980950 ↗

Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page