Skip to content

Arthrex® AR-501-UFLDRecalls & reported problems

Arthrex, Inc. · IBAL UKA FEM CEMENTED SIZ 4,L-MEDIAL · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867034273), received between Dec 16, 2019 and Aug 29, 2022 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Naturally Worn” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Naturally Worn1
  • Output Problem1
  • Unstable1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 29, 2022Output Problem

    ON 08/08/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT DURING A MEDIAL UNI-COMPARTMENTAL KNEE REPLACEMENT A AR-501-UFLD IBALANCE¿ UKA FEMORAL,CEMENTED SIZE 4S THE CEMENT BECAME TO FIRM WHEN TRYING TO IMPLANT THE FEMORAL COMPONENT. NOT BEING ABLE TO FULLY SEAT THE FEMORAL COMPONENT, THE COMPONENT WAS THEN REMOVED AS WELL AS THE CEMENT REMOVED AND A NEW BATCH OF CEMENT MIXED. A 2ND COMPONENT WAS OPENED, AS ALL THE PRIOR CEMENT COULD NOT BE REMOVED FROM THE ORIGINAL IMPLANT. IT WAS ALSO FOUND THAT THE TIBIAL ACL INSERTION HAD AVULSED, THAT WAS REPAIRED BACK DOWN TO THE TIBIA WITH A 4.5 CANNULATED SCREW AND WASHER, AS WELL AS A FIBERTAPE FIXATED WITH A SWIVELOCK INTO THE PROXIMAL TIBIA. THE CASE WAS COMPLETED, WITH NO PATIENT EFFECT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15310539 ↗
  • InjuryDec 16, 2019Unstable; Naturally Worn

    THE PATIENT HAD ORIGINAL SURGERY FOR A UNI KNEE REPLACEMENT ON (B)(6) 2018. SHE REPORTED PAIN AND INSTABILITY; XRAYS SHOWED THAT THE POLY HAD WORN DOWN. ARTHREX PARTS AR-521-TBA8, AR-501-UFLD AND AR-501-TTLA WERE REMOVED. A TOTAL KNEE REVISION WAS DONE ON (B)(6) 2019 WITH A COMPETITOR'S DEVICE. PATIENT FEMALE, 60S.

    Patient problems: Pain

    MAUDE report 9473632 ↗

Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates