Skip to content

Arthrex® AR-6420CDSRecalls & reported problems

Arthrex, Inc. · DISSECTOR, CURVED, HL 4.2MM X 19CM · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867039193), received between Jan 23, 2023 and Jan 13, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation3
  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 13, 2026Material Fragmentation

    ON 19-DEC-2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6420CDS CURVED DISSECTOR LEFT METAL SHAVINGS BEHIND. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM. ADDITIONAL INFORMATION WAS PROVIDED BY THE REP ON 23-DEC-25 THAT THIS WAS DISCOVERED DURING A HIP SCOPE. THE SHAVINGS WERE DISCOVERED DURING THE DEBRIDEMENT OF LABRUM. ALL FRAGMENTS WERE RETRIEVED. BOTH SUCTION AND IRRIGATION WERE USED, AND NO ALARMS WERE PRESENT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24053050 ↗
  • MalfunctionJan 13, 2026Material Fragmentation

    ON 19-DEC-2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT (QTY. 2) OF AN AR-6420CDS CURVED DISSECTORS LEFT METAL SHAVINGS BEHIND. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM. ADDITIONAL INFORMATION WAS PROVIDED BY THE REP ON 23-DEC-25, THAT THIS WAS DISCOVERED DURING A HIP SCOPE. THE SHAVINGS WERE DISCOVERED DURING THE DEBRIDEMENT OF LABRUM. ALL FRAGMENTS WERE RETRIEVED. BOTH SUCTION AND IRRIGATION WERE USED, AND NO ALARMS WERE PRESENT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24052835 ↗
  • MalfunctionApr 24, 2025Break; Break

    ON 04/08/2025, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA (B)(4) THAT AN AR-6420CDS DISSECTOR SHAVER BLADE BROKE. THIS WAS DISCOVERED DURING THE CASE. THERE WERE NO FRAGMENTS AND ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21901835 ↗
  • MalfunctionJan 23, 2023Material Fragmentation

    IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY DEBRIS WAS PRODUCED WITH THE DEVICE. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE AVOE (B)(6) 2023 IT WAS CONFIRMED THAT ALL FRAGMENTS COULD BE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16223081 ↗

Other BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices

All BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page