Arthrex® AR-6480Recalls & reported problems
Arthrex, Inc. · DW ARTHROSCOPY FLUID MANAGEMENT DEV · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Arthrex AR-6480 is an arthroscopy fluid management device that controls fluid flow during arthroscopic surgery. The 937 reports mostly describe pressure problems, unusually high fluid pressure or flow, suction failures and intermittent power loss. Some mention swelling from fluid that escaped into surrounding tissue. The reports include no deaths and 93 injuries, and no recalls name this device. Reports fell to 59 over the last year from 114 in the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 925
- Deaths reported
- 0
- Injuries reported
- 91
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 925 adverse event reports that list this exact device (DI 00888867039377), received between Jul 13, 2016 and Jul 30, 2026 — classified as 91 injuries, 834 malfunctions.
The most frequently reported specific device problem is “Pressure Problem” (185 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Pressure Problem185
- Increase in Pressure104
- Defective Device85
- Infusion or Flow Problem64
- Suction Failure46
- Intermittent Loss of Power44
- Unexpected Shutdown30
- No Flow28
Patient problems
- No Clinical Signs, Symptoms or Conditions685
- Insufficient Information122
- Swelling/ Edema47
- Extravasation34
- Swelling13
- No Code Available8
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available7
- No Known Impact Or Consequence To Patient6
Recent adverse event reports
The 25 most recent of 925 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 30, 2026Increase in Pressure
Patient problems: Joint Dislocation; No Clinical Signs, Symptoms or Conditions
MAUDE report 50083099 ↗MalfunctionApr 28, 2026Increase in Pressure
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC SURGERY THE PRESSURE FROM THE PUMP INCREASED BY ITSELF FROM THE SET 35MMHG TO 90MMHG. THE ISSUE WAS NOTICED AND THE INFLOW COULD BE STOPPED IN TIME TO PREVENT A SERIOUS INJURY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE.
Patient problems: Insufficient Information
MAUDE report 25018933 ↗MalfunctionApr 28, 2026Increase in Pressure
IT WAS REPORTED THAT DURING THE HIP ARTHROSCOPY, THE BLACK PRESSURE VALVE ON THE INFLOW TUBE DID NOT INFLATE. THE PUMP STOPPED OPERATING DUE TO A HIGH-PRESSURE ALARM. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25014990 ↗MalfunctionApr 9, 2026Increase in Pressure
IT WAS REPORTED THAT DURING A SHOULDER INSTABILITY SURGERY THE DEVICE BUILT UP A LOT OF PRESSURE AND WOULDN¿T STOP IN ANY WAY. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. THIS LINE WAS CREATED FOR THE UNKNOWN PUMP, THAT WAS USED WITH THE REPORTED TUBING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24833517 ↗InjuryMar 31, 2026Patient Device Interaction Problem
IT WAS REPORTED THAT AFTER A SURGERY THE PATIENT DEVELOPED A COLLAPSED LUNG (PNEUMOTHORAX) WHICH HAD TO BE FURTHER MONITORED. IT WAS REPORTED THAT THE SURGERY WAS ON THE RIGHT SHOULDER AND THE PNEUMOTHORAX OCCURRED IPSILATERALLY. IT WAS CONFIRMED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY. AS A RESULT OF THE PNEUMOTHORAX, THE PATIENT EXPERIENCED ACUTE RESPIRATORY FAILURE, HOWEVER, NO OTHER INTERVENTIONS FOR THE PNEUMOTHORAX WERE PERFORMED BEYOND OBSERVATION IN THE STEP-DOWN UNIT. IT WAS ALSO REPORTED THAT SUBCUTANEOUS EMPHYSEMA WAS NOTED AFTER THE SURGERY. DURING THE SURGERY, THE PATIENT WAS POSITIONED HALF-SITTING, NO COMPLICATIONS WITH THE ANESTHESIA OCCURRED, AND THE INTRAOPERATIVE AIRWAY PRESSURE WAS MAINTAINED STABLE THOUGH THE PROCEDURE. THE PATIENT DID NOT HAVE OTHER UNDERLYING LUNG DISEASES. IT WAS FURTHER REPORTED THAT IT IS UNKNOWN IF AN ARTHREX TUBING WAS USED HOWEVER THE USED TUBING WAS SCRAPPED.
Patient problems: Pneumothorax; Respiratory Failure
MAUDE report 24742175 ↗MalfunctionFeb 24, 2026Excess Flow or Over-Infusion; Failure to Auto Stop; Noise, Audible
ADDITIONAL INFORMATION WAS PROVIDED 19-FEB-2026: IT WAS REPORTED THE ISSUE OCCURRED DURING TWO SEPARATE KNEE ARTHROSCOPY PROCEDURES (REFER TO (B)(4)). AR-6410 MAIN PUMP TUBING WAS USED WITH THE AR-6480 DW ARTHROSCOPY PUMP. THE PUMP SETTING USED WAS REPORTED TO BE THE DEFAULT KNEE SETTING. THE CASE WAS COMPLETED WITH ANOTHER PUMP. NO CLAMP OR PRESSURE TEST WAS PERFORMED, AND NO ALARMS OR ERROR MESSAGES WERE OBSERVED BEFORE OR DURING THE EVENT. IT WAS FURTHER REPORTED EXTRAVASATION OCCURRED WITHIN THE PATIENT¿S JOINT DURING THE EVENT. THE SURGICAL TEAM ATTEMPTED TO DRAIN AS MUCH EXCESS FLUID AS POSSIBLE AND WRAPPED THE PATIENT¿S LEG TO HELP MITIGATE KNEE SWELLING. IT REMAINS UNKNOWN WHETHER THE PATIENT REQUIRED EXTENDED HOSPITALIZATION FOLLOWING THE EVENT.
Patient problems: Extravasation; No Clinical Signs, Symptoms or Conditions
MAUDE report 24429564 ↗MalfunctionFeb 20, 2026Suction Failure
ON 04-FEB-2026, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 DW ARTHROSCOPY PUMP WAS NOT WORKING, THERE WAS NO SUCTION IN THE OUTFLOW OF THE SYSTEM. THIS WAS DISCOVERED DURING A PROCEDURE WITH NO PATIENT HARM. ADDITIONAL INFORMATION PROVIDED 13-FEB-2026: IT WAS FURTHER REPORTED THIS OCCURRED DURING A SHOULDER ROTATOR CUFF REPAIR PROCEDURE THAT WAS COMPLETED WITH ANOTHER ARTHREX PUMP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24404383 ↗MalfunctionFeb 19, 2026Increase in Pressure
ON 03-FEB-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 PUMP WAS SET AT 40 PRESSURE AND THEN SUDDENLY RAISES TO 100 IN PRESSURE. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24388877 ↗MalfunctionFeb 11, 2026Suction Problem
ON 22-JAN-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 PUMP WHEEL THAT CONTROLS SUCTION IS MALFUNCTIONING. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24312808 ↗MalfunctionFeb 10, 2026Output below Specifications
ON 20-JAN-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 PUMP HAS POOR OUTPUT. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT. THE PUMP WAS SWAPPED OUT AND THE CASE WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24305188 ↗MalfunctionFeb 10, 2026Increase in Pressure
ON 20-JAN-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 PUMP IS HAVING A PRESSURE ISSUE. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT. PRESSURE PERFORMANCE VERIFICATION WAS GOOD UNTIL ABOUT 86 PRESSURE READING. STARTED READING HIGHER THAN SPEC. THERE WAS NO EXTRAVASATION OR DELAY TO THE CASE. THE PUMP WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24304790 ↗MalfunctionFeb 5, 2026Increase in Pressure; Use of Device Problem
ON 15-JAN-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 PUMP IS SHOWING AN OVP DETECT FAILURE. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24268271 ↗MalfunctionFeb 4, 2026Increase in Pressure
ON 14-JAN-2026, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT THE AR-6480 DUALWAVE ARTHROSCOPIC FLUID MANAGEMENT DEVICE EXPERIENCED A PRESSURE MALFUNCTION WHERE THE PUMP INCREASED THE PRESSURE FROM 20 TO 70 WITH NO INTERACTION. THIS ISSUE WAS DISCOVERED DURING A CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24256818 ↗MalfunctionFeb 4, 2026Decrease in Pressure
ON 14-JAN-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(6) THAT AN AR-6480 PUMP HAS LOW PRESSURE. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24256676 ↗MalfunctionFeb 3, 2026Infusion or Flow Problem
ON 13-JAN-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 DW ARTHROSCOPY PUMP INFLOW AND OUTFLOW DID NOT WORK. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24248329 ↗MalfunctionFeb 3, 2026Difficult to Open or Close
ON 12-JAN-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 PUMP INFLOW BRACKET WON'T SHUT. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24248213 ↗MalfunctionFeb 3, 2026Device Alarm System; Use of Device Problem
ON 12-JAN-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 PUMP IS SHOWING A OVP DETECT FAILURE MESSAGE. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24247198 ↗MalfunctionFeb 2, 2026Inflation Problem
ON 07-JAN-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 DW ARTHROSCOPY PUMP HAD OVERINFLATED. THIS WAS DISCOVERED DURING A KNEE SCOPE. THE OVERINFLATION CAUSED THE PATIENTS KNEE TO SWELL AND THE PATIENT HAS SUBSEQUENTLY TAKEN TO A NEARBY HOSPITAL. ADDITIONAL INFORMATION WAS PROVIDED BY THE REP ON 23-JAN-26 THAT ARTHREX TUBING WAS USED WITH THE PUMP. THE PUMP SETTINGS WERE NOT RECORDED, AND THE PUMP WAS USED TO FINISH THE CASE. THERE WERE NO PICTURES OR VIDEOS CAPTURED AND NO ARTHREX REP WAS PRESENT DURING THE CASE. THE PATIENT WAS TAKEN TO A NEARBY HOSPITAL FOR OBSERVATIONS DUE TO THE OVERINFLATION WHICH CAUSED THE PATIENTS KNEE TO SWELL.
Patient problems: Swelling/ Edema
MAUDE report 24234941 ↗MalfunctionJan 23, 2026Increase in Pressure
ON 08-JAN-2025, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 DW ARTHROSCOPY PUMP PRESSURE WAS TOO HIGH. THIS WAS DISCOVERED DURING A PROCEDURE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24161876 ↗MalfunctionJan 20, 2026Use of Device Problem
IT WAS REPORTED THAT DURING AN ARTHROSCOPY OF THE UPPER ANKLE JOINT THE PUMP DID NOT DRAIN PROPERLY. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24115177 ↗InjuryJan 12, 2026Defective Alarm; Excess Flow or Over-Infusion; Pressure Problem; Key or Button Unresponsive/not Working
IT WAS REPORTED THAT THE FLOW IS HIGH AND DID NOT CHANGE IN INTENSITY EVEN THOUGH THE SETTINGS WERE CHANGED. THIS WAS NOTICED BEFORE THE SURGERY. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THERE WAS NO CASE INVOLVEMENT. NO FURTHER INFORMATION RECEIVED. UPDATE DW ON 19-DEC-2025: FURTHER INFORMATION WAS PROVIDED THAT THE REPORTED ISSUE OCCURRED DURING A SURGERY. DURING AN ONGOING SHOULDER ARTHROSCOPY, THE PUMP SUDDENLY ACCELERATES AND DELIVERS VERY HIGH FLUID FLOW RATES. THE SURGEON ATTEMPTS TO REDUCE THE FLOW, BUT THE CONTROLS DO NOT RESPOND, AND THEY ARE FORCED TO SHUT IT OFF USING THE POWER BUTTON ON THE PUMP. ALL CONNECTIONS ARE CHECKED AND APPEAR NORMAL, BUT WHEN RESTARTED, THE SAME ISSUE OCCURS AGAIN. THERE ARE NO ALARMS OR WARNINGS FROM THE PUMP. IT IS NOTED THAT THE FLOW INCREASES SIGNIFICANTLY AND THE SOUND CLEARLY INDICATES THE PUMP IS RACING. THE SHOULDER SWELLS DUE TO THE LARGE FLUID VOLUME, AND THE SURGEON IS FORCED TO TERMINATE THE PROCEDURE BECAUSE OF THIS. THE SURGEON ALSO ASSESSES THAT, DUE TO THE AMOUNT OF FLUID AND SWELLING, CHANGING EQUIPMENT AT THAT POINT WOULD NOT ENABLE COMPLETION OF THE SURGERY. THEREFORE, THE OPERATION IS NOT COMPLETED. THE PATIE
Patient problems: Swelling/ Edema
MAUDE report 24035765 ↗MalfunctionJan 8, 2026Increase in Pressure
ON 16-DEC-2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 PUMP PRESSURE IS TOO HIGH. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24002994 ↗MalfunctionJan 5, 2026Defective Device
ON 16-DEC-2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 DW ARTHROSCOPY PUMP DISPLAYED AN ERROR MESSAGE. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23960260 ↗MalfunctionJan 2, 2026Use of Device Problem; Infusion or Flow Problem
ON 12-DEC-2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 ARTHROSCOPY PUMP'S OUTFLOW PORT WAS NOT SPINNING CORRECTLY DURING USE. THIS WAS DISCOVERED DURING AN UNSPECIFIED PROCEDURE WITH NO PATIENT HARM OR CASE DELAY REPORTED. THE CASE WAS COMPLETED WITH ANOTHER DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23944464 ↗MalfunctionDec 31, 2025Device Alarm System; Use of Device Problem; Device Alarm System; Use of Device Problem
ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-6480 PUMP IS SHOWING AN ERROR DURING USE. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23935496 ↗
FDA clearance and classification
The FDA regulates product code HRX (arthroscope) under 21 CFR 888.1100 as a Class II device. 21 CFR 888.1100 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the Arthrex® AR-6480
Has the Arthrex® AR-6480 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00888867039377), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Arthrex® AR-6480?
The FDA's MAUDE database lists 925 adverse event reports for this device, received between Jul 13, 2016 and Jul 30, 2026. In those reports, the most common specific device problems are “Pressure Problem” (185), “Increase in Pressure” (104) and “Defective Device” (85); the most common patient problems are “Swelling/ Edema” (47), “Extravasation” (34) and “Swelling” (13). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Arthrex® AR-6480?
None of the reports are classified as deaths, 91 as injuries and 834 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Arthrex® AR-6480, and is it still sold?
The FDA's device database lists Arthrex, Inc. as the company responsible for it (brand Arthrex®, model AR-6480). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ARTHROSCOPE devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,791 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE814 reports
- MITEK 242018Medos International Sàrl · Arthroscope729 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope655 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope547 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE468 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
- FMS 283812Medos International Sàrl · Arthroscope367 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page