Quick summary
This DePuy Mitek FMS VUE fluid management system is used during arthroscopic joint surgery to control fluid flow. Its 820 reports include 13 injuries and no deaths. Many describe fluid flow, pressure, suction or overfill problems, and some describe leaks. Reports fell to 39 in the last 12 months, down from 74 in the 12 months before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 814
- Deaths reported
- 0
- Injuries reported
- 13
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 814 adverse event reports that list this exact device (DI 10886705020591), received between Dec 28, 2017 and May 19, 2026 — classified as 13 injuries, 801 malfunctions.
The most frequently reported specific device problem is “Increase in Pressure” (141 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Increase in Pressure141
- Overfill122
- Improper Flow or Infusion116
- Suction Failure83
- Infusion or Flow Problem73
- Break65
- Use of Device Problem45
- Output Problem33
Patient problems
- No Clinical Signs, Symptoms or Conditions430
- No Consequences Or Impact To Patient337
- Not Applicable38
- Extravasation9
- No Known Impact Or Consequence To Patient9
- No Code Available4
- No Patient Involvement3
- Hemorrhage/Blood Loss/Bleeding2
Recent adverse event reports
The 25 most recent of 814 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 19, 2026Leak/Splash; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE FMS VUE PUMP DEVICE WAS LEAKING A LIQUID. IT WAS REPORTED IN THE SERVICE ORDER THAT THE DEVICE HAD AN UNDETERMINED MALFUNCTION. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25219910 ↗MalfunctionMay 8, 2026Use of Device Problem; Improper Flow or Infusion
IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPIC PROCEDURE THE FMS VUE PUMP DEVICE THAT THE OUTFLOW WHEEL IS NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25129670 ↗MalfunctionApr 23, 2026Nonstandard Device; Increase in Pressure; Device Remains Activated
IT WAS REPORTED THAT DURING THE START OF HIP SURGERY/SCOPE, THE STAFF INDICATED THAT THE MS VUE PUMP WASN'T REGISTERING PRESSURE PROPERLY AND KEPT RUNNING WHEN IT SHOULDN'T HAVE. THEY ENDED UP FILLING THE TUBING THROUGH THE CHAMBER AND BACK UP THE TUBING. THE WHEEL WAS NOT RUNNING TO MAINTAIN PRESSURE. THE RESOLUTION WAS TO SWAP FOR ANOTHER PUMP, AND NO PATIENT HARM WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24974972 ↗MalfunctionApr 2, 2026Overfill; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC HIP PROCEDURE THAT THE FMS VUE PUMP DEVICE 2 HAD THE SAME ISSUE, THE FILL CHAMBER WAS FILLING ENTIRELY WITH FLUID BEFORE GENERATING A LOW-PRESSURE WARNING AND TURNING OFF THE PUMP. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A SLIGHT DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24770191 ↗MalfunctionApr 2, 2026Overfill; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC HIP PROCEDURE THAT THE FMS VUE PUMP DEVICE 2 HAD THE SAME ISSUE, THE FILL CHAMBER WAS FILLING ENTIRELY WITH FLUID BEFORE GENERATING A LOW-PRESSURE WARNING AND TURNING OFF THE PUMP. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A SLIGHT DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24770190 ↗MalfunctionMar 26, 2026Increase in Pressure
IT WAS REPORTED THAT DURING AN UNKNOWN SHOULDER ARTHROSCOPIC PROCEDURE THAT THE FMS VUE PUMP DEVICE THE PUMP PRESSURE WAS READING EXTREMELY HIGH (300MMHG) AND KEPT DECREASING AND INCREASING ERRATICALLY. TUBING WAS INSPECTED AND NOT HOLES OR OCCLUSIONS WERE FOUND. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24702303 ↗MalfunctionMar 18, 2026Display or Visual Feedback Problem; Increase in Pressure
IT WAS REPORTED THAT DURING AN UNSPECIFIED PROCEDURE, WHILE PRIMING THE FILL CHAMBER OF THE FMS VUE PUMP, THE PRESSURE WAS NOT MAINTAINED IN THE LINES AND THE AIR WAS BEING INTRODUCED IN THE LINES. IT IS UNKNOWN IF ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24625465 ↗MalfunctionMar 17, 2026Increase in Pressure
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGERY, THE FMS VUE PUMP DEVICE WAS NOT HOLDING PRESSURE. IT IS UNKNOWN IF ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24615860 ↗MalfunctionMar 16, 2026Increase in Pressure
IT WAS REPORTED THAT THE FMS VUE PUMP DID NOT HOLDING PRESSURE AS IT SHOULD. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24608203 ↗MalfunctionMar 11, 2026Display or Visual Feedback Problem; Improper Flow or Infusion
IT WAS REPORTED THAT DURING AN UNKNOWN ARTHROSCOPIC PROCEDURE THAT THE FMS VUE PUMP DEVICE PUMP IS NOT CLEARING OUT THE JOINT, IT ONLY SEEMS TO CLEAR IT OUT WHEN THE SHAVER IS BEING USED. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24571770 ↗MalfunctionMar 10, 2026Increase in Pressure
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY THE PRESSURE OF A FMS VUE PUMP SPIKED UNEXPECTEDLY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24559803 ↗MalfunctionMar 6, 2026Device Remains Activated; Noise, Audible
IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE FMS VUE PUMP DEVICE SUCTION FAILED. IT WAS REPORTED IN THE SERVICE ORDER THAT THE DEVICE OUTFLOW PUMP WAS LOUD AND DOES NOT STOP AS EXPECTED. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24537107 ↗MalfunctionFeb 25, 2026Increase in Pressure
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY, THE FMS VUE PUMP WAS NOT HOLDING PRESSURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO FURTHER INFORMATION WAS PROVIDED. THERE WAS NO FURTHER INFORMATION REGARDING THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24445090 ↗MalfunctionFeb 23, 2026Backflow; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE FMS VUE PUMP DEVICE ENCOUNTERED A BACK-FLOW/CHECK VALVE FAILURE. IT WAS REPORTED IN THE SERVICE ORDER THAT THE DEVICE HAD AN UNDETERMINED MALFUNCTION. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2026. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24420488 ↗MalfunctionFeb 18, 2026Infusion or Flow Problem
IT WAS REPORTED THAT PRIOR TO SURGERY, THE FMS VUE PUMP DEVICE HAS A SUCTION PROBLEM. A SPARE DEVICE USED TO FINISH THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE AND NO DELAY REPORTED. NO MORE INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24380631 ↗MalfunctionFeb 5, 2026Break; Power Problem
IT WAS REPORTED THAT DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE FMS VUE PUMP DEVICE WAS BROKEN (2+ PIECES) : DEVICE FRACTURED. IT WAS INITIALLY REPORTED THAT, PRIOR TO A SCOPE PROCEDURE, THE PUMP HAD NO POWER AND WOULD NOT TURN ON. REPLACING THE POWER CORDS DID NOT RESTORE POWER. A SPARE DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. THERE WAS A DELAY TO THE SURGICAL PROCEDURE. HOWEVER, THE DURATION OF THE DELAY WAS NOT SPECIFIED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24268235 ↗MalfunctionFeb 4, 2026Increase in Pressure; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE FMS VUE PUMP DEVICE HAD INADEQUATE PRESSURE, WAS OUT OF TOLERANCE AND HAD AN ELECTRICAL/ELECTRONIC COMPONENT FAILURE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24261038 ↗MalfunctionJan 29, 2026Increase in Pressure; Device Remains Activated
IT WAS REPORTED THAT THE FMS VUE PUMP'S PRESSURE WAS FLUCTUATING BETWEEN 100 AND 200 WITHOUT BEING COMMANDED BY THE USER. THE ISSUE WAS OBSERVED DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. IT WAS REPORTED THAT THERE WAS NO FURTHER INFORMATION REGARDING THE ISSUE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24209440 ↗MalfunctionJan 28, 2026Computer Software Problem; Electrical /Electronic Property Problem; Environmental Particulates
IT WAS REPORTED THAT THE FMS VUE PUMP GAVE ERROR CODES FC02 AND FC14 AND THERE WAS AN ELECTRICAL BURNING SMELL. THEY USED ANOTHER PUMP TO COMPLETE THE SURGERY. THE ISSUE WAS OBSERVED DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. IT WAS REPORTED THAT THERE WAS NO FURTHER INFORMATION REGARDING THE ISSUE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24199610 ↗MalfunctionJan 27, 2026Device Contaminated During Manufacture or Shipping; Power Problem
IT WAS REPORTED THAT PRIOR TO AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE FMS VUE PUMP DEVICE WOULD NOT TURN ON. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE DEVICE HAD FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION. THERE WAS NO PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24192273 ↗MalfunctionJan 7, 2026Break; Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE FMS VUE PUMP DEVICE WAS FRACTURED. IT WAS REPORTED IN THE SERVICE ORDER THAT THE DEVICE SCREEN WAS NOT WORKING. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23990950 ↗MalfunctionDec 30, 2025Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED THAT THE FMS VUE PUMP FAILED TO GENERATE SUCTION AND DID NOT PULL DEBRIS THROUGH THE TUBING. THE PUMP WAS EXCHANGED FOR ANOTHER UNIT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23923524 ↗InjuryDec 23, 2025Computer Software Problem; Electrical /Electronic Property Problem; Appropriate Device Problem Term/Code Not Available; Computer Software Problem; Electrical /Electronic Property Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING A SCOPE SURGICAL PROCEDURE IT WAS OBSERVED THAT WHILE USING THE FMS VUE PUMP DEVICE, ERROR CODES (FC02: ERROR CODE INDICATES SYSTEM ERROR, FC15, FC14: ERROR CODE INDICATES REMOTE DEVICE FAILURE) WERE FLASHING ON AND OFF WHEN THE SHAVER HANDPIECE WAS PLUGGED IN. THERE WAS AN UNSPECIFIED AMOUNT OF DELAY IN THE PROCEDURE REPORTED. IT WAS FURTHER REPORTED THAT THE USER THEN PERFORMED AN OPEN ROTATOR CUFF REPAIR SURGERY INSTEAD OF A SCOPE PROCEDURE. THERE WERE NO REPORTS OF INJURIES OR PROLONGED HOSPITALIZATION. IT WAS NOTED THAT THE DEVICE WAS TROUBLESHOOTED ON SITE AND AS A RESULT THE HANDPIECE HAD NO ISSUES WHEN PLUGGED INTO A DIFFERENT PUMP MACHINE. ADDITIONAL HANDPIECES WERE TRIED PLUGGING INTO THIS PUMP WHICH ALL HAD SAME ISSUES. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: Insufficient Information
MAUDE report 23881694 ↗MalfunctionDec 18, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING EVALUATION OF THE DEVICE, IT WAS DETERMINED THAT THE FMS VUE PUMP DEVICE ALARM WAS NOT TRIGGERED. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN BUT WAS NOTED TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23843180 ↗MalfunctionDec 18, 2025Increase in Pressure; Increase in Pressure
IT WAS REPORTED THAT DURING A SHOULDER REPAIR PROCEDURE, THE FMS VUE PUMP WAS OPERATING AT A HIGHER PRESSURE THAN INDICATED. THE DEVICE'S PRESSURE READING WAS INACCURATE. THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23837530 ↗
FDA clearance and classification
Cleared through 510(k) K233675 on Feb 26, 2024 (decision: Substantially Equivalent). Submitted by DePuy Mitek, Inc. as “FMS VUE Fluid Management and Tissue Debridement System”. The FDA review took 102 days. FDA record ↗
Cleared through 510(k) K130169 on May 1, 2013 (decision: Substantially Equivalent). Submitted by Depuy Mitek, A Johnson & Johnson Company as “Fms Vue Fluid Managment System, Fms Connect Interface Cable”. The FDA review took 97 days. FDA record ↗
The FDA regulates product code HRX (arthroscope) under 21 CFR 888.1100 as a Class II device. 21 CFR 888.1100 ↗
Common questions about the FMS VUE 284002
Has the FMS VUE 284002 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886705020591), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the FMS VUE 284002?
The FDA's MAUDE database lists 814 adverse event reports for this device, received between Dec 28, 2017 and May 19, 2026. In those reports, the most common specific device problems are “Increase in Pressure” (141), “Overfill” (122) and “Improper Flow or Infusion” (116); the most common patient problems are “Not Applicable” (38), “Extravasation” (9) and “Hemorrhage/Blood Loss/Bleeding” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the FMS VUE 284002?
None of the reports are classified as deaths, 13 as injuries and 801 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the FMS VUE 284002 FDA approved?
It was cleared through 510(k) K130169 on May 1, 2013; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the FMS VUE 284002, and is it still sold?
The FDA's device database lists Depuy Mitek, LLC as the company responsible for it (brand FMS VUE, model 284002). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ARTHROSCOPE devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,791 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE925 reports
- MITEK 242018Medos International Sàrl · Arthroscope729 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope655 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope547 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE468 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
- FMS 283812Medos International Sàrl · Arthroscope367 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page